Ultrasound in Obstetrics & Gynecology :... Jul 2020Incompatibility between currently available fetoscopes and the anatomical constraints of the distended fetal bladder, with the resulting curvature around the bladder...
OBJECTIVES
Incompatibility between currently available fetoscopes and the anatomical constraints of the distended fetal bladder, with the resulting curvature around the bladder neck, account for most technical difficulties during fetal cystoscopy in lower urinary tract obstruction (LUTO). The aim of this anatomical study was to assess by magnetic resonance imaging (MRI) the variation in three bladder angles (bladder-neck angle (BNA), vesicourethral angle (VUA) and angle between bladder dome and posterior urethra (DUA)), according to gestational age (GA), bladder volume and the presence of LUTO.
METHODS
From our fetal medicine database, we retrieved for review 46 MRI examinations of male fetuses between 2015 and 2019, including 17 with LUTO, examined at a mean GA of 28.1 (range, 17.3-35.0) weeks and 29 age-matched controls, examined at 29.9 (range, 21.9-35.0) weeks. We measured bladder volume, bladder-wall thickness and the three bladder angles, and used the Mann-Whitney U-test to compare values between groups. Variations according to GA and bladder volume were determined using analysis of variance (ANOVA). A reliability study was performed using the Bland-Altman method and Lin's correlation coefficient was calculated.
RESULTS
Both bladder volume and bladder-wall thickness were significantly greater in the LUTO group (P < 0.01). BNA was significantly larger in LUTO compared with control fetuses: the mean (range) was 127.1° (101.6-161.6°) vs 111.2° (88.5-157.3°) (P < 0.01). DUA averaged 117° and showed no difference between the groups (P = 0.92). No statistical comparison was performed on VUA since this was not measurable in most control fetuses. ANOVA showed no variation of any angle with bladder volume in both LUTO fetuses and control fetuses. BNA in LUTO fetuses was the only angle to vary with GA, being larger after, compared with at or before, 25 weeks (P = 0.04). The reliability study showed an acceptable bias for both intra- and interobserver reproducibility for all three angles.
CONCLUSION
The findings that BNA is increased by approximately 15° in fetuses with LUTO and DUA averages 117° could aid in development of a customized fetal cystoscope and help to overcome the current technical challenges of fetal cystoscopy. Copyright © 2019 ISUOG. Published by John Wiley & Sons Ltd.
Topics: Adult; Biometry; Case-Control Studies; Cystoscopy; Female; Gestational Age; Humans; Magnetic Resonance Imaging; Male; Pregnancy; Prenatal Diagnosis; Retrospective Studies; Ultrasonography, Prenatal; Urinary Bladder Neck Obstruction
PubMed: 31006924
DOI: 10.1002/uog.20297 -
American Journal of Obstetrics &... May 2022Preterm prelabor rupture of membranes is the most frequent complication of fetoscopic surgery. Strategies to seal the membrane defect created by fetoscopy have been...
BACKGROUND
Preterm prelabor rupture of membranes is the most frequent complication of fetoscopic surgery. Strategies to seal the membrane defect created by fetoscopy have been attempted with little success. We previously developed an integrated semirigid bioadhesive patch composed of silicone and hydroxypropyl methylcellulose that achieved ex vivo sealing of membrane defects.
OBJECTIVE
To evaluate the feasibility of the insertion of our integrated semirigid bioadhesive patches using a fetoscopic technique and to test the adhesion in ex vivo human membranes and in an in vivo ovine model.
STUDY DESIGN
An experimental study involving 2 experiments: (1) ex vivo-human fetal membranes were mounted in a custom-designed model with saline solution simulating intraamniotic pressure. The insertion of 2 different bioadhesive patches made of silicone-hydroxypropyl methylcellulose and silicone-polyurethane-hydroxypropyl methylcellulose was performed through a 12-Fr cannula mimicking fetoscopic surgery technique. The experiment was repeated 10 times with membranes from different donors. Measures included insertion time, successful insertion, and adhesion at 5 minutes; (2) in vivo-16 patches of silicone-hydroxypropyl methylcellulose were inserted by fetoscopy in the amniotic cavity of pregnant sheep (4 bioadhesives per animal, in 4 ewes). Measures included successful insertion, adhesion at 5 minutes, and adhesion at the end of surgery.
RESULTS
In the ex vivo insertion study, there was no difference in the insertion time between silicone-hydroxypropyl methylcellulose and silicone-polyurethane-hydroxypropyl methylcellulose patches (P=.49). Insertion was successful in all cases, but complete adhesion at 5 minutes was superior for silicone-hydroxypropyl methylcellulose (P=.02). In the in vivo study, insertion of silicone-hydroxypropyl methylcellulose by fetoscopy was feasible and successful in all cases, and no complications were reported. Adhesion persisted at 5 minutes and at the end of the surgery in 68.8% and 56.3% of the patches, respectively.
CONCLUSION
We describe the feasibility of deploying through a fetoscopic trocar a semirigid silicone-hydroxypropyl methylcellulose patch that seals fetal membranes after an invasive fetal procedure. The results warrant further research for improving long-term adhesion and developing a clinically applicable system.
Topics: Animals; Female; Fetoscopes; Fetoscopy; Hypromellose Derivatives; Polyurethanes; Pregnancy; Sheep; Silicones
PubMed: 35144009
DOI: 10.1016/j.ajogmf.2022.100593 -
Medical Devices (Auckland, N.Z.) 2020The annual global neonatal mortality stands at 2.5 million deaths, 1 million of them dying within the first day of life. An additional 2.6 million are stillborn...
BACKGROUND
The annual global neonatal mortality stands at 2.5 million deaths, 1 million of them dying within the first day of life. An additional 2.6 million are stillborn globally, the majority of them due to intrapartum events. Optimal fetal heart rate (FHR) monitoring has the potential to timely detect fetuses at risk and, if coupled with timely obstetric responses may save more newborns. Moyo is a new Doppler with nine crystals capable of monitoring FHR both intermittently and continuously.
AIM
To assess women's and midwives' opinions on the use of Moyo for intrapartum FHR monitoring.
METHODS
We conducted a cross-sectional study using a structured questionnaire to assess women's and midwives' perception. Women who gave birth at the hospital who used Moyo were interviewed using a questionnaire immediately before discharge from the hospital. Twenty-eight midwives who have been using Moyo for more than 6 months were also interviewed using a structured questionnaire. Data were analyzed using excel and result presented in figures.
RESULTS
In total 113 postpartum women who were monitored using Moyo were interviewed before discharge. Out of these, 46 (40.7%) were first-time mothers and the rest were multipara. In total, 95 women (84.1%) used Moyo and other devices for FHR monitoring, 81 (72%) said Moyo was better than Fetoscopes and handheld Doppler, two-third 75 (66.4%) felt that Moyo was comfortable and 93 (82.3%) would like Moyo to be used on them in the future. Out of 28 midwives, 11 (39.3%) used Moyo continuous only, 3 (10.7%) used Moyo intermittently only and 14 (50.0%) used both intermittent and continuous. Thirteen (46.4%) midwives prefer to use Moyo both intermittent and continuous. Sixteen (55.6%) said Moyo was effective, 21 (75%) felt comfortable to use Moyo, and 13 (46.4%) said Moyo was easy to use.
CONCLUSION
The majority of midwives and women who used Moyo felt that Moyo was comfortable for intrapartum FHR monitoring. Moyo can be used both intermittently and continuously depending on the user's preferences.
PubMed: 32256129
DOI: 10.2147/MDER.S241741 -
Artificial Intelligence in Medicine Jan 2024Fetoscopic Laser Coagulation (FLC) for Twin to Twin Transfusion Syndrome is a challenging intervention due to the working conditions: low quality images acquired from a...
Fetoscopic Laser Coagulation (FLC) for Twin to Twin Transfusion Syndrome is a challenging intervention due to the working conditions: low quality images acquired from a 3 mm fetoscope inside a turbid liquid environment, local view of the placental surface, unstable surgical field and delicate tissue layers. FLC is based on locating, coagulating and reviewing anastomoses over the placenta's surface. The procedure demands the surgeons to generate a mental map of the placenta with the distribution of the anastomoses, maintaining, at the same time, precision in coagulation and protecting the placenta and amniotic sac from potential damages. This paper describes a teleoperated platform with a cognitive-based control that provides assistance to improve patient safety and surgery performance during fetoscope navigation, target re-location and coagulation processes. A comparative study between manual and teleoperated operation, executed in dry laboratory conditions, analyzes basic fetoscopic skills: fetoscope navigation and laser coagulation. Two exercises are proposed: first, fetoscope guidance and precise coagulation. Second, a resolved placenta (all anastomoses are indicated) to evaluate navigation, re-location and coagulation. The results are analyzed in terms of economy of movement, execution time, coagulation accuracy, amount of coagulated placental surface and risk of placenta puncture. In addition, new metrics, based on navigation and coagulation maps evaluate robotic performance. The results validate the developed platform, showing noticeable improvements in all the metrics.
Topics: Female; Pregnancy; Humans; Laser Coagulation; Fetoscopes; Robotics; Placenta; Exercise
PubMed: 38184348
DOI: 10.1016/j.artmed.2023.102725 -
Journal of Perinatal Medicine Jun 2024Fetoscopic laser coagulation of placental anastomoses is usually performed for a treatment of twin-to-twin transfusion syndrome (TTTS). A common complication of... (Comparative Study)
Comparative Study
Fetoscopic laser coagulation for twin-to-twin transfusion syndrome: a comparison of flexible 1.0/1.2 mm fetoscopes with curved sheaths of 2.7/3.3 mm vs. 2 mm fetoscopic lens technique with sheaths of 6.6/11.3 mm.
OBJECTIVES
Fetoscopic laser coagulation of placental anastomoses is usually performed for a treatment of twin-to-twin transfusion syndrome (TTTS). A common complication of fetoscopic laser coagulation for TTTS is preterm preliminary rupture of fetal membranes (PPROM) aggravating the neonatal outcome significantly. However, use of an flexible 1 mm fetoscope with an curved sheath could reduce iatrogenic damage of the amniotic membrane and improve neonatal outcomes after laser treatment. The aim of this study was to compare neonatal outcomes using this flexible fetoscope with curved sheath vs. use of a standard lens technique.
METHODS
Outcomes were retrospective analyzed after use of a standard lens fetoscope of 2 mm (sheath 6.63 mm or 11.27 mm for anterior placenta) and a flexible fetoscope of 1 mm or 1.2 mm (sheath 2.65 mm or 3.34 mm) in two German centers of fetal surgery, performed during 2006-2019.
RESULTS
Neonatal outcome of 247 TTTS patients were analyzed including the rates of double and single fetal survival. The survival of at least one fetus was 97.2 % in the group with the ultrathin technique (n=154) compared to 88.3 % (n=93) in the group with the standard lens fetoscope (p=0.008). Survival of both fetuses was not different between groups (81.0 vs. 75.3 %). The procedure to delivery interval was significantly increased using the ultrathin fetoscope (89.1±35.0 d vs. 71.4±35.4 d, p=0.001) resulting in an increased gestational age at delivery by 11 days on average (231.9±28.1 d vs. 221.1±32.7 d, p=0.012).
CONCLUSIONS
Fetal survival can be significantly increased following TTTS using flexible fetoscope of 1 mm or 1.2 mm (sheath 2.65 mm or 3.34 mm).
Topics: Humans; Fetofetal Transfusion; Pregnancy; Female; Fetoscopy; Retrospective Studies; Laser Coagulation; Fetoscopes; Adult; Infant, Newborn; Fetal Membranes, Premature Rupture
PubMed: 38536953
DOI: 10.1515/jpm-2023-0328