Revista Latino-americana de Enfermagem 2022to analyze the cleaning process of gastroscopes, colonoscopes and duodenoscopes in eight in-hospital health services.
OBJECTIVE
to analyze the cleaning process of gastroscopes, colonoscopes and duodenoscopes in eight in-hospital health services.
METHOD
a cross-sectional study conducted with 22 endoscopes (eight gastroscopes, eight colonoscopes and six duodenoscopes), and microbiological analysis of 60 samples of air/water channels (all endoscopes) and elevator (duodenoscopes), in addition to protein testing. Descriptive statistics with calculation of frequencies and central tendency measures was used in data analysis.
RESULTS
the processing of 22 endoscopes was monitored with microbiological analysis for 60 channels. In the pre-cleaning procedure, in 82.3% (14/17) of the devices, gauze was used in cleaning the insertion tube. Incomplete immersion of the endoscope in detergent solution occurred in 72.3% (17/22) of the cases, and in 63.6% (14/22) there was no standardization of filling-in of the channels. Friction of the biopsy channel was not performed in 13.6% (3/22) of the devices. In the microbiological analysis, 25% (7/32) of the samples from the stored endoscopes were positive for microbial growth (from 2x101 to 9.5x104 CFU/mL), while after processing, contamination was 32% (9/28). Protein residues in the elevator channel were detected in 33% of duodenoscopes.
CONCLUSION
the results indicate important gaps in the stages of pre-cleaning and cleaning of endoscopes that, associated with presence of protein residues and growth of microorganisms of epidemiological importance, indicate limitations in safety of the processing procedures, which can compromise the disinfection processes and, consequently, their safe use among patients subjected to such tests.
Topics: Humans; Equipment Contamination; Cross-Sectional Studies; Detergents; Endoscopes; Disinfection; Hospitals
PubMed: 36287399
DOI: 10.1590/1518-8345.5969.3684 -
Annals of Hepato-biliary-pancreatic... Feb 2023Data regarding outcomes of endoscopic retrograde cholangiography (ERC) in liver transplant (LT) recipients with biliary-enteric (BE) anastomosis are limited. We report...
Outcomes of endoscopic retrograde cholangiography and percutaneous transhepatic biliary drainage in liver transplant recipients with a Roux-en-Y biliary-enteric anastomosis.
BACKGROUNDS/AIMS
Data regarding outcomes of endoscopic retrograde cholangiography (ERC) in liver transplant (LT) recipients with biliary-enteric (BE) anastomosis are limited. We report outcomes of ERC and percutaneous transhepatic biliary drainage (PTBD) as first-line therapies in LT recipients with BE anastomosis.
METHODS
All LT recipients with Roux-BE anastomosis from 2001 to 2020 were divided into ERC and PTBD subgroups. Technical success was defined as the ability to cannulate the bile duct. Clinical success was defined as the ability to perform cholangiography and therapeutic interventions.
RESULTS
A total of 36 LT recipients (25 males, age 53.5 ± 13 years) with Roux-BE anastomosis who underwent biliary intervention were identified. The most common indications for a BE anastomosis were primary sclerosing cholangitis (n = 14) and duct size mismatch (n = 10). Among the 29 patients who initially underwent ERC, technical success and clinical success were achieved in 24 (82.8%) and 22 (75.9%) patients, respectively. The initial endoscope used for the ERC was a single balloon enteroscope in 16 patients, a double balloon enteroscope in 7 patients, a pediatric colonoscope in 5 patients, and a conventional reusable duodenoscope in 1 patient. Among the 7 patients who underwent PTBD as the initial therapy, six (85.7%) achieved technical and clinical success ( = 0.57).
CONCLUSIONS
In LT patients with Roux-BE anastomosis requiring biliary intervention, ERC with a balloon-assisted enteroscope is safe with a success rate comparable to PTBD. Both ERC and PTBD can be considered as first-line therapies for LT recipients with a BE anastomosis.
PubMed: 36245257
DOI: 10.14701/ahbps.22-037 -
The Journal of Hospital Infection Jan 2023Microbiological surveillance of endoscopes is a safety measure for verifying the quality of reprocessing procedures and identifying contaminated devices, but...
INTRODUCTION
Microbiological surveillance of endoscopes is a safety measure for verifying the quality of reprocessing procedures and identifying contaminated devices, but duodenoscope-related outbreaks are still reported.
AIM
To assess the effectiveness of duodenoscope reprocessing procedures in Italy.
METHODS
Between December 2019 and April 2020, data obtained from microbiological surveillance post-reprocessing in 15 Italian endoscopy units were collected. Sampling was carried out after reprocessing or during storage in a cabinet. In keeping with international guidelines and the Italian position paper, the micro-organisms were classified as high-concern organisms (HCOs) and low-concern organisms (LCOs).
FINDINGS
In total, 144 samples were collected from 51 duodenoscopes. Of these, 36.81% were contaminated: 22.92% were contaminated with HCOs and 13.89% were contaminated with LCOs [2.08% with an LCO load of 11-100 colony-forming units (CFU)/device and 0.69% with an LCO load of >100 CFU/device]. The contamination rate was 27.5% in samples collected after reprocessing, 40% in samples collected during storage in a cabinet that was compliant with EN 16442:2015 (C-I), and 100% in samples collected during storage in a cabinet that was not compliant with EN 16442:2015 (NC-I). The respective HCO rates were 15.00%, 27.27% and 66.67%. Correlation between LCO contamination and storage time was demonstrated (Spearman's rho=0.3701; P=0.0026). The Olympus duodenoscope TJFQ180V demonstrated the lowest rate of contamination (29.82%), although the contamination rate was 100% for duodenoscopes stored in an NC-I cabinet.
CONCLUSION
Microbiological surveillance, along with strict adherence to reprocessing protocols, may help to detect endoscope contamination at an early stage, and reduce the risk of duodenoscope-associated infections.
Topics: Humans; Equipment Contamination; Disinfection; Endoscopes; Duodenoscopes; Endoscopy, Gastrointestinal
PubMed: 36244520
DOI: 10.1016/j.jhin.2022.09.024 -
GE Portuguese Journal of... Sep 2022Duodenal duplication cysts (DDCs) are rare congenital anomalies typically manifesting during childhood. Clinical manifestations are uncommon in adulthood. DDCs were...
BACKGROUND
Duodenal duplication cysts (DDCs) are rare congenital anomalies typically manifesting during childhood. Clinical manifestations are uncommon in adulthood. DDCs were classically treated surgically, but endoscopic treatment has been increasingly reported. Endoscopic cyst marsupialization establishes a communication between the cyst cavity and the duodenal lumen so that the cystic content can be drained continuously into the duodenum. We herein describe two cases of symptomatic DDCs diagnosed in adulthood and submitted to endoscopic marsupialization using different techniques and devices.
CASE SUMMARY
Case 1: A 23-year-old female patient was admitted with the diagnosis of acute pancreatitis. Endoscopic ultrasound revealed a 35-mm duodenal subepithelial lesion whose proximal limit was immediately distal to the ampulla of Vater and filled with fluid and calcifications. Using a duodenoscope, deroofing of the lesion was made with a diathermic snare. Pathology confirmed the diagnosis of DDC. Case 2: A 41-year-old female, submitted to laparoscopic cholecystectomy 1 month earlier due to suspected lithiasic acute pancreatitis, was admitted due to suspicion of iatrogenic biliary fistula. An endoscopic retrograde cholangiopancreatography was performed and the bile leak was treated. Immediately distal to the papillary orifice, a 20-mm subepithelial lesion was also detected. A biopsy forceps was used to fenestrate its wall, allowing the exit of mucous fluid and stones, and a sphincterotome was used to expand the incision. No recurrence was documented in both cases.
CONCLUSION
These cases highlight DDC as a potential cause for acute pancreatitis in adults and endoscopy as an easy treatment option.
PubMed: 36159200
DOI: 10.1159/000518586 -
Endoscopy International Open Sep 2022Adenosine triphosphate (ATP) tests are increasingly used to detect biological material; however, their reliability to detect bacterial contamination in endoscopes is...
Adenosine triphosphate (ATP) tests are increasingly used to detect biological material; however, their reliability to detect bacterial contamination in endoscopes is not proven. We investigated the predictive value of ATP tests after manual cleaning for presence or absence of microorganisms as shown by culture after automated high-level disinfection (HLD) in duodenoscopes and linear echoendoscopes (DLEs). After manual cleaning, ATP tests were performed on swab samples taken from the detachable cap and forceps elevator, and on flush samples of the DLE working channels. These results were compared to the growth of any microorganisms in cultures acquired after automated HLD. ATP tests with > 200 relative light units (RLU) were considered positive. Receiver operator characteristic (ROC) curves were used to compare the RLU levels with microbial presence in cultures. In total, 903 procedures were performed involving 26 distinct DLEs. Depending on sample site, 20.8 % (cap) to 63.8 % (channel brush) of the ATP negative samples were accompanied by positive post-HLD cultures. 54.4 % of the cap samples with a positive culture (growth of any kind of microorganism) and 91.8 % of the channel samples with a positive culture had a negative ATP test after manual cleaning. ROC curves per sample site, DLE type and microorganism type all had area under the curves below 0.6. In our study, ATP tests performed after manual cleaning could not predict presence or absence of microorganisms after automated HLD as shown by culture. More than half of the positive cultures were preceded by a negative ATP test.
PubMed: 36118629
DOI: 10.1055/a-1897-5000 -
Indian Journal of Anaesthesia Jul 2022
PubMed: 36111095
DOI: 10.4103/ija.ija_520_22 -
Indian Journal of Anaesthesia Jul 2022
PubMed: 36111087
DOI: 10.4103/ija.ija_232_22 -
Pediatric Gastroenterology, Hepatology... Jul 2022Although endoscopic retrograde cholangiopancreatography (ERCP) has been used for more than five decades, its applicability in Bangladeshi children has recently become...
PURPOSE
Although endoscopic retrograde cholangiopancreatography (ERCP) has been used for more than five decades, its applicability in Bangladeshi children has recently become more common. Therefore, this manuscript aims to describe our experience in performing ERCPs in Bangladeshi children with hepatopancreaticobiliary diseases, focusing on presenting diseases, as well as the diagnostic and therapeutic efficacy.
METHODS
Between 2018 and 2021, 20 children underwent 30 ERCP procedures at the Bangladesh Specialized Hospital, Dhaka. A single trained adult gastroenterologist performed all procedures using a therapeutic video duodenoscope. The indications for ERCP, diagnostic findings, therapeutic procedures, and complications were documented.
RESULTS
The median age of the study patients was 10 years (range, 1.7-15 years). Successful cannulation of the papilla was achieved in 28 procedures and failed in 2 cases. Repeated ERCP was required in seven patients. Nine patients had biliary indications and 11 had pancreatic indications. Choledocholithiasis was the most common indication for ERCP in patients with biliary disease, while chronic pancreatitis was common among patients with pancreatic indications. Pancreatic divisum was observed in only one patient. Pancreatic and biliary sphincterotomy was performed in 14 and 9 cases, respectively. A single pigtail or straight therapeutic stent was inserted in seven cases and removed in five cases. Stone extraction was performed in six procedures, and balloon dilatation was performed in five procedures. The post-procedural period for these patients was uneventful.
CONCLUSION
We found that ERCP is a practical and successful therapeutic intervention for treating hepatopancreaticobiliary disorders in children when performed by experienced endoscopists.
PubMed: 35903495
DOI: 10.5223/pghn.2022.25.4.332 -
Annals of Gastroenterology 2022Despite the standardization and optimization of disinfection protocols, duodenoscope-related infections (DRIs) remain an emerging threat for patients undergoing...
BACKGROUND
Despite the standardization and optimization of disinfection protocols, duodenoscope-related infections (DRIs) remain an emerging threat for patients undergoing endoscopic retrograde cholangiopancreatography (ERCP). Single-use duodenoscopes could represent a potential alternative avenue to circumvent the problem of reprocessing and thus risk of exogenous patient-to-patient transmission. The aim of our study was to test the feasibility and technical success rate of a recently made available single-use duodenoscope.
METHODS
We carried out a thorough and standardized evaluation of the usability, performance and safety of a recently developed single-use duodenoscope in 52 consecutive patients scheduled for ERCP in a single center. Outcomes included performance ratings of the single-use duodenoscopes, adverse events (assessed at 3 days and 1 week), and crossover rate to a reusable duodenoscope.
RESULTS
The ERCP completion rate with a single-use duodenoscope was 90.4%, rising to 94.2% after crossover to reusable duodenoscope. The mean American Society for Gastrointestinal Endoscopy (ASGE) grade was 2.7, with 27 procedures (51.9%) considered as advanced level complexity (ASGE grade 3 & 4). Performance rating found that 94% of the therapeutic treatments were comparable to those using a traditional reusable duodenoscope. Overall satisfaction amounted to 80%. No major adverse events were attributable to the single-use endoscope.
CONCLUSIONS
Single-use duodenoscopes can provide an alternative to avoid the intensive and often inconsistent results of cleaning and disinfection procedures. We confirmed the feasibility, adequate performance characteristics and safety of a recently developed first-generation single-use duodenoscope over a broad range of ERCP procedures, in terms of both indication and complexity.
PubMed: 35784631
DOI: 10.20524/aog.2022.0723 -
World Journal of Methodology May 2022Several strategies have been implemented to reduce or abolish the life-threatening risk of endoscopic retrograde cholangiopancreatography (ERCP)-related...
BACKGROUND
Several strategies have been implemented to reduce or abolish the life-threatening risk of endoscopic retrograde cholangiopancreatography (ERCP)-related multidrug-resistant infections due to duodenoscopes contaminations; among those strategies, serial microbiologic tests, thorough reprocessing schedules, and use of removable scope cap have been adopted, but the potential cross-infection risk was not eliminated.
AIM
To review available evidence in the field of single-use duodenoscopes (SUD) use for ERCP.
METHODS
An overview on ongoing clinical studies was also performed to delineate which data will become available in the next future.
RESULTS
One bench comparative study and four clinical trials performed with EXALT model-D (Boston Scientific Corp., United States) have been identified. Of them, one is a randomized controlled trial, while the other three studies are prospective single-arm, cross-over studies. Pooled technical success rate (4 studies, 368 patients) was 92.9% [95% confidence interval (CI): 89.9-95.5; : 11.8%]. Pooled serious adverse event (4 studies, 381 patients) rate was 5.9% [3.7%-8.5%; : 0.0%].
CONCLUSION
Although few clinical trials are available, evidence is concordant in identifying an absolute feasibility and safety and feasibility for SUD use for ERCP. The expertise and quality of evidence in this field are going to be improved by further large clinical trials;data on cost-effectiveness and environmental impact will be needed for a worldwide spread of SUD use for ERCP.
PubMed: 35721249
DOI: 10.5662/wjm.v12.i3.122