Europace : European Pacing,... Dec 2023Single-connector (DF4) defibrillator leads have become the predominantly implanted transvenous implantable cardioverter-defibrillator lead. However, data on their... (Review)
Review
AIMS
Single-connector (DF4) defibrillator leads have become the predominantly implanted transvenous implantable cardioverter-defibrillator lead. However, data on their long-term performance are derived predominantly from manufacturer product performance reports.
METHODS AND RESULTS
We reviewed medical records in 5289 patients with DF4 leads between 2011 and 2023 to determine the frequency of lead-related abnormalities. We defined malfunction as any single or combination of electrical abnormalities requiring revision including a sudden increase (≥2×) in stimulation threshold, a discrete jump in high-voltage impedance, or sensing of non-physiologic intervals or noise. We documented time to failure, predictors of failure, and management strategies. Mean follow-up after implant was 4.15 ± 3.6 years (median = 3.63), with 37% of leads followed for >5 years. A total of 80 (1.5%) leads demonstrated electrical abnormalities requiring revision with an average time to failure of 4 ± 2.8 years (median = 3.5). Of the leads that malfunctioned, 62/80 (78%) were extracted and replaced with a new lead and in the other 18 cases, malfunctioned DF4 leads were abandoned, and a new lead implanted. In multivariable models, younger age at implant (OR 1.03 per year; P < 0.001) and the presence of Abbott/St. Jude leads increased the risk of malfunction.
CONCLUSION
DF4 defibrillator leads demonstrate excellent longevity with >98.3% of leads followed for at least 5 years still functioning normally. Younger age at implant and lead manufacturer are associated with an increased risk of DF4 lead malfunction. The differences in lead survival between manufacturers require further investigation.
Topics: Humans; Defibrillators, Implantable; Equipment Failure; Retrospective Studies
PubMed: 38000900
DOI: 10.1093/europace/euad347 -
Life (Basel, Switzerland) Mar 2024Homochirality is an obvious feature of life on Earth. On the other hand, extraterrestrial samples contain largely racemic compounds. The same is true for any common... (Review)
Review
Homochirality is an obvious feature of life on Earth. On the other hand, extraterrestrial samples contain largely racemic compounds. The same is true for any common organic synthesis. Therefore, it has been a perplexing puzzle for decades how these racemates could have formed enantiomerically enriched fractions as a basis for the origin of homochiral life forms. Numerous hypotheses have been put forward as to how preferentially homochiral molecules could have formed and accumulated on Earth. In this article, it is shown that homochirality of the abiotic organic pool at the time of formation of the first self-replicating molecules is not necessary and not even probable. It is proposed to abandon the notion of a molecular ensemble and to focus on the level of individual molecules. Although the formation of the first self-replicating, most likely homochiral molecule, is a seemingly improbable event, on a closer look, it is almost inevitable that some homochiral molecules have formed simply on a statistical basis. In this case, the non-selective leap to homochirality would be one of the first steps in chemical evolution directly out of a racemic "ocean". Moreover, most studies focus on the chirality of the primordial monomers with respect to an asymmetric carbon atom. However, any polymer with a minimal size that allows folding to a secondary structure would spontaneously lead to asymmetric higher structures (conformations). Most of the functions of these polymers would be influenced by this inherently asymmetric folding. Furthermore, a concept of physical compartmentalization based on rock nanopores in analogy to nanocavities of digital immunoassays is introduced to suggest that complex cell walls or membranes were also not required for the first steps of chemical evolution. To summarize, simple and universal mechanisms may have led to homochiral self-replicating systems in the context of chemical evolution. A homochiral monomer pool is deemed unnecessary and probably never existed on primordial Earth.
PubMed: 38541666
DOI: 10.3390/life14030341 -
World Journal For Pediatric &... Jul 2023Lead performance is suboptimal in young patients and a main cause of device system failure. Our objective was to assess early and midterm outcomes after epicardial...
BACKGROUND
Lead performance is suboptimal in young patients and a main cause of device system failure. Our objective was to assess early and midterm outcomes after epicardial device implantation in a contemporary pediatric cohort.
METHODS
A total of 116 consecutive pediatric patients underwent 137 epicardial device implantations from 2010 to 2019. Forty pacemakers and 97 implantable cardioverter defibrillators (ICDs) were implanted. Lead failure was defined as leads repaired, replaced, or abandoned due to fracture, dislodgement, or dysfunction. Freedom from device system failure was determined using Kaplan-Meier analysis.
RESULTS
Mean age at implantation was 10 ± 5 years, 46 (34%) were younger than 8 years old, 41 (30%) had prior cardiac surgery, and 38 (28%) had prior devices. Main indications were acquired heart block (17/40 [43%]), sinus node dysfunction (14/40 [35%]), and congenital heart block (7/40 [18%]) for pacemakers, and hypertrophic cardiomyopathy (46/97 [47%]), long QT syndrome (31/97 [32%]), and ventricular arrhythmia (17/97 [18%]) for ICDs. There were no early deaths. Three-year freedom from device system failure was 80% (95% CI 73%, 88%) for all patients and 88% (95% CI 79%, 99%) for patients <8 years old. Device system failure causes included lead fracture (20/34 [59%]), lead dysfunction (5/34 [15%]), lead dislodgement (5/34 [15%]), infection (3/34 [9%]), and pericarditis (1/34 [3%]). Reintervention was required in 26/34 (76%) device system failures.
CONCLUSIONS
Epicardial device implantation is safe, shows acceptable midterm outcomes in children, and is an effective option in patients younger than 8 years old. Close device surveillance continues to be essential to detect lead failure early and ensure timely reintervention.
Topics: Humans; Child; Arrhythmias, Cardiac; Defibrillators, Implantable; Cardiac Surgical Procedures
PubMed: 36851830
DOI: 10.1177/21501351231157374 -
Animals : An Open Access Journal From... Feb 2024The historical relationship between humans and dogs has involved selective breeding for various purposes, such as hunting, guarding, and service roles. However, over... (Review)
Review
The historical relationship between humans and dogs has involved selective breeding for various purposes, such as hunting, guarding, and service roles. However, over time, there has been a shift in preferences from functionality to aesthetics, which has influenced the diverse sizes, shapes, and coats of dog breeds. This review looks at fashionable dog breeding and questions the ethics of prioritising looks over health and behaviour. It aims to alert potential owners, breeders, and regulators to the importance of considering a dog's overall well-being, not just its appearance, which has resulted in fad breeding, leading to genetic disorders, health issues, and a loss of biodiversity. Ethical concerns arise from breeding brachycephalic breeds with respiratory conditions, inbreeding causing inherited disorders, and overbreeding popular breeds while shelter dogs remain unadopted. Additionally, the impact of cosmetic surgeries on popular dog breeds, as well as the neglect of behavioural traits in favour of physical characteristics and strict breeding practices are also considered. The current breeding model can have a negative impact on the emotional and cognitive well-being of dogs, resulting in issues such as aggression, anxiety, and other behavioural problems that can significantly reduce their overall quality of life. Unregulated breeding practices and the demand for rare breeds can lead to illegal breeding, compromising animal welfare. Prospective owners, veterinarians, kennel clubs, and legislators all need to play a responsible role in protecting animals.
PubMed: 38473141
DOI: 10.3390/ani14050756 -
Netherlands Heart Journal : Monthly... Sep 2023The introduction of dedicated tools for pacing and defibrillator lead extraction has resulted in relatively high success and low complication rates. The confidence this... (Review)
Review
The introduction of dedicated tools for pacing and defibrillator lead extraction has resulted in relatively high success and low complication rates. The confidence this elicits has broadened the indications from device infections to non-functional or redundant leads and the latter make up an increasing share of extraction procedures. Proponents of extracting these leads point to the higher complication burden of lead extraction in patients with longstanding abandoned leads when compared one-to-one with extraction when these leads become redundant. However, this does not translate into better patient outcomes on a population level: complications are rare with properly abandoned leads and thus most patients will never be subjected to an extraction procedure and the ensuing complications. Therefore, not extracting redundant leads minimises the risk for the patients and avoids many expensive procedures.
PubMed: 37010738
DOI: 10.1007/s12471-023-01770-7 -
Current Opinion in Pulmonary Medicine Nov 2023Biannual clock changes to and from daylight saving time have been pervasive in many societies for over 50 years. Governments are considering abandoning this practice... (Review)
Review
PURPOSE OF REVIEW
Biannual clock changes to and from daylight saving time have been pervasive in many societies for over 50 years. Governments are considering abandoning this practice and choosing a single permanent time.
RECENT FINDINGS
Our endogenous circadian clock follows our photoperiod, which changes over the year. The acute disruption caused by changing our clocks can affect safety (motor vehicle and on the job accidents), health (cardiovascular disease, drug overdoses, suicide), and human behavior (sport performance, generosity, and procrastination). Although abandoning the clock change could help avoid these acute harms, choosing the wrong permanent time could lead to chronic circadian misalignment, which could have even more profound implications for health, safety, and human behavior.
SUMMARY
Ceasing the biannual clock change may be a good choice, but governments need to be mindful of which permanent time to adopt. Many regions of the world already follow the wrong time during standard time, and circadian misalignment would be amplified by moving to permanent daylight saving time. In many regions, Standard Time better aligns with our circadian clock, thus providing a more natural light cycle that minimizes circadian misalignment.
Topics: Humans; Circadian Rhythm; Photoperiod; Circadian Clocks
PubMed: 37578383
DOI: 10.1097/MCP.0000000000001003 -
MRI in Patients with Cardiovascular Implantable Electronic Devices and Fractured or Abandoned Leads.Radiology. Cardiothoracic Imaging Jun 2024Purpose To examine the clinical effect of lead length and lead orientation in patients with cardiac implantable electronic devices (CIEDs) and lead fragments or...
Purpose To examine the clinical effect of lead length and lead orientation in patients with cardiac implantable electronic devices (CIEDs) and lead fragments or abandoned leads undergoing 1.5-T MRI. Materials and Methods This Health Insurance Portability and Accountability Act-compliant retrospective study included patients with CIEDs and abandoned leads or lead fragments undergoing 1.5-T MRI from March 2014 through July 2020. CIED settings before and after MRI were reviewed, with clinically significant variations defined as a composite of the change in capture threshold of at least 50%, in sensing of at least 40%, or in lead impedance of at least 30% between before MRI and after MRI interrogation. Adverse clinical events were assessed at MRI and up to 30 days after. Univariable and multivariable analysis was performed. Results Eighty patients with 126 abandoned CIED leads or lead fragments underwent 107 1.5-T MRI examinations. Sixty-seven patients (median age, 74 years; IQR, 66-78 years; 44 male patients, 23 female patients) had abandoned leads, and 13 (median age, 66 years; IQR, 52-74 years; nine male patients, four female patients) had lead fragments. There were no reported deaths, clinically significant arrhythmias, or adverse clinical events within 30 days of MRI. Three patients with abandoned leads had a significant change in the composite of capture threshold, sensing, or lead impedance. In a multivariable generalized estimating equation analysis, lead orientation, lead length, MRI type, and MRI duration were not associated with a significant change in the composite outcome. Conclusion Use of 1.5-T MRI in patients with abandoned CIED leads or lead fragments of varying length and orientation was not associated with adverse clinical events. Cardiac Assist Devices, MRI, Cardiac Implantable Electronic Device © RSNA, 2024.
Topics: Humans; Male; Female; Aged; Defibrillators, Implantable; Retrospective Studies; Pacemaker, Artificial; Magnetic Resonance Imaging; Middle Aged; Equipment Failure
PubMed: 38869431
DOI: 10.1148/ryct.230303 -
Journal of Clinical Medicine Dec 2023(1) Background: Transvenous lead extraction (TLE) can become far more complex when unanticipated difficulties arise. The aim was to develop a simple scoring system that...
(1) Background: Transvenous lead extraction (TLE) can become far more complex when unanticipated difficulties arise. The aim was to develop a simple scoring system that allows for the prediction of the difficulty and complexity of this significant procedure. (2) Methods: Based on analysis of 3741 TLE procedures with and without complicating factors (extended fluoroscopy time, need for second-line instruments, and advanced techniques and instruments), a five-point Complex Indicator of Difficulty of (TLE) Procedure (CID-TLEP) scale was developed. Two or more points on the CID-TLEP scale indicate a higher level of procedure complexity. (3) Results: Patient age below 51 years at first CIED implantation, number of abandoned leads, number of previous procedures, passive fixation and multiple leads to be extracted, and a ratio of dwell time of oldest lead to patient age during TLE of >0.13 are significant predictors of higher levels of lead extraction complexity. The ROC analysis demonstrates that a point total (being the sum of the odds ratios of the above variables) of >9.697 indicates a 21.83% higher probability of complex TLE (sensitivity 74.08%, specificity 74.46%). Finally, a logistic function was calculated, and we constructed a simple equation for lead extraction complexity that can predict the probability of a difficult procedure. The risk of complex extraction (as a percentage) is calculated as [1/(1 + 55.34 · 0.754)] · 100 ( < 0.001). (4) Conclusion: The LECOM score can effectively predict the risk of a difficult transvenous lead extraction procedure, and predicting the probability of a more complex procedure may help clinicians in planning lead removal and improving patient management.
PubMed: 38137637
DOI: 10.3390/jcm12247568