Gastrointestinal Endoscopy Oct 2021Consensus regarding an optimal algorithm for endoscopic treatment of papillary adenomas has not been established. We aimed to assess the existing degree of consensus...
BACKGROUND AND AIMS
Consensus regarding an optimal algorithm for endoscopic treatment of papillary adenomas has not been established. We aimed to assess the existing degree of consensus among international experts and develop further concordance by means of a Delphi process.
METHODS
Fifty-two international experts in the field of endoscopic papillectomy were invited to participate. Data were collected between August and December 2019 using an online survey platform. Three rounds were conducted. Consensus was defined as ≥70% agreement.
RESULTS
Sixteen experts (31%) completed the full process, and consensus was achieved on 47 of the final 79 statements (59%). Diagnostic workup should include at least an upper endoscopy using a duodenoscope (100%) and biopsy sampling (94%). There should be selected use of additional abdominal imaging (75%-81%). Patients with (suspected) papillary malignancy or over 1 cm intraductal extension should be referred for surgical resection (76%). To prevent pancreatitis, rectal nonsteroidal anti-inflammatory drugs should be administered before resection (82%) and a pancreatic stent should be placed (100%). A biliary stent is indicated in case of ongoing bleeding from the papillary region (76%) or concerns for a (micro)perforation after resection (88%). Follow-up should be started 3 to 6 months after initial papillectomy and repeated every 6 to 12 months for at least 5 years (75%).
CONCLUSIONS
This is the first step in developing an international consensus-based algorithm for endoscopic management of papillary adenomas. Surprisingly, in many areas consensus could not be achieved. These aspects should be the focus of future studies.
Topics: Ampulla of Vater; Common Bile Duct Neoplasms; Consensus; Delphi Technique; Endoscopy; Humans; Treatment Outcome
PubMed: 33887269
DOI: 10.1016/j.gie.2021.04.009 -
Gastrointestinal Endoscopy Sep 2020Contaminated duodenoscopes and linear echoendoscopes (DLEs) pose a risk for infectious outbreaks. To identify DLEs and reprocessing risk factors, we combined the data...
BACKGROUND AND AIMS
Contaminated duodenoscopes and linear echoendoscopes (DLEs) pose a risk for infectious outbreaks. To identify DLEs and reprocessing risk factors, we combined the data from the previously published nationwide cross-sectional PROCESS 1 study (Prevalence of contamination of complex endoscopes in the Netherlands) with the follow-up PROCESS 2 study.
METHODS
We invited all 74 Dutch DLE centers to sample ≥2 duodenoscopes during PROCESS 1, and all duodenoscopes as well as linear echoendoscopes during PROCESS 2. The studies took place 1 year after another. Local staff sampled each DLE at ≤6 sites according to uniform methods explained by online videos. We used 2 contamination definitions: (1) any microorganism with ≥20 colony-forming units (CFU)/20 mL (AM20) and (2) presence of microorganisms with GI or oral origin, independent of CFU count (MGOs). We assessed the factors of age and usage by performing an analysis of pooled data of both PROCESS studies; additional factors including reprocessing characteristics were only recorded in PROCESS 2.
RESULTS
Ninety-seven percent of all Dutch centers (72 of 74; PROCESS 1, 66; PROCESS 2, 61) participated in one of the studies, sampling 309 duodenoscopes and 64 linear echoendoscopes. In total, 54 (17%) duodenoscopes and 8 (13%) linear echoendoscopes were contaminated according to the AM20 definition. MGOs were detected on 47 (15%) duodenoscopes and 9 (14%) linear echoendoscopes. Contamination was not age or usage dependent (all P values ≥.27) and was not shown to differ between the reprocessing characteristics (all P values ≥.01).
CONCLUSIONS
In these nationwide studies, we found that DLE contamination was independent of age and usage. These results suggest that old and heavily used DLEs, if maintained correctly, have a similar risk for contamination as new DLEs. The prevalence of MGO contamination of ∼15% was similarly high for duodenoscopes as for linear echoendoscopes, rendering patients undergoing ERCP and EUS at risk for transmission of microorganisms.
Topics: Cross Infection; Cross-Sectional Studies; Disinfection; Duodenoscopes; Equipment Contamination; Humans; Netherlands; Risk Assessment
PubMed: 32502549
DOI: 10.1016/j.gie.2020.05.030 -
Journal of Personalized Medicine May 2021Endoscopic retrograde cholangiopancreatography (ERCP) in patients who have undergone a Billroth II gastrectomy is a major challenge. This study aimed to evaluate the...
Endoscopic retrograde cholangiopancreatography (ERCP) in patients who have undergone a Billroth II gastrectomy is a major challenge. This study aimed to evaluate the outcomes of the road-map technique for duodenal intubation using a side-viewing duodenoscope for ERCP in Billroth II gastrectomy patients with naïve papilla, and to analyze the formation and release patterns of common bowel loops that occur when the duodenoscope navigates the afferent limb. The duodenoscopy approach success rate was 85.8% (97/113). In successful duodenoscopy approach patients, there were five bowel looping patterns that occurred when the preceding catheter-connected duodenoscope was advanced into the duodenum: (1) reverse ɣ-loop (29.9%), (2) fixed reverse ɣ-loop (5.2%), (3) simple U-loop (22.7%), (4) N-loop (28.9%), and (5) reverse alpha loop (13.4%). The duodenoscopy cannulation and duodenoscopy therapeutic success rates were 81.4% (92/113) and 80.5% (91/113), respectively, while the overall cannulation and therapeutic success rates were 92.0% (104/113) and 87.6% (99/113), respectively. Bowel perforation occurred in three patients (2.7%). The road-map technique may benefit duodenoscope-based ERCP in Billroth II gastrectomy patients by minimizing the tangential axis alignment between the duodenoscopic tip and driving of the afferent limb, and by predicting and counteracting bowel loops that occur when the duodenoscope navigates the afferent limb.
PubMed: 34066235
DOI: 10.3390/jpm11050404 -
Infection and Drug Resistance 2018Duodenoscopes have been widely used for both diagnostic and therapeutic endoscopic retrograde cholangiopancreatography procedures. Numerous outbreaks of...
INTRODUCTION
Duodenoscopes have been widely used for both diagnostic and therapeutic endoscopic retrograde cholangiopancreatography procedures. Numerous outbreaks of duodenoscope-associated infections involving multidrug-resistant bacteria have recently been reported. Plasma activated water (PAW) has been widely considered an effective agent for surface decontamination and is increasingly used for disinfection of medical equipment. The aim of this study was to evaluate whether the duodenoscopes currently on market are suited for the repeated use of PAW and to test the efficacy of PAW for their disinfection.
MATERIALS AND METHODS
In order to evaluate the disinfection efficacy and the required time of contact, the duodenoscope samples were contaminated by immersing them in fasted-state simulated intestinal fluid containing , , , and , prior to PAW exposure. In order to test the duodenoscope polymer compatibility with PAW, a challenge test was conducted by immersing the samples in PAW for 30 minutes daily for 45 consecutive days.
RESULTS
Significant reductions in bacterial populations were achieved after 30 minutes of PAW treatment, indicating a high-level disinfection. Atomic force microscopy and scanning electron microscopy were used to demonstrate that repeated PAW treatment of duodenoscope coating polymer samples did not result in significant differences in morphological surface between the treated and untreated samples. Energy-dispersive X-ray spectroscopy analysis also showed no significant differences between the elemental composition of the duodenoscope coating polymer samples before and after repeated PAW treatment.
CONCLUSION
Considering these preliminary results, PAW could be considered as a new alternative for duodenoscope reprocessing.
PubMed: 29844690
DOI: 10.2147/IDR.S159243 -
Molecules (Basel, Switzerland) Jun 2019Duodenoscopes have been widely used for both diagnostic and therapeutic endoscopic retrograde cholangiopancreatography (ERCP) procedures, but recently, numerous...
OBJECTIVES
Duodenoscopes have been widely used for both diagnostic and therapeutic endoscopic retrograde cholangiopancreatography (ERCP) procedures, but recently, numerous outbreaks of multidrug-resistant organisms (MDRO) infections have been reported which has led to extensive research for their possible causes. Consequently, the aim of this study is to search for possible duodenoscope surface damages that could provide an alternative and plausible source of infections.
MATERIALS AND METHODS
In order to assess both outer and inner surfaces, a duodenoscope was dismantled and samples were taken from the outer resin polymer and from the air/water, elevator, and working (biopsy) channels that were characterized by FTIR, DSC, TGA, AFM, SEM techniques and the antimicrobial activity were tested.
RESULTS
Alterations were noticed on both the coating and working channel polymers, with external alterations increasing progressively from the proximal sample to the distal sample near the tip of the scope. However, the results showed that the coating surface was still efficient against bacterial adhesion. Changes in surface texture and also morphological changes were shown.
CONCLUSIONS
The study describes the impact of routine procedural use and reprocessing cycles on the duodenoscope, showing that these may possibly make it susceptible to bacterial contamination and MDRO biofilm formation due to difficult reprocessing of the altered surfaces.
Topics: Biofilms; Calorimetry, Differential Scanning; Cholangiopancreatography, Endoscopic Retrograde; Cross Infection; Disinfection; Duodenoscopes; Environmental Microbiology; Equipment Reuse; Humans; Microscopy, Atomic Force; Spectroscopy, Fourier Transform Infrared; Thermogravimetry
PubMed: 31242689
DOI: 10.3390/molecules24122343 -
Endoscopy Feb 2022The risk of exogenous infections from endoscopic procedures is often cited as almost negligible (1 infection in 1.8 million procedures); however, this risk is based on...
BACKGROUND
The risk of exogenous infections from endoscopic procedures is often cited as almost negligible (1 infection in 1.8 million procedures); however, this risk is based on older literature and does not seem to match the number of infectious outbreaks due to contaminated duodenoscopes reported after endoscopic retrograde cholangiopancreatography (ERCP). Using Dutch data, we aimed to estimate the minimum risk of duodenoscope-associated infection (DAI) and colonization (DAC) in patients undergoing ERCP.
METHODS
A systematic literature search identified all DAI outbreaks in the Netherlands reported between 2008 and 2019. Included cases were confirmed by molecular matching of patient and duodenoscope cultures. Risk ratios were calculated based on the total number of ERCPs performed during the study period.
RESULTS
Three outbreaks were reported and published between 2008 and 2018, including 21 confirmed DAI cases and 52 confirmed DAC cases. The estimated number of ERCPs performed during the same period was 181 209-227 006. The calculated minimum estimated DAI risk was approximately 0.01 % and the minimum estimated DAC risk was 0.023 %-0.029 %.
CONCLUSIONS
The estimated risk of DAI in Dutch ERCP practice was at least 180 times higher than previously published risk estimates. The actual risk is likely to be (much) higher due to underreporting of infections caused by multidrug-resistant organisms and sensitive bacteria. Greater awareness by healthcare personnel involved in endoscopy and endoscope cleaning is required, as well as innovative technical solutions to contain and ultimately eliminate DAIs.
Topics: Cholangiopancreatography, Endoscopic Retrograde; Disease Outbreaks; Duodenoscopes; Equipment Contamination; Humans; Netherlands
PubMed: 33915575
DOI: 10.1055/a-1467-6294 -
Clinical Infectious Diseases : An... Apr 2019Clinicians increasingly utilize polymyxins for treatment of serious infections caused by multidrug-resistant gram-negative bacteria. Emergence of plasmid-mediated,...
BACKGROUND
Clinicians increasingly utilize polymyxins for treatment of serious infections caused by multidrug-resistant gram-negative bacteria. Emergence of plasmid-mediated, mobile colistin resistance genes creates potential for rapid spread of polymyxin resistance. We investigated the possible transmission of Klebsiella pneumoniae carrying mcr-1 via duodenoscope and report the first documented healthcare transmission of mcr-1-harboring bacteria in the United States.
METHODS
A field investigation, including screening targeted high-risk groups, evaluation of the duodenoscope, and genome sequencing of isolated organisms, was conducted. The study site included a tertiary care academic health center in Boston, Massachusetts, and extended to community locations in New England.
RESULTS
Two patients had highly related mcr-1-positive K. pneumoniae isolated from clinical cultures; a duodenoscope was the only identified epidemiological link. Screening tests for mcr-1 in 20 healthcare contacts and 2 household contacts were negative. Klebsiella pneumoniae and Escherichia coli were recovered from the duodenoscope; neither carried mcr-1. Evaluation of the duodenoscope identified intrusion of biomaterial under the sealed distal cap; devices were recalled to repair this defect.
CONCLUSIONS
We identified transmission of mcr-1 in a United States acute care hospital that likely occurred via duodenoscope despite no identifiable breaches in reprocessing or infection control practices. Duodenoscope design flaws leading to transmission of multidrug-resistant organsisms persist despite recent initiatives to improve device safety. Reliable detection of colistin resistance is currently challenging for clinical laboratories, particularly given the absence of a US Food and Drug Administration-cleared test; improved clinical laboratory capacity for colistin susceptibility testing is needed to prevent the spread of mcr-carrying bacteria in healthcare settings.
Topics: Colistin; Drug Resistance, Multiple, Bacterial; Duodenoscopes; Equipment Contamination; Humans; Klebsiella pneumoniae; Microbial Sensitivity Tests; United States
PubMed: 30204838
DOI: 10.1093/cid/ciy683 -
Experimental and Therapeutic Medicine Jan 2021The study aimed to investigate the clinical efficacy of duodenoscopy combined with laparoscopy in the treatment of patients with severe acute pancreatitis (SAP) and...
Clinical efficacy of duodenoscopy combined with laparoscopy in the treatment of patients with severe acute pancreatitis and pancreatic pseudocyst, and the effects on IL-6 and CRP.
The study aimed to investigate the clinical efficacy of duodenoscopy combined with laparoscopy in the treatment of patients with severe acute pancreatitis (SAP) and pancreatic pseudocyst (PP), and its effects on serum inflammatory factors. Altogether 94 patients complicated with SAP and PP who were admitted to Weifang People's Hospital (Weifang, China) from September 2015 to December 2018 were included. Based on the different operation methods, 49 patients who underwent traditional laparotomy under laparoscopic surgery were included in group A, and 45 patients who underwent duodenoscopy and laparoscopy under duodenoscope to treat the drainage of nipple and pancreatic pseudocysts were included in group B. The expression levels of related serum indexes and serum stress indexes before and at 48 h after surgery, the postoperative nausea, vomiting and abdominal pain scores, as well as the clinical efficacy, perioperative related indexes, recovery and complications were compared between the two groups. The prognostic factors in both groups were assessed via Logistic univariate and multivariate analyses. C-reactive protein (CRP), tumor necrosis factor-α (TNF-α), interleukin-6 (IL-6), interleukin-β (IL-β), endotoxin and nuclear factor κB (NF-κB) were significantly lower in group B than those in group A (P<0.001). Upregulating cortisol and norepinephrine in group B was lower than that in group A (P<0.001). The total effective rate in group B was higher than that in group A (P<0.05). The perioperative related indexes, recovery, and postoperative complications in group B were better than those in group A (P<0.05). Scores of abdominal pain, nausea and vomiting in group B were markedly lower than those in group A (P<0.001). Multivariate Logistic regression analysis showed that CRP, TNF-α, IL-6, IL-β and surgical methods were independent risk factors for the prognosis of patients with SAP and PP. In conclusion, the combined treatment with duodenoscopy and laparoscopic surgery has little inflammatory and stress reaction, and it is highly safe, worthy to be popularized.
PubMed: 33273983
DOI: 10.3892/etm.2020.9487 -
Clinical Microbiology and Infection :... Oct 2018Several outbreaks of severe infections due to contamination of gastrointestinal (GI) endoscopes, mainly duodenoscopes, have been described. The rate of microbial...
OBJECTIVES
Several outbreaks of severe infections due to contamination of gastrointestinal (GI) endoscopes, mainly duodenoscopes, have been described. The rate of microbial endoscope contamination varies dramatically in literature. The aim of this multicentre prospective study was to evaluate the hygiene quality of endoscopes and automated endoscope reprocessors (AERs) in Tyrol/Austria.
METHODS
In 2015 and 2016, a total of 463 GI endoscopes and 105 AERs from 29 endoscopy centres were analysed by a routine (R) and a combined routine and advanced (CRA) sampling procedure and investigated for microbial contamination by culture-based and molecular-based analyses.
RESULTS
The contamination rate of GI endoscopes was 1.3%-4.6% according to the national guideline, suggesting that 1.3-4.6 patients out of 100 could have had contacts with hygiene-relevant microorganisms through an endoscopic intervention. Comparison of R and CRA sampling showed 1.8% of R versus 4.6% of CRA failing the acceptance criteria in phase I and 1.3% of R versus 3.0% of CRA samples failing in phase II. The most commonly identified indicator organism was Pseudomonas spp., mainly Pseudomonas oleovorans. None of the tested viruses were detected in 40 samples. While AERs in phase I failed (n = 9, 17.6%) mainly due to technical faults, phase II revealed lapses (n = 6, 11.5%) only on account of microbial contamination of the last rinsing water, mainly with Pseudomonas spp.
CONCLUSIONS
In the present study the contamination rate of endoscopes was low compared with results from other European countries, possibly due to the high quality of endoscope reprocessing, drying and storage.
Topics: Austria; Cross Infection; Decontamination; Endoscopes, Gastrointestinal; Equipment Contamination; Europe; Humans; Prospective Studies; Pseudomonas
PubMed: 29408276
DOI: 10.1016/j.cmi.2018.01.017 -
BMC Gastroenterology Apr 2023Endoscopic retrograde cholangiography (ERC) possesses a translocation risk of microbes to the biliary system. We studied bile contamination during ERC and its impact on...
BACKGROUND
Endoscopic retrograde cholangiography (ERC) possesses a translocation risk of microbes to the biliary system. We studied bile contamination during ERC and its impact on patients' outcome in a real-life-situation.
METHODS
Ninety-nine ERCs were analyzed and microbial samples were taken from the throat before and from bile during ERC and from irrigation fluid of the duodenoscope before and after ERC.
RESULTS
91.2% of cholangitis patients had detectable microbes in the bile (sensitivity 91%), but the same was true for 86.2% in the non-cholangitis group. Bacteroides fragilis (p=0.015) was significantly associated with cholangitis. In 41.7% of ERCs with contaminated endoscopes these microbes were found in the bile after the procedure. Analysis of duodenoscopes' irrigation liquid after ERC matched the microbial bile analysis of these patients in 78.8%. Identical microbial species were in throat and in bile samples of the same ERC in 33% of all cases and in 45% in the non-cholangitis group. Transmission of microbes to the biliary tract did not result in more frequent cholangitis, longer hospital stays, or worse outcome.
CONCLUSIONS
During ERC bile samples are regularly contaminated with microbes of the oral cavity but it did not affect clinical outcome.
Topics: Humans; Cholangiopancreatography, Endoscopic Retrograde; Biliary Tract; Cholangiography; Biliary Tract Surgical Procedures; Cholangitis; Microbiota
PubMed: 37013522
DOI: 10.1186/s12876-023-02721-7