American Journal of Infection Control Apr 2024Over the past decade, infections linked to duodenoscopes have become a significant concern, primarily due to the intricate design of the elevator mechanism. Currently,...
Over the past decade, infections linked to duodenoscopes have become a significant concern, primarily due to the intricate design of the elevator mechanism. Currently, there is limited evidence regarding the bacterial contamination level of the elevator mechanism after clinical use and throughout its various reprocessing stages. This study utilized the swab culture technique to examine the bacterial contamination on the duodenoscope elevator mechanism after clinical use and after 3 reprocessing stages at a Center of tertiary hospital. Our findings revealed severe bacterial contamination after clinical usage, emphasizing that the effectiveness of manual cleaning greatly influences the subsequent high-level disinfection quality.
Topics: Humans; Duodenoscopes; Equipment Contamination; Bacteria; Disinfection
PubMed: 37838249
DOI: 10.1016/j.ajic.2023.10.003 -
Ibrain 2023This study aimed to determine the values of the half-effective dose (ED) and 95% effective dose (ED) of remimazolam besylate used in the procedural sedation of...
This study aimed to determine the values of the half-effective dose (ED) and 95% effective dose (ED) of remimazolam besylate used in the procedural sedation of endoscopic retrograde cholangiopancreatography (ERCP). Sixty patients who fulfilled the inclusion and exclusion criteria of this study were selected. Sufentanil was administered intravenously and remimazolam besylate was administered 2 min later. ERCP treatment was feasible when the modified alertness/sedation (MOAA/S) score was ≤2. If choking or movement occurred during duodenoscope placement, it was considered as a positive reaction. The dose was increased in the next patient; otherwise, it was considered as a negative reaction, and the dose was reduced in the next patient. The ED and ED values and 95% confidence interval (CI) of remimazolam besylate were calculated by Probit regression analysis. All 60 patients completed the trial. The ED and ED values of remimazolam besylate were 0.196 and 0.239 mg/kg, respectively, for the procedural sedation of ERCP. The time of MOAA/S score ≤ 2 was (82.58 ± 21.70) s, and the mean time of awakening was (9.03 ± 5.64) min. Transient hypotension was observed in two patients without medical intervention. The ED and ED values of remimazolam besylate used in the procedural sedation of ERCP were 0.196 and 0.239 mg/kg, and the dose of the medications has definite efficacy and good safety.
PubMed: 37786755
DOI: 10.1002/ibra.12072 -
American Journal of Infection Control Mar 2024An inpatient endoscopy unit is a care hub for patients from throughout the hospital and can be the site of health care-associated infections (HAIs). Shared surfaces and...
BACKGROUND
An inpatient endoscopy unit is a care hub for patients from throughout the hospital and can be the site of health care-associated infections (HAIs). Shared surfaces and other nonmedical devices (keyboards) have been increasingly recognized as sites of pathogen transmission. Beyond standard cleaning of high-touch target areas, we queried whether the addition of automated devices delivering low-intensity doses of ultraviolet (UV)-C radiation could further reduce bioburden in an academic endoscopy unit.
METHODS
Bioburden on previously identified high-touch/communal surfaces was measured before and after the installation of automated, low-intensity UV-light emitting devices (UV Angel) that passively monitor and disinfect targeted surfaces with Ultraviolet-C light (UV-C) light.
RESULTS
High-touch sites (keyboards) had a baseline bacterial contamination of >80%, whereas individual procedure rooms and common areas had a >57% contamination rate. Following the implementation of automated UV-C light decontamination, bioburden was reduced on average by >91% at high-touch surfaces and within procedure rooms.
DISCUSSION/CONCLUSIONS
Nonsterile hubs of patient care could serve as sites of "silent" HAI transmission. We have identified high-touch surfaces within an endoscopy unit that have a high bioburden of bacterial contamination and demonstrated that the installation of passive, automated UV-C light disinfection devices can reduce bioburden significantly, possibly mitigating HAI transmission between patients.
Topics: Humans; Duodenoscopes; Tertiary Healthcare; Hospitals; Bacteria; Cross Infection; Endoscopy, Gastrointestinal; Ultraviolet Rays; Disinfection
PubMed: 37776900
DOI: 10.1016/j.ajic.2023.09.016 -
Revista Espanola de Enfermedades... Sep 2023A 52-year-old male diagnosed with severe acute pancreatitis developed persistent abdominal distension and intermittent emesis six weeks after treatment. Computed...
Transpapillary pancreatic duct stenting in the treatment of acute severe pancreatitis complicated by pancreatic pseudocyst compressing portal vein leading to local portal hypertension.
A 52-year-old male diagnosed with severe acute pancreatitis developed persistent abdominal distension and intermittent emesis six weeks after treatment. Computed tomography (CT) showed that the shape of the pancreas was not clear, and a pseudocyst (11.2*6.6 cm) existed in the tail of the pancreas, which compressed the portal vein and led to local portal hypertension. A transpapillary pancreatic duct stent (7F*9 cm) was successfully placed for drainage of fluid collection despite pancreatic duct obstruction, with one end into the pseudocyst cavity and the other into the duodenum. Duodenoscope observed outflow of the fluid collection immediately. Four days after the surgery, CT scan validated the location of the pancreatic duct stent, and the pancreatic pseudocyst was smaller than before. Another week later, the pancreatic pseudocyst nearly disappeared. No significant esophagogastric varix was observed by gastroscope, indicating a regression of local portal hypertension. The patient resumed eating normally, did not complain of vomiting, and was discharged from our hospital.
PubMed: 37771282
DOI: 10.17235/reed.2023.9930/2023 -
Life (Basel, Switzerland) Aug 2023Duodenoscope-related infections are a major concern in medicine and GI endoscopy, especially in fragile patients. Disposable duodenoscopes seem to be the right tool to... (Review)
Review
Duodenoscope-related infections are a major concern in medicine and GI endoscopy, especially in fragile patients. Disposable duodenoscopes seem to be the right tool to minimize the problem: a good choice for patients with many comorbidities or with a high risk of carrying multidrug resistant bacteria. Urgent endoscopy could also be a good setting for the use of single-use duodenoscopes, especially when the risk of the infection cannot be evaluated. Their safety and efficacy in performing ERCP has been proven in many studies. However, randomized clinical trials and comparative large studies with reusable scopes are lacking. Moreover, the present early stage of their introduction on the market does not allow a large economical evaluation for each health system. Thus, accurate economical and safety comparisons with cap-disposable duodenoscopes are needed. Moreover, the environmental impact of single-use duodenoscopes should be carefully evaluated, considering the ongoing climate change. In conclusion, definitive guidelines are needed to choose wisely the appropriate patients for ERCP with disposable duodenoscopes as the complete switch to single-use duodenoscopes seems to be difficult, to date. Many issues are still open, and they need to be carefully evaluated in further, larger studies.
PubMed: 37629551
DOI: 10.3390/life13081694 -
Cureus Jul 2023Previous gastric procedures often make endoscopic interventions challenging. Our case study focuses specifically on performing an endoscopic retrograde...
Previous gastric procedures often make endoscopic interventions challenging. Our case study focuses specifically on performing an endoscopic retrograde cholangiopancreatography (ERCP) through a gastroscope (EVIS EXERA III GIF-HQ190, Olympus, Center Valley, USA) in a patient with a history of Billroth II gastrojejunostomy. Successful ERCP in Billroth II using a gastroscope with traditional ERCP instrumentation has been very rarely reported in case reports in the literature review. This case study provides an alternative method of access to the common bile duct (CBD) and treatment of obstruction to prevent the risk of morbidities from an open CBD exploration. The primary diagnosis for this patient was choledocholithiasis. He initially underwent a standard ERCP with a side-viewing duodenoscope (EVIS EXERA II, TJF-Q190V, Olympus, Center Valley, USA); however, due to the difficult anatomy from his previous Billroth II reconstruction, the CBD was very difficult to access. A gastroscope was then used instead to perform the ERCP, providing an innovative endoscopic therapy. Given the patient's multiple comorbidities, he was at high risk for morbidity and mortality with an open CBD exploration. Hence, this case report provides insight into an innovative endoscopic approach to CBD exploration with difficult anatomy.
PubMed: 37575799
DOI: 10.7759/cureus.41793 -
VideoGIE : An Official Video Journal of... Aug 2023Duodenal polyps have a reported incidence of 0.3% to 4.6%. Sporadic, nonampullary duodenal adenomas (SNDAs) comprise less than 10% of all duodenal polyps, and ampullary... (Review)
Review
BACKGROUND AND AIMS
Duodenal polyps have a reported incidence of 0.3% to 4.6%. Sporadic, nonampullary duodenal adenomas (SNDAs) comprise less than 10% of all duodenal polyps, and ampullary adenomas are even less common. Nonetheless, the incidence continues to rise because of widespread endoscopy use. Duodenal polyps with villous features or those that are larger than 10 mm may raise concern for malignancy and require removal. We demonstrate endoscopic resection of SNDAs and ampullary adenomas using some of our preferred techniques.
METHODS
The duodenum has several components that can make EMR of duodenal polyps technically challenging. Not only does the duodenum have a thin muscle layer, but it is also highly mobile and vascular, which may explain higher rates of perforation and bleeding of duodenal EMR reported in the literature compared with colon EMR. A standard adult gastroscope with a distal cap is commonly used for duodenal EMRs. Based on the location, however, side-viewing duodenoscopes or pediatric colonoscopes may be used. To prepare for EMR, a submucosal injection is performed for an adequate lift. The polyp is then resected via stiff monofilament snares and subsequently closed with hemostatic clips if feasible. The ampullectomy technique differs slightly from duodenal EMRs and carries the additional risk of pancreatitis. Submucosal injection in the ampulla may not lift well; thus, its utility is debatable. Biliary sphincterotomy should be performed, and based on endoscopist preference, the pancreatic duct (PD) guidewire can be left during resection to maintain access. After resection, a PD stent is placed to minimize pancreatitis risk.
RESULTS
The video shows the aforementioned duodenal EMR techniques. Two clips of ampullectomy are also shown in the video.
CONCLUSIONS
A few common techniques used to perform duodenal EMR and ampullectomy are highlighted in the video. It is important to understand and manage adverse events associated with these procedures and to have established surveillance plans.
PubMed: 37575136
DOI: 10.1016/j.vgie.2023.05.006 -
Endoscopy Dec 2023The first commercialized single-use duodenoscope was cleared by the US Food and Drug Administration in December 2019. Data regarding endoscopic retrograde... (Clinical Trial)
Clinical Trial
BACKGROUND
The first commercialized single-use duodenoscope was cleared by the US Food and Drug Administration in December 2019. Data regarding endoscopic retrograde cholangiopancreatography (ERCP) using a single-use duodenoscope are needed on a broader range of cases conducted by endoscopists with varying levels of experience in a wide range of geographic areas.
METHODS
61 endoscopists at 22 academic centers in 11 countries performed ERCP procedures in adult patients aged ≥ 18. Outcomes included ERCP completion for the intended indication, rate of crossover to a reusable endoscope, device performance ratings, and serious adverse events (SAEs).
RESULTS
Among 551 patients, 236 (42.8 %) were aged > 65, 281 (51.0 %) were men, and 256 (46.5 %) had their procedure as an inpatient. ERCPs included 196 (35.6 %) with American Society for Gastrointestinal Endoscopy complexity of grades 3-4. A total of 529 ERCPs (96.0 %) were completed: 503 (91.3 %) using only the single-use duodenoscope, and 26 (4.7 %) with crossover to a reusable endoscope. There were 22 ERCPs (4.0 %) that were not completed, of which 11 (2.0 %) included a crossover and 11 (2.0 %) were aborted cases (no crossover). Median ERCP completion time was 24.0 minutes. Median overall satisfaction with the single-use duodenoscope was 8.0 (scale of 1 to 10 [best]). SAEs were reported in 43 patients (7.8 %), including 17 (3.1 %) who developed post-ERCP pancreatitis.
CONCLUSIONS
In academic medical centers over a wide geographic distribution, endoscopists with varying levels of experience using the first marketed single-use duodenoscope had good ERCP procedural success and reported high performance ratings for this device.
Topics: Adult; Male; Humans; Female; Cholangiopancreatography, Endoscopic Retrograde; Duodenoscopes; Endoscopy, Gastrointestinal; Pancreatitis
PubMed: 37463599
DOI: 10.1055/a-2131-7180 -
Clinical Endoscopy Jan 2024Selective bile duct or pancreatic duct cannulation remains a significant initial hurdle in endoscopic retrograde cholangiopancreatography (ERCP) despite advances in...
BACKGROUND/AIMS
Selective bile duct or pancreatic duct cannulation remains a significant initial hurdle in endoscopic retrograde cholangiopancreatography (ERCP) despite advances in endoscopy and accessories. This study evaluated our experience with a rotatable sphincterotome in cases of difficult cannulation.
METHODS
We retrospectively reviewed ERCP cases using TRUEtome, a rotatable sphincterotome, as a rescue device for cannulation at a cancer institute in Japan from October 2014 to December 2021.
RESULTS
TRUEtome was used in 88 patients. Duodenoscopes were used for 51 patients, while single-balloon enteroscopes (SBE) were used for 37 patients. TRUEtome was used for biliary and pancreatic duct cannulation (84.1%), intrahepatic bile duct selection (12.5%), and strictures of the afferent limb (3.4%). Cannulation success rates were similar in the duodenoscope and SBE groups (86.3% vs. 75.7%, p=0.213). TRUEtome was more commonly used in cases with steep cannulation angles in the duodenoscope group and in cases requiring cannulation in different directions in the SBE group. There were no significant differences in adverse events between the two groups.
CONCLUSION
The cannulation sphincterotome was useful for difficult cannulations in both unaltered and surgically altered anatomies. It may be an option to consider before high-risk procedures such as precut and endoscopic ultrasound-guided rendezvous techniques.
PubMed: 37157962
DOI: 10.5946/ce.2022.248 -
Infection Control and Hospital... Sep 2023To evaluate variables that affect risk of contamination for endoscopic retrograde cholangiopancreatography and endoscopic ultrasound endoscopes. (Observational Study)
Observational Study
OBJECTIVE
To evaluate variables that affect risk of contamination for endoscopic retrograde cholangiopancreatography and endoscopic ultrasound endoscopes.
DESIGN
Observational, quality improvement study.
SETTING
University medical center with a gastrointestinal endoscopy service performing ∼1,000 endoscopic retrograde cholangiopancreatography and ∼1,000 endoscopic ultrasound endoscope procedures annually.
METHODS
Duodenoscope and linear echoendoscope sampling (from the elevator mechanism and instrument channel) was performed from June 2020 through September 2021. Operational changes during this period included standard reprocessing with high-level disinfection with ethylene oxide gas sterilization (HLD-ETO) was switched to double high-level disinfection (dHLD) (June 16, 2020-July 15, 2020), and duodenoscopes changed to disposable tip model (March 2021). The frequency of contamination for the co-primary outcomes were characterized by calculated risk ratios.
RESULTS
The overall pathogenic contamination rate was 4.72% (6 of 127). Compared to duodenoscopes, linear echoendoscopes had a contamination risk ratio of 3.64 (95% confidence interval [CI], 0.69-19.1). Reprocessing using HLD-ETO was associated with a contamination risk ratio of 0.29 (95% CI, 0.06-1.54). Linear echoendoscopes undergoing dHLD had the highest risk of contamination (2 of 18, 11.1%), and duodenoscopes undergoing HLD-ETO and the lowest risk of contamination (0 of 53, 0%). Duodenoscopes with a disposable tip had a 0% contamination rate (0 of 27).
CONCLUSIONS
We did not detect a significant reduction in endoscope contamination using HLD-ETO versus dHLD reprocessing. Linear echoendoscopes have a risk of contamination similar to that of duodenoscopes. Disposable tips may reduce the risk of duodenoscope contamination.
Topics: Humans; Cholangiopancreatography, Endoscopic Retrograde; Disinfection; Duodenoscopes; Endosonography; Equipment Contamination
PubMed: 36645014
DOI: 10.1017/ice.2022.319