Journal of Clinical Microbiology Oct 2015Inadequate flexible endoscope reprocessing has been associated with infection outbreaks, most recently caused by carbapenem-resistant Enterobacteriaceae. Lapses in... (Review)
Review
Inadequate flexible endoscope reprocessing has been associated with infection outbreaks, most recently caused by carbapenem-resistant Enterobacteriaceae. Lapses in essential device reprocessing steps such as cleaning, disinfection/sterilization, and storage have been reported, but some outbreaks have occurred despite claimed adherence to established guidelines. Recommended changes in these guidelines include the use of sterilization instead of high-level disinfection or the use of routine microbial culturing to monitor efficacy of reprocessing. This review describes the current standards for endoscope reprocessing, associated outbreaks, and the complexities associated with both microbiological culture and sterilization approaches to mitigating the risk of infection associated with endoscopy.
Topics: Cross Infection; Decontamination; Disinfection; Endoscopes; Humans; Practice Guidelines as Topic
PubMed: 26202125
DOI: 10.1128/JCM.01394-15 -
Acta Gastro-enterologica Belgica 2021Data on procedural outcome and quality of endoscopic retrograde cholangiopancreatography (ERCP) in Belgian practice are scarce. The aim of this study is to assess...
BACKGROUND AND STUDY AIMS
Data on procedural outcome and quality of endoscopic retrograde cholangiopancreatography (ERCP) in Belgian practice are scarce. The aim of this study is to assess current status of ERCP-performance in Belgium.
METHODS
National multi-institutional survey (online questionnaire) among members of the Belgian Society of Gastrointestinal Endoscopy (BSGIE), conducted in the period June-August 2018. The RIZIV/INAMI provided real-life data on the total number of ERCPs performed in Belgium.
RESULTS
Forty-five responders completed the survey (for 43 centers performing ERCP), providing information for 8368 ERCPs performed in 45% (43/95) of institutions performing ERCP in Belgium. Fifty-eight percent of centers performed > 100 ERCPs/year and 7% of centers (n=3) performed < 50 ERCPs/year. According to the RIZIV/INAMI data, low case-volume centers are underrepresented in this survey. The most common ERCPindication was stone extraction (52%). 74% of endoscopists had more than 10 years of experience in performing ERCP. The majority of centers had their own written protocol (84%) for microbiological duodenoscope surveillance. Monitoring of cannulation rate and post-ERCP pancreatitis (PEP) was only performed in a minority of centers (30%). The majority of centers (76%) provided verbal informed consent relating to the ERCP-procedure ; a minority also requested a written informed consent (23%). 65% of centers systematically use NSAIDs for PEP prophylaxis.
CONCLUSION
This is the first survey of ERCP performance in Belgium. There were wide variations in practice. Adherence to key performance measures and measurement and evaluation of ERCP performance in daily practice at center and endoscopist level are not uniformly widespread.
Topics: Belgium; Cholangiopancreatography, Endoscopic Retrograde; Endoscopy, Gastrointestinal; Humans; Pancreatitis; Surveys and Questionnaires
PubMed: 33639696
DOI: 10.51821/84.1.128 -
Clinical and Translational... Aug 2020
Review
Topics: Cholangiopancreatography, Endoscopic Retrograde; Disinfection; Duodenoscopes; Equipment Design; Focus Groups; Humans; Infection Control; Qualitative Research; Surgical Wound Infection
PubMed: 32955193
DOI: 10.14309/ctg.0000000000000214 -
Journal of Personalized Medicine May 2021Endoscopic retrograde cholangiopancreatography (ERCP) in patients who have undergone a Billroth II gastrectomy is a major challenge. This study aimed to evaluate the...
Endoscopic retrograde cholangiopancreatography (ERCP) in patients who have undergone a Billroth II gastrectomy is a major challenge. This study aimed to evaluate the outcomes of the road-map technique for duodenal intubation using a side-viewing duodenoscope for ERCP in Billroth II gastrectomy patients with naïve papilla, and to analyze the formation and release patterns of common bowel loops that occur when the duodenoscope navigates the afferent limb. The duodenoscopy approach success rate was 85.8% (97/113). In successful duodenoscopy approach patients, there were five bowel looping patterns that occurred when the preceding catheter-connected duodenoscope was advanced into the duodenum: (1) reverse ɣ-loop (29.9%), (2) fixed reverse ɣ-loop (5.2%), (3) simple U-loop (22.7%), (4) N-loop (28.9%), and (5) reverse alpha loop (13.4%). The duodenoscopy cannulation and duodenoscopy therapeutic success rates were 81.4% (92/113) and 80.5% (91/113), respectively, while the overall cannulation and therapeutic success rates were 92.0% (104/113) and 87.6% (99/113), respectively. Bowel perforation occurred in three patients (2.7%). The road-map technique may benefit duodenoscope-based ERCP in Billroth II gastrectomy patients by minimizing the tangential axis alignment between the duodenoscopic tip and driving of the afferent limb, and by predicting and counteracting bowel loops that occur when the duodenoscope navigates the afferent limb.
PubMed: 34066235
DOI: 10.3390/jpm11050404 -
Surgical Endoscopy Dec 2016The aim of this study was to investigate the value of a single stage with combined tri-endoscopic (duodenoscopy, laparoscopy and choledochoscopy) approach for patients... (Randomized Controlled Trial)
Randomized Controlled Trial
OBJECTIVE
The aim of this study was to investigate the value of a single stage with combined tri-endoscopic (duodenoscopy, laparoscopy and choledochoscopy) approach for patients with concomitant cholecystolithiasis and choledocholithiasis.
METHODS
Fifty-three patients with combined gallbladder stones and common bile duct stones from February 2014 to April 2015 were randomized assigned to two groups: 29 patients underwent single-stage surgery with combined duodenoscope, laparoscope and choledochoscope (combined tri-endoscopic group), and 29 patients underwent endoscopic sphincterotomy to remove common bile duct stones followed by laparoscopic cholecystectomy several days later (control group). The success rate of complete stone removal, procedure-related complication, hospital stay and the cost of hospitalization were compared between the two groups.
RESULTS
Altogether, 53 patients (29 patients in combined tri-endoscopic group and 24 patients in control group) successfully underwent the surgery and ERCP procedure. Three patients in the control group developed post-ERCP pancreatitis. One case of bile leaking and one case of residual stone were noted in the combined tri-endoscopic group. There were no significant differences between the two groups with regard to both complete stone removal [96.6 % (28/29) vs. 100 % (24/24)] and procedure-related complication rate [3.4 % (1/29) vs. 12.5 % (3/24)] (p > 0.05). No open surgery was required in either group. There were significant differences between the two groups with regard to hospital stay (6.72 ± 1.3 days vs. 10.91 ± 1.6 days, p < 0.01) and cost of hospitalization (15,724 ± 1613 CNY vs. 19,829 ± 2433 CNY, p < 0.05).
CONCLUSION
The single-stage combined tri-endoscopic approach for concomitant cholecystolithiasis and choledocholithiasis was just as safe and successful as the control group. In addition, it resulted in a shorter hospital stay and less cost.
Topics: Cholecystectomy, Laparoscopic; Cholecystolithiasis; Choledocholithiasis; Duodenoscopes; Endoscopy, Gastrointestinal; Female; Hospital Costs; Humans; Laparoscopes; Laparoscopy; Length of Stay; Male; Middle Aged; Sphincterotomy, Endoscopic
PubMed: 27126621
DOI: 10.1007/s00464-016-4918-6 -
The American Surgeon Apr 2024The use of endoscopic retrograde cholangiography (ERCP) for diagnostic and therapeutic interventions on the pancreaticobiliary system has steadily increased, but the...
The use of endoscopic retrograde cholangiography (ERCP) for diagnostic and therapeutic interventions on the pancreaticobiliary system has steadily increased, but the standard approach through the oropharynx is prohibited after Roux-en-Y (RYGB) gastric bypass surgery. Laparoscopic access to the gastric remnant allows for the completion of ERCP using the standard side-viewing duodenoscope to facilitate the completion of standard and advanced endoscopic maneuvers. Here, we describe our experience with the technical aspects of safe and effective performance of laparoscopic-assisted ERCP.
PubMed: 38587416
DOI: 10.1177/00031348241244651 -
World Journal of Gastroenterology Oct 2020The management strategies for recurrent ampullary adenoma after endoscopic papillectomy are still controversial. Patients with the recurrent papillary lesions need to...
BACKGROUND
The management strategies for recurrent ampullary adenoma after endoscopic papillectomy are still controversial. Patients with the recurrent papillary lesions need to receive repetitive endoscopic interventions due to the limitations of conventional endoscopic techniques.
AIM
To assess the feasibility, efficacy, and safety of hybrid endoscopic submucosal dissection (ESD) by duodenoscope for recurrent, laterally spreading papillary lesions.
METHODS
We enrolled two patients with recurrent, laterally spreading, duodenal papillary adenomas with no intraductal extension confirmed by follow-up between March 2017 and September 2018. After marking the resection borders of the lesion using a dual knife, a submucosal cushion was created by injecting a mixture of saline solution, methylene blue, and adrenaline. A total circumferential incision and submucosal excision was performed by dual knife combined with insulated-tip diathermic knife, and then the lesion was ligated and resected using an electric snare. Endoscopic hemostasis was applied during the endoscopic procedures. Moreover, the endoscopic retrograde cholangiopancreatography (ERCP) procedures, including selective cannulation and stent implantation of biliary and pancreatic ducts, were performed. Additionally, we performed endoclip closure for mucosal defect after ESD.
RESULTS
Hybrid ESD using a duodenoscope and biliary and pancreatic stent placement were performed successfully in two patients. The endoscopic size of recurrent papillary lesions was no more than 2 cm. Generally, the average total procedure time was 95.5 min, and the procedure time of ESD and ERCP was 38.5 min and 15.5 min, respectively. No serious complications occurred during the intraoperative and postoperative periods. The histopathological examination revealed tubulovillous adenoma negative for neoplastic extension at the cut margin in both patients. The duodenoscopic follow-up and histopathology of biopsy specimens at 3 mo after ESD showed no residual or recurrent lesions in ampullary areas in both cases. Both cases have been followed up with no recurrence to June 2020.
CONCLUSION
Hybrid ESD by duodenoscope is technically challenging, and may be curative for recurrent, laterally spreading papillary adenomas < 2 cm. It should be performed cautiously in selected patients by experienced endoscopists.
Topics: Adenoma; Duodenal Neoplasms; Duodenoscopes; Endoscopic Mucosal Resection; Humans; Intestinal Mucosa; Treatment Outcome
PubMed: 33088160
DOI: 10.3748/wjg.v26.i37.5673 -
Gastrointestinal Endoscopy Jun 2017Recent outbreaks of duodenoscope-transmitted infections underscore the importance of adequate endoscope reprocessing. Adenosine triphosphate (ATP) bioluminescence...
BACKGROUND AND AIMS
Recent outbreaks of duodenoscope-transmitted infections underscore the importance of adequate endoscope reprocessing. Adenosine triphosphate (ATP) bioluminescence testing allows rapid evaluation of endoscopes for bacteriologic/biologic residue. In this prospective study we evaluate the utility of ATP in bacteriologic surveillance and the effects of endoscopy staff education and dual cycles of cleaning and high-level disinfection (HLD) on endoscope reprocessing.
METHODS
ATP bioluminescence was measured after precleaning, manual cleaning, and HLD on rinsates from suction-biopsy channels of all endoscopes and elevator channels of duodenoscopes/linear echoendoscopes after use. ATP bioluminescence was remeasured in duodenoscopes (1) after re-education and competency testing of endoscopy staff and subsequently (2) after 2 cycles of precleaning and manual cleaning and single cycle of HLD or (3) after 2 cycles of precleaning, manual cleaning, and HLD.
RESULTS
The ideal ATP bioluminescence benchmark of <200 relative light units (RLUs) after manual cleaning was achieved from suction-biopsy channel rinsates of all endoscopes, but 9 of 10 duodenoscope elevator channel rinsates failed to meet this benchmark. Re-education reduced RLUs in duodenoscope elevator channel rinsates after precleaning (23,218.0 vs 1340.5 RLUs, P < .01) and HLD (177.0 vs 12.0 RLUs, P < .01). After 2 cycles of manual cleaning/HLD, duodenoscope elevator channel RLUs achieved levels similar to sterile water, with corresponding negative cultures.
CONCLUSIONS
ATP testing offers a rapid, inexpensive alternative for detection of endoscope microbial residue. Re-education of endoscopy staff and 2 cycles of cleaning and HLD decreased elevator channel RLUs to levels similar to sterile water and may therefore minimize the risk of transmission of infections by duodenoscopes.
Topics: Adenosine Triphosphate; Cross Infection; Disease Transmission, Infectious; Disinfection; Duodenoscopes; Endoscopy, Gastrointestinal; Equipment Contamination; Equipment Reuse; Humans; Luminescent Measurements; Professional Competence
PubMed: 27818222
DOI: 10.1016/j.gie.2016.10.035 -
JAMA Oct 2014Carbapenem-resistant Enterobacteriaceae (CRE) producing the New Delhi metallo-β-lactamase (NDM) are rare in the United States, but have the potential to add to the...
IMPORTANCE
Carbapenem-resistant Enterobacteriaceae (CRE) producing the New Delhi metallo-β-lactamase (NDM) are rare in the United States, but have the potential to add to the increasing CRE burden. Previous NDM-producing CRE clusters have been attributed to person-to-person transmission in health care facilities.
OBJECTIVE
To identify a source for, and interrupt transmission of, NDM-producing CRE in a northeastern Illinois hospital.
DESIGN, SETTING, AND PARTICIPANTS
Outbreak investigation among 39 case patients at a tertiary care hospital in northeastern Illinois, including a case-control study, infection control assessment, and collection of environmental and device cultures; patient and environmental isolate relatedness was evaluated with pulsed-field gel electrophoresis (PFGE). Following identification of a likely source, targeted patient notification and CRE screening cultures were performed.
MAIN OUTCOMES AND MEASURES
Association between exposure and acquisition of NDM-producing CRE; results of environmental cultures and organism typing.
RESULTS
In total, 39 case patients were identified from January 2013 through December 2013, 35 with duodenoscope exposure in 1 hospital. No lapses in duodenoscope reprocessing were identified; however, NDM-producing Escherichia coli was recovered from a reprocessed duodenoscope and shared more than 92% similarity to all case patient isolates by PFGE. Based on the case-control study, case patients had significantly higher odds of being exposed to a duodenoscope (odds ratio [OR], 78 [95% CI, 6.0-1008], P < .001). After the hospital changed its reprocessing procedure from automated high-level disinfection with ortho-phthalaldehyde to gas sterilization with ethylene oxide, no additional case patients were identified.
CONCLUSIONS AND RELEVANCE
In this investigation, exposure to duodenoscopes with bacterial contamination was associated with apparent transmission of NDM-producing E coli among patients at 1 hospital. Bacterial contamination of duodenoscopes appeared to persist despite the absence of recognized reprocessing lapses. Facilities should be aware of the potential for transmission of bacteria including antimicrobial-resistant organisms via this route and should conduct regular reviews of their duodenoscope reprocessing procedures to ensure optimal manual cleaning and disinfection.
Topics: Adult; Aged; Aged, 80 and over; Carbapenems; Case-Control Studies; Cohort Studies; Cross Infection; Disease Outbreaks; Disinfection; Drug Resistance, Bacterial; Duodenoscopes; Enterobacteriaceae Infections; Equipment Contamination; Escherichia coli; Female; Hospitals; Humans; Illinois; Male; Middle Aged; beta-Lactamases
PubMed: 25291580
DOI: 10.1001/jama.2014.12720 -
The Journal of Hospital Infection May 2024Duodenoscope-associated infections (DAIs) are exogenous infections resulting from the use of contaminated duodenoscopes. Though numerous outbreaks of DAI have involved... (Observational Study)
Observational Study
BACKGROUND
Duodenoscope-associated infections (DAIs) are exogenous infections resulting from the use of contaminated duodenoscopes. Though numerous outbreaks of DAI have involved multidrug-resistant micro-organisms (MDROs), outbreaks involving non-MDROs are also likely to occur. Detection challenges arise as these infections often resolve before culture or because causative strains are not retained for comparison with duodenoscope strains.
AIM
To identify and analyse DAIs spanning a seven-year period in a tertiary care medical centre.
METHODS
This was a retrospective observational study. Duodenoscope cultures positive for gastrointestinal flora between March 2015 and September 2022 were paired with duodenoscope usage data to identify patients exposed to contaminated duodenoscopes. Analysis encompassed patients treated after a positive duodenoscope culture and those treated within the interval from a negative to a positive culture. Patient identification numbers were cross-referenced with a clinical culture database to identify patients developing infections with matching micro-organisms within one year of their procedure. A 'pair' was established upon a species-level match between duodenoscope and patient cultures. Pairs were further analysed via antibiogram comparison, and by whole-genome sequencing (WGS) to determine genetic relatedness.
FINDINGS
Sixty-eight pairs were identified; of these, 21 exhibited matching antibiograms which underwent WGS, uncovering two genetically closely related pairs categorized as DAIs. Infection onset occurred up to two months post procedure. Both causative agents were non-MDROs.
CONCLUSION
This study provides crucial insights into DAIs caused by non-MDROs and it highlights the challenge of DAI recognition in daily practice. Importantly, the delayed manifestation of the described DAIs suggests a current underestimation of DAI risk.
Topics: Humans; Retrospective Studies; Duodenoscopes; Tertiary Care Centers; Microbial Sensitivity Tests; Male; Female; Bacteria; Equipment Contamination
PubMed: 38447805
DOI: 10.1016/j.jhin.2024.02.015