Journal of Arrhythmia Jun 2024Guidelines recommended remote monitoring (RM) in managing patients with Cardiac Implantable Electronic Devices. In recent years, smart device (phone or tablet)...
BACKGROUND
Guidelines recommended remote monitoring (RM) in managing patients with Cardiac Implantable Electronic Devices. In recent years, smart device (phone or tablet) monitoring-based RM (SM-RM) was introduced. This study aims to systematically review SM-RM versus bedside monitor RM (BM-RM) using radiofrequency in terms of compliance, connectivity, and episode transmission time.
METHODS
We conducted a systematic review, searching three international databases from inception until July 2023 for studies comparing SM-RM (intervention group) versus BM-RM (control group).
RESULTS
Two matched studies (21 978 patients) were retrieved (SM-RM arm: 9642 patients, BM-RM arm: 12 336 patients). There is significantly higher compliance among SM-RM patients compared with BM-RM patients in both pacemaker and defibrillator patients. Manyam et al. found that more SM-RM patients than BM-RM patients transmitted at least once (98.1% vs. 94.3%, < .001), and Tarakji et al. showed that SM-RM patients have higher success rates of scheduled transmissions than traditional BM-RM methods (SM-RM: 94.6%, pacemaker manual: 56.3%, pacemaker wireless: 77.0%, defibrillator wireless: 87.1%). There were higher enrolment rates, completed scheduled and patient-initiated transmissions, shorter episode transmission time, and higher connectivity among SM-RM patients compared to BM-RM patients. Younger patients (aged <75) had more patient-initiated transmissions, and a higher proportion had ≥10 transmissions compared with older patients (aged ≥75) in both SM-RM and BM-RM groups.
CONCLUSION
SM-RM is a step in the right direction, with good compliance, connectivity, and shorter episode transmission time, empowering patients to be in control of their health. Further research on cost-effectiveness and long-term clinical outcomes can be carried out.
PubMed: 38939794
DOI: 10.1002/joa3.13054 -
PloS One 2024The elderly is commonly susceptible to depression, the symptoms for which may overlap with natural aging or other illnesses, and therefore miss being captured by routine...
BACKGROUND
The elderly is commonly susceptible to depression, the symptoms for which may overlap with natural aging or other illnesses, and therefore miss being captured by routine screening questionnaires. Passive sensing data have been promoted as a tool for depressive symptoms detection though there is still limited evidence on its usage in the elderly. Therefore, this study aims to review current knowledge on the use of passive sensing data via smartphones and smartwatches in depressive symptom screening for the elderly.
METHOD
The search of literature was performed in PubMed, IEEE Xplore digital library, and PsycINFO. Literature investigating the use of passive sensing data to screen, monitor, and/or predict depressive symptoms in the elderly (aged 60 and above) via smartphones and/or wrist-worn wearables was included for initial screening. Studies in English from international journals published between January 2012 to September 2022 were included. The reviewed studies were further analyzed by a narrative analysis.
RESULTS
The majority of 21 included studies were conducted in Western countries with a few in Asia and Australia. Most studies adopted a cohort study design (n = 12), followed by cross-sectional design (n = 7) and a case-control design (n = 2). The most popular passive sensing data was related to sleep and physical activity using an actigraphy. Sleep characteristics, such as prolonged wakefulness after sleep onset, along with lower levels of physical activity, exhibited a significant association with depression. However, cohort studies expressed concerns regarding data quality stemming from incomplete follow-up and potential confounding effects.
CONCLUSION
Passive sensing data, such as sleep, and physical activity parameters should be promoted for depressive symptoms detection. However, the validity, reliability, feasibility, and privacy concerns still need further exploration.
Topics: Humans; Smartphone; Depression; Aged; Mass Screening; Wearable Electronic Devices; Sleep; Middle Aged; Exercise; Female
PubMed: 38935797
DOI: 10.1371/journal.pone.0304845 -
Medicina (Kaunas, Lithuania) May 2024Muscle properties are critical for performance and injury risk, with changes occurring due to physical exertion, aging, and neurological conditions. The MyotonPro... (Review)
Review
Muscle properties are critical for performance and injury risk, with changes occurring due to physical exertion, aging, and neurological conditions. The MyotonPro device offers a non-invasive method to comprehensively assess muscle biomechanical properties. This systematic review evaluates the reliability of MyotonPro across various muscles for diagnostic purposes. Following PRISMA guidelines, a comprehensive literature search was conducted in Medline (PubMed), Ovid (Med), Epistemonikos, Embase, Cochrane Library, Clinical trials.gov, and the WHO International Clinical Trials platform. Studies assessing the reliability of MyotonPro across different muscles were included. A methodological quality assessment was performed using established tools, and reviewers independently conducted data extraction. Statistical analysis involved summarizing intra-rater and inter-rater reliability measures across muscles. A total of 48 studies assessing 31 muscles were included in the systematic review. The intra-rater and inter-rater reliability were consistently high for parameters such as frequency and stiffness in muscles of the lower and upper extremities, as well as other muscle groups. Despite methodological heterogeneity and limited data on specific parameters, MyotonPro demonstrated promising reliability for diagnostic purposes across diverse patient populations. The findings suggest the potential of MyotonPro in clinical assessments for accurate diagnosis, treatment planning, and monitoring of muscle properties. Further research is needed to address limitations and enhance the applicability of MyotonPro in clinical practice. Reliable muscle assessments are crucial for optimizing treatment outcomes and improving patient care in various healthcare settings.
Topics: Humans; Reproducibility of Results; Muscle, Skeletal; Diagnostic Tests, Routine
PubMed: 38929468
DOI: 10.3390/medicina60060851 -
Journal of the International AIDS... Jun 2024Evidence-based intervention strategies to improve adherence among individuals living with chronic conditions are critical in ensuring better outcomes. In this systematic...
INTRODUCTION
Evidence-based intervention strategies to improve adherence among individuals living with chronic conditions are critical in ensuring better outcomes. In this systematic review, we assessed the impact of interventions that aimed to promote adherence to treatment for chronic conditions.
METHODS
We systematically searched PubMed, Web of Science, Scopus, Google Scholar and CINAHL databases to identify relevant studies published between the years 2000 and 2023 and used the QUIPS assessment tool to assess the quality and risk of bias of each study. We extracted data from eligible studies for study characteristics and description of interventions for the study populations of interest.
RESULTS
Of the 32,698 total studies/records screened, 2814 were eligible for abstract screening and of those, 497 were eligible for full-text screening. A total of 82 studies were subsequently included, describing a total of 58,043 patients. Of the total included studies, 58 (70.7%) were related to antiretroviral therapy for HIV, 6 (7.3%) were anti-hypertensive medication-related, 12 (14.6%) were anti-diabetic medication-related and 6 (7.3%) focused on medication for more than one condition. A total of 54/82 (65.9%) reported improved adherence based on the described study outcomes, 13/82 (15.9%) did not have clear results or defined outcomes, while 15/82 (18.3%) reported no significant difference between studied groups. The 82 publications described 98 unique interventions (some studies described more than one intervention). Among these intervention strategies, 13 (13.3%) were multifaceted (4/13 [30.8%] multi-component health services- and community-based programmes, 6/13 [46.2%] included individual plus group counselling and 3/13 [23.1%] included SMS or alarm reminders plus individual counselling).
DISCUSSION
The interventions described in this review ranged from adherence counselling to more complex interventions such as mobile health (mhealth) interventions. Combined interventions comprised of different components may be more effective than using a single component in isolation. However, the complexity involved in designing and implementing combined interventions often complicates the practicalities of such interventions.
CONCLUSIONS
There is substantial evidence that community- and home-based interventions, digital health interventions and adherence counselling interventions can improve adherence to medication for chronic conditions. Future research should answer if existing interventions can be used to develop less complicated multifaceted adherence intervention strategies.
Topics: Humans; Africa South of the Sahara; Chronic Disease; HIV Infections; Medication Adherence
PubMed: 38924296
DOI: 10.1002/jia2.26266 -
Resuscitation Plus Sep 2024To examine speed and accuracy of newborn heart rate measurement by various assessment methods employed at birth. (Review)
Review
AIM
To examine speed and accuracy of newborn heart rate measurement by various assessment methods employed at birth.
METHODS
A search of Medline, SCOPUS, CINAHL and Cochrane was conducted between January 1, 1946, to until August 16, 2023. (CRD 42021283364) Study selection was based on predetermined criteria. Reviewers independently extracted data, appraised risk of bias and assessed certainty of evidence.
RESULTS
Pulse oximetry is slower and less precise than ECG for heart rate assessment. Both auscultation and palpation are imprecise for heart rate assessment. Other devices such as digital stethoscope, Doppler ultrasound, an ECG device using dry electrodes incorporated in a belt, photoplethysmography and electromyography are studied in small numbers of newborns and data are not available for extremely preterm or bradycardic newborns receiving resuscitation. Digital stethoscope is fast and accurate. Doppler ultrasound and dry electrode ECG in a belt are fast, accurate and precise when compared to conventional ECG with gel adhesive electrodes.
LIMITATIONS
Certainty of evidence was low or very low for most comparisons.
CONCLUSION
If resources permit, ECG should be used for fast and accurate heart rate assessment at birth. Pulse oximetry and auscultation may be reasonable alternatives but have limitations. Digital stethoscope, doppler ultrasound and dry electrode ECG show promise but need further study.
PubMed: 38912532
DOI: 10.1016/j.resplu.2024.100668 -
Journal of the American Medical... Jun 2024To examine randomized controlled trials (RCTs) of "hospital at home" (HAH) for admission avoidance in adults presenting with acute physical illness to identify the use... (Review)
Review
OBJECTIVES
To examine randomized controlled trials (RCTs) of "hospital at home" (HAH) for admission avoidance in adults presenting with acute physical illness to identify the use of vital sign monitoring approaches and evidence for their effectiveness.
DESIGN
Systematic review.
SETTING AND PARTICIPANTS
This review compared strategies for vital sign monitoring in admission avoidance HAH for adults presenting with acute physical illness. Vital sign monitoring can support HAH acute multidisciplinary care by contributing to safety, determining requirement of further assessment, and guiding clinical decisions. There are a wide range of systems currently available, including reliable and automated continuous remote monitoring using wearable devices.
METHODS
Eligible studies were identified through updated database and trial registries searches (March 2, 2016, to February 15, 2023), and existing systematic reviews. Risk of bias was assessed using the Cochrane risk of bias 2 tool. Random effects meta-analyses were performed, and narrative summaries provided stratified by vital sign monitoring approach.
RESULTS
Twenty-one eligible RCTs (3459 participants) were identified. Two approaches to vital sign monitoring were characterized: manual and automated. Reporting was insufficient in the majority of studies for classification. For HAH compared to hospital care, 6-monthly mortality risk ratio (RR) was 0.94 (95% CI 0.78-1.12), 3-monthly readmission to hospital RR 1.02 (0.77-1.35), and length of stay mean difference 1.91 days (0.71-3.12). Readmission to hospital was reduced in the automated monitoring subgroup (RR 0.30 95% CI 0.11-0.86).
CONCLUSIONS AND IMPLICATIONS
This review highlights gaps in the reporting and evidence base informing remote vital sign monitoring in alternatives to admission for acute illness, despite expanding implementation in clinical practice. Although continuous vital sign monitoring using wearable devices may offer added benefit, its use in existing RCTs is limited. Recommendations for the implementation and evaluation of remote monitoring in future clinical trials are proposed.
PubMed: 38908399
DOI: 10.1016/j.jamda.2024.105080 -
Journal of Medical Internet Research Jun 2024Despite the well-documented health benefits associated with wearable monitoring devices (WMDs), adherence among community-dwelling older adults remains low. By providing... (Review)
Review
Effects of Peer- or Professional-Led Support in Enhancing Adherence to Wearable Monitoring Devices Among Community-Dwelling Older Adults: Systematic Review of Randomized Controlled Trials.
BACKGROUND
Despite the well-documented health benefits associated with wearable monitoring devices (WMDs), adherence among community-dwelling older adults remains low. By providing guidance on the purpose and benefits of using WMDs, facilitating goal-setting aligned with the device's features, promoting comprehension of the health data captured by the device, and assisting in overcoming technological challenges, peers and health care professionals can potentially enhance older adults' adherence to WMDs. However, the effectiveness of such support mechanisms in promoting adherence to WMDs among older adults remains poorly understood.
OBJECTIVE
The aims of this systematic review were to examine the effects of peer- or professional-led intervention programs designed to improve adherence to WMDs among community-dwelling older adults and to identify the intervention components that may positively influence the effects of the intervention.
METHODS
We conducted a comprehensive search across 7 electronic databases (Cochrane Central Register of Controlled Trials [CENTRAL], PubMed, EMBASE, PsycINFO, British Nursing Index, Web of Science, and CINAHL) to identify articles published between January 1, 2010, and June 26, 2023. We specifically targeted randomized controlled trials that examined the impact of peer- or professional-led interventions on enhancing adherence to WMDs among individuals aged 60 years and older residing in the community. Two independent reviewers extracted data from the included studies and assessed the potential risk of bias in accordance with the Cochrane Risk of Bias tool for randomized trials, version 2.
RESULTS
A total of 10,511 studies were identified through the database search. Eventually, we included 3 randomized controlled trials involving 154 community-dwelling older adults. The participants had a mean age of 65 years. Our review revealed that increasing awareness of being monitored and implementing the SystemCHANGE approach, a habit change tool focusing on personal goals and feedback, were effective strategies for enhancing adherence to WMDs among older adults. All of the included studies exhibited a low risk of bias.
CONCLUSIONS
By collaboratively designing specific goals related to WMDs with health care professionals, including nurses and physicians, older adults exhibited a higher likelihood of adhering to the prescribed use of WMDs. These goal-setting tools provided a framework for structure and motivation, facilitating the seamless integration of WMDs into their daily routines. Researchers should prioritize interventions that target awareness and goal-setting as effective approaches to enhance adherence to WMDs among older adults, thereby maximizing the realization of associated health benefits.
Topics: Humans; Aged; Independent Living; Wearable Electronic Devices; Randomized Controlled Trials as Topic; Patient Compliance; Peer Group; Middle Aged; Female; Male; Aged, 80 and over
PubMed: 38900546
DOI: 10.2196/53607 -
Frontiers in Endocrinology 2024There has been continuous progress in diabetes management over the last few decades, not least due to the widespread dissemination of continuous glucose monitoring (CGM)...
There has been continuous progress in diabetes management over the last few decades, not least due to the widespread dissemination of continuous glucose monitoring (CGM) and automated insulin delivery systems. These technological advances have radically changed the daily lives of people living with diabetes, improving the quality of life of both children and their families. Despite this, hypoglycemia remains the primary side-effect of insulin therapy. Based on a systematic review of the available scientific evidence, this paper aims to provide evidence-based recommendations for recognizing, risk stratifying, treating, and managing patients with hypoglycemia. The objective of these recommendations is to unify the behavior of pediatric diabetologists with respect to the timely recognition and prevention of hypoglycemic episodes and the correct treatment of hypoglycemia, especially in patients using CGM or advanced hybrid closed-loop systems. All authors have long experience in the specialty and are members of the Italian Society of Pediatric Endocrinology and Diabetology. The goal of treating hypoglycemia is to raise blood glucose above 70 mg/dL (3.9 mmol/L) and to prevent further decreases. Oral glucose at a dose of 0.3 g/kg (0.1 g/kg for children using "smart pumps" or hybrid closed loop systems in automated mode) is the preferred treatment for the conscious individual with blood glucose <70 mg/dL (3.9 mmol/L), although any form of carbohydrate (e.g., sucrose, which consists of glucose and fructose, or honey, sugary soft drinks, or fruit juice) containing glucose may be used. Using automatic insulin delivery systems, the oral glucose dose can be decreased to 0.1 g/kg. Practical flow charts are included to aid clinical decision-making. Although representing the official position of the Italian Society of Pediatric Endocrinology and Diabetology (ISPED), these guidelines are applicable to the global audience and are especially pertinent in the era of CGM and other advanced technologies.
Topics: Humans; Hypoglycemia; Child; Adolescent; Blood Glucose Self-Monitoring; Insulin; Hypoglycemic Agents; Blood Glucose; Diabetes Mellitus, Type 1; Insulin Infusion Systems; Risk Assessment; Practice Guidelines as Topic; Disease Management
PubMed: 38894740
DOI: 10.3389/fendo.2024.1387537 -
Digital Health 2024Wearable technology is used in healthcare to monitor the health of individuals. This study presents an updated systematic literature review of the use of wearable... (Review)
Review
BACKGROUND
Wearable technology is used in healthcare to monitor the health of individuals. This study presents an updated systematic literature review of the use of wearable technology in promoting child and adolescent health, accompanied by recommendations for future research.
METHODS
This review focuses on studies involving children and adolescents aged between 2 and 18 years, regardless of their health condition or disabilities. Studies that were published from 2016 to 2024, and which met the inclusion criteria, were extracted from four academic databases (i.e. PubMed, Cochrane, Embase, and Web of Science) using the Preferred Reporting Items for Systematic Review and Meta-Analyses (PRISMA) protocol. Data on intervention purposes, interventions deployed, intervention duration, measurements, and the main outcomes of the studies were collected.
RESULTS
A total of 53 studies involving 14,852 participants were reviewed. They focused on various aspects, including the ownership and use of wearable devices ( = 3), the feasibility ( = 22), effectiveness ( = 4), and adherence ( = 2) of intervention strategies, or a combination of multiple aspects ( = 22). Among the interventions deployed, Fitbit was the most frequently used, featuring in 26 studies, followed by ActiGraph ( = 11). In intervention studies, the majority of studies focused on pre-morbidity prevention ( = 26) and the treatment of illnesses ( = 20), with limited attention given to postoperative monitoring ( = 4).
CONCLUSIONS
The use of wearable technology by children and adolescents has proven to be both feasible and effective for health promotion. This systematic review summarizes existing research by exploring the use of wearable technology in promoting health across diverse youth populations, including healthy and unhealthy individuals. It examines health promotion at various stages of the disease continuum, including pre-disease prevention, in-disease treatment, and postoperative monitoring. Additionally, the review provides directions for future research.
PubMed: 38868368
DOI: 10.1177/20552076241260507 -
Frontiers in Medicine 2024Ensuring high-quality healthcare for newborns is essential for improving their chances of survival within Ethiopia's healthcare system. Although various intervention...
BACKGROUND
Ensuring high-quality healthcare for newborns is essential for improving their chances of survival within Ethiopia's healthcare system. Although various intervention approaches have been implemented, neonatal mortality rates remain stable. Therefore, the present review seeks to identify initiatives for enhancing healthcare quality, their effects on neonatal wellbeing, and the factors hindering or supporting these Quality Improvement (QI) efforts' success in Ethiopia.
METHODS
We searched for original research studies up to June 23, 2023, using PubMed/Medline, WHO-Global Health Library, Cochrane, Clinical Trials.gov, and Hinari. After selecting eligible studies, we assessed their quality using a mixed-method appraisal tool. Quality of care refers to how healthcare services effectively improve desired outcomes for individuals and patient populations. It encompasses vital principles such as safety, effectiveness, timeliness, efficiency, equity, and patient-centeredness.
RESULTS
We found 3,027 publication records and included 13 studies during our search. All these interventions primarily aimed to provide safe healthcare, with a strong focus on Domain One, which deals with the evidence-based routine upkeep and handling of complications, and Domain Seven, which revolves around ensuring staff competency, emerged as a frequent target for intervention. Many interventions aimed at improving quality also concentrate on essential quality measure elements such as processes, focusing on the activities that occur during care delivery, and quality planning, involving distributing resources, such as basic medicine and equipment, and improving infrastructure. Moreover, little about the facilitators and barriers to QI interventions is investigated.
CONCLUSIONS
This review highlights the significance of introducing QI initiatives in Ethiopia, enhancing the healthcare system's capabilities, engaging the community, offering financial incentives, and leveraging mobile health technologies. Implementing QI interventions in Ethiopia poses difficulties due to resource constraints, insufficient infrastructure, and medical equipment and supplies shortages. It necessitates persistent endeavors to improve neonatal care quality, involving ongoing training, infrastructure enhancement, the establishment of standardized protocols, and continuous outcome monitoring. These efforts are crucial to achieving the optimal outcomes for newborns and their families.
PubMed: 38841585
DOI: 10.3389/fmed.2024.1293473