Cardiology in the Young Sep 2023The aim of this study is to establish a guidance for device retrieval based on comprehensive bench tests.
OBJECTIVES
The aim of this study is to establish a guidance for device retrieval based on comprehensive bench tests.
BACKGROUND
Device embolisation remains a major complication in transcatheter closure of atrial septal defect and patent ductus arteriosus. Although percutaneous retrieval is feasible in the majority of cases, surgical retrieval may be required in complicated circumstances. However, the methods of transcatheter device retrieval have not been completely established.
METHODS
Bench tests of device retrieval were performed to verify the appropriate retrieval method according to device type/size. The devices used for testing were Amplatzer Septal Occluder (Abbott, Chicago, IL, United States of America), Figulla Flex II (Occlutech GmbH, Jena, Germany), Amplatzer Duct Occluder-I (Abbott), Amplatzer Duct Occluder-II (Abbott), and Amplatzer Vascular Plug-II (Abbott). The retrieval equipment constituted diagnostic catheters (multipurpose catheter and right Judkins catheter, 4-Fr or 5-Fr, Gadelius Medical, Tokyo, Japan), delivery sheath and cables for each device, Amplatz goose neck snares (Medtronic, Minneapolis, MN, United States of America), OSYPKA CATCHER (Osypka ag, Rheinfelden-Herten, Germany), and OSYPKA LASSOS (Osypka). We investigated the retrieval equipment and sheath sizes required for a successful retrieval procedure for variously sized devices.
RESULTS
For patent ductus arteriosus devices, the type of snare and the snaring position are considered important. For atrial septal defect devices, simple snare capture or a double-snare technique with a sufficiently large sheath is effective. Special care should be taken when using the OSYPKA CATCHER for device retrieval.
CONCLUSIONS
The results of this study may assist in the selection of both capture devices and a retrieval sheath or a catheter for complete retrieval.
Topics: Humans; Ductus Arteriosus, Patent; Treatment Outcome; Cardiac Catheterization; Septal Occluder Device; Heart Septal Defects, Atrial
PubMed: 36093848
DOI: 10.1017/S1047951122002864 -
Interventional Cardiology Clinics Jul 2024Congenital heart disease (CHD) is the most common congenital birth defect with an incidence of 1 in 100. Current survival to adulthood is expected in 9 out of 10... (Review)
Review
Congenital heart disease (CHD) is the most common congenital birth defect with an incidence of 1 in 100. Current survival to adulthood is expected in 9 out of 10 children with severe CHD as the diagnostic, interventional, and surgical success improves. The adult CHD (ACHD) population is increasingly diverse, reflecting the broad spectrum of CHD and evolution of surgical techniques to improve survival. Similarly, transcatheter interventions have seen exponential growth and creativity to reduce the need for repeat sternotomies. This article focuses on newer data and evolving techniques for transcatheter interventions specific to certain ACHD populations.
Topics: Humans; Heart Defects, Congenital; Cardiac Catheterization; Adult; Cardiac Surgical Procedures
PubMed: 38839171
DOI: 10.1016/j.iccl.2024.03.004 -
Cardiovascular & Hematological... Apr 2024Cryptogenic stroke, whose underlying pathology is unknown, accounts for 30-40% of all ischemic strokes. Studies have mentioned the association between atrial septal...
BACKGROUND
Cryptogenic stroke, whose underlying pathology is unknown, accounts for 30-40% of all ischemic strokes. Studies have mentioned the association between atrial septal abnormalities and cryptogenic stroke, but there are still disparities in the results among different studies.
OBJECTIVE
We aimed to clarify the prevalence of atrial septal abnormalities in patients with cryptogenic stroke.
METHODS
We conducted a cross-sectional study on 91 patients with cryptogenic stroke/transient ischemic attack from March 2021 to March 2022. We evaluated the demographic data of the patients and also the existence of neurologic attacks. Furthermore, echocardiography was performed to determine the type of atrial septal abnormality.
RESULTS
Out of 91 patients with cryptogenic stroke/transient ischemic attack, 16 patients (17.5%) had patent foramen ovale, 1 man (1.1%) had atrial septal aneurysm, and 1 woman (1.1%) had an atrial septal defect. Patients with patent foramen ovale were significantly younger than those without. The size of patent foramen ovale in patients with cryptogenic stroke was larger than those with transient ischemic attack, but this difference was not significant. Also, the size of the patent foramen ovale (length and width) was not significantly related to any of the demographic variables (p-value = 0.544, 0.604).
CONCLUSION
Based on our results, the prevalence of atrial septal abnormalities was relatively high. Considering these issues and the importance of preventing neurological accidents in patients, especially young people, it is recommended to always consider atrial septal disorders and, if diagnosed, to carry out the necessary treatment in this field.
PubMed: 38685781
DOI: 10.2174/011871529X294809240415070950 -
Journal of Cardiology Jun 2024The left atrium approach for atrial fibrillation (AF) ablation requires an atrial transseptal puncture that may cause an iatrogenic atrial septal defect (iASD). This...
BACKGROUND
The left atrium approach for atrial fibrillation (AF) ablation requires an atrial transseptal puncture that may cause an iatrogenic atrial septal defect (iASD). This study aimed to investigate the incidence and predictors of iASD in catheter ablation, assessed by transthoracic echocardiography (TTE), a relatively non-invasive technique frequently employed in follow-up.
METHODS
This retrospective study included 639 patients (489 male; 60.2 ± 10.7 years) who underwent initial catheter ablation for AF between May 2005 and June 2018. All patients underwent preprocedural transesophageal echocardiography (pre-TEE), preprocedural TTE (pre-TTE), and TTE one day after the procedure (post-TTE). iASD incidence after 6 months (6 M), preprocedural characteristics, and procedure methods were evaluated.
RESULTS
Patent foramen ovale (PFO) was diagnosed in 42 patients (6.6 %) using pre-TEE and in 11 patients using pre-TTE (26.2 % of the patients with PFO in pre-TEE). Among the 597 patients without PFO, 497 underwent 6 M-TTE. iASD was observed in 59.6 % of patients using post-TTE and 4.6 % using 6 M-TTE. In the univariate logistic regression analysis, the total diameter of the sheath through the septum (odds ratio 1.15, p < 0.001) or two sheaths through a single puncture (odds ratio 4.17, p = 0.001) were independent risk factors on iASD incidence in 6 M-TTE. iASD was also more likely to occur via cryoballoon ablation using a larger sheath than radiofrequency catheter ablation.
CONCLUSIONS
iASD was not a rare complication. A larger sheath diameter or two sheaths through a single puncture were associated with the incidence of iASD.
PubMed: 38914280
DOI: 10.1016/j.jjcc.2024.06.007 -
EClinicalMedicine Sep 2023Atrial septal defect (ASD) increases the risk of adverse cardiovascular outcomes. Despite the potential for risk mitigation through minimally invasive percutaneous...
BACKGROUND
Atrial septal defect (ASD) increases the risk of adverse cardiovascular outcomes. Despite the potential for risk mitigation through minimally invasive percutaneous closure, ASD remains underdiagnosed due to subtle symptoms and examination findings. To bridge this diagnostic gap, we propose a novel screening strategy aimed at early detection and enhanced diagnosis through the implementation of a convolutional neural network (CNN) to identify ASD from 12-lead electrocardiography (ECG).
METHODS
ECGs were collected from patients with at least one recorded echocardiogram at 3 hospitals from 2 continents (Keio University Hospital from July 2011 to December 2020, Brigham and Women's Hospital from January 2015 to December 2020, and Dokkyo Medical University Saitama Medical Center from January 2010 and December 2021). ECGs from patients with a diagnosis of ASD were labeled as positive cases while the remainder were labeled as negative. ECGs after the closure of ASD were excluded. After randomly splitting the ECGs into 3 datasets (50% derivation, 20% validation, and 30% test) with no patient overlap, a CNN-based model was trained using the derivation datasets from 2 hospitals and was tested on held-out datasets along with an external validation on the 3rd hospital. All eligible ECGs were used for derivation and validation whereas the earliest ECG for each patient was used for the test and external validation. The discrimination of ASD was assessed by the area under the receiver operating characteristic curve (AUROC). Multiple subgroups were examined to identify any heterogeneity.
FINDINGS
A total of 671,201 ECGs from 80,947 patients were collected from the 3 institutions. The AUROC for detecting ASD was 0.85-0.90 across the 3 hospitals. The subgroup analysis showed excellent performance across various characteristics Screening simulation using the model greatly increased sensitivity from 80.6% to 93.7% at specificity 33.6% when compared to using overt ECG abnormalities.
INTERPRETATION
A CNN-based model using 12-lead ECG successfully identified the presence of ASD with excellent generalizability across institutions from 2 separate continents.
FUNDING
This work was supported by research grants from JST (JPMJPF2101), JSR corporation, Taiju Life Social Welfare Foundation, Kondou Kinen Medical Foundation, Research fund of Mitsukoshi health and welfare foundation, Tokai University School of Medicine Project Research and Internal Medicine Project Research, Secom Science and Technology Foundation, and Grants from AMED (JP23hma922012 and JP23ym0126813). This work was partially supported by One Brave Idea, co-funded by the American Heart Association and Verily with significant support from AstraZeneca and pillar support from Quest Diagnostics.
PubMed: 37753448
DOI: 10.1016/j.eclinm.2023.102141 -
Journal of Clinical Medicine Jan 2024Transcatheter atrial septal closures for secundum atrial septal defects (ASD) have demonstrated favorable clinical outcomes. However, the impact of device implantation...
BACKGROUND
Transcatheter atrial septal closures for secundum atrial septal defects (ASD) have demonstrated favorable clinical outcomes. However, the impact of device implantation on the stiffness of the left atrium remains unclear.
METHOD
Patients with secundum ASD undergoing transcatheter closure and follow-up right heart catheterization at six months were included. We investigated the relationship between post-procedural (E/e' ratio)/(LAs strain) ratio, an index of left atrial stiffness, and baseline characteristics, including echocardiographic and hemodynamic parameters.
RESULTS
Forty patients were included (median 69 (56, 75) years, 12 men, and pulmonary systemic flow ratio 2.27 (1.96, 2.86)). Trans-catheter ASD closure was successfully performed without any major complications, accompanying a significant reduction in right ventricular to left ventricular size ratio from 1.04 (0.87, 1.13) to 0.74 (0.66, 0.86) ( < 0.01). The (E/e' ratio)/(LAs strain) ratio was markedly elevated the day after the procedure and was further increased 6 months later (before: 0.25 (0.17, 0.34), 1 day later: 0.34 (0.27, 0.50), 6 months later: 0.43 (0.27, 0.76), < 0.01). The groups with higher (E/e' ratio)/(LAs strain) ratios at 6 months had significantly more severe heart failure conditions including lower cardiac output and higher plasma B-type natriuretic peptides.
CONCLUSIONS
Patients undergoing transcatheter ASD closure experienced improvement in hemodynamics and clinical symptoms but an elevation in left atrial stiffness post-procedure. The clinical ramifications of this finding, particularly during the longer-term observation period subsequent to ASD closure, warrant further investigation.
PubMed: 38256461
DOI: 10.3390/jcm13020327 -
International Journal of Cardiology Jan 2024Worsening mitral regurgitation (MR) is a complication of intervention for atrial septal defect (ASD). Little is known about mitral valve (MV) characteristics associated...
BACKGROUND
Worsening mitral regurgitation (MR) is a complication of intervention for atrial septal defect (ASD). Little is known about mitral valve (MV) characteristics associated with worsening MR. We aimed to elucidate MR outcomes and predictors of worsening MR after transcatheter ASD closure.
METHODS
We analyzed changes in MR from prior to transcatheter ASD closure to 6 months after the procedure and predictors of worsening MR via baseline transthoracic echocardiography in 238 patients (64.7% females; mean age, 53 ± 22 years).
RESULTS
Worsening MR was defined as worsening to moderate in patients with less than or equal to mild MR at baseline or vena contracta width increasing of ≥2 mm by 6-month follow-up in patients with moderate MR. Worsening MR was observed in 29 patients (12.2%). The associated echocardiographic findings were pseudoprolapse, hamstringing, stiffness, and anteroposterior and intercommissural mitral annulus diameter in the univariable logistic regression analysis (all P < 0.05). Multivariable analysis after adjusting for age; long-standing persistent atrial fibrillation; and ASD size showed that models combining MV leaflet findings such as pseudoprolapse or hamstringing, or anterior leaflet stiffness with the ratio of the sum of anterior and posterior leaflet lengths to intercommissural mitral annulus diameter were statistically significant for predicting worsening MR (R = 0.393, P < 0.001 and R = 0.385, P < 0.001, respectively).
CONCLUSIONS
Worsening MR after transcatheter ASD closure might depend on MV leaflet findings and annulus size in patients with long-standing persistent atrial fibrillation.
Topics: Female; Humans; Adult; Middle Aged; Aged; Male; Mitral Valve Insufficiency; Atrial Fibrillation; Retrospective Studies; Mitral Valve; Heart Septal Defects, Atrial
PubMed: 37777073
DOI: 10.1016/j.ijcard.2023.131404 -
HGG Advances Jul 2023Contraction of the human sarcomere is the result of interactions between myosin cross-bridges and actin filaments. Pathogenic variants in genes such as , , and that...
Contraction of the human sarcomere is the result of interactions between myosin cross-bridges and actin filaments. Pathogenic variants in genes such as , , and that encode parts of the cardiac sarcomere cause muscle diseases that affect the heart, such as dilated cardiomyopathy and hypertrophic cardiomyopathy. In contrast, pathogenic variants in homologous genes such as , , and that encode parts of the skeletal muscle sarcomere cause muscle diseases affecting skeletal muscle, such as distal arthrogryposis (DA) syndromes and skeletal myopathies. To date, there have been few reports of genes (e.g., ) encoding sarcomeric proteins in which the same pathogenic variant affects skeletal and cardiac muscle. Moreover, none of the known genes underlying DA have been found to contain pathogenic variants that also cause cardiac abnormalities. We report five families with DA because of heterozygous missense variants in the gene , , (). encodes a highly conserved actin that binds to myosin in cardiac and skeletal muscle. Pathogenic variants in have been found previously to underlie atrial septal defect, dilated cardiomyopathy, hypertrophic cardiomyopathy, and left ventricular noncompaction. Our discovery delineates a new DA condition because of variants in and suggests that some functions of are shared in cardiac and skeletal muscle.
Topics: Humans; Arthrogryposis; Actins; Heart Defects, Congenital; Cardiomyopathies; Cardiomyopathy, Dilated; Muscular Diseases; Myosins; Cardiomyopathy, Hypertrophic
PubMed: 37457373
DOI: 10.1016/j.xhgg.2023.100213 -
European Heart Journal Nov 2023
Topics: Humans; Heart Septal Defects, Atrial; Migraine Disorders; Prostheses and Implants; Cardiac Catheterization; Treatment Outcome; Septal Occluder Device
PubMed: 37793152
DOI: 10.1093/eurheartj/ehad645