Journal of Glaucoma Jun 2024Different mechanisms of angle closure represented distinct aspects of intraocular pressure (IOP) control after phacoemulsification. Classification of angle closure...
PRECIS
Different mechanisms of angle closure represented distinct aspects of intraocular pressure (IOP) control after phacoemulsification. Classification of angle closure mechanisms is necessary for postoperative IOP management and glaucoma progression in primary angle closure eyes.
PURPOSE
To investigate the relationship between the anterior chamber angle (ACA) characteristics, measured by swept-source anterior segment optical coherence tomography (SS AS-OCT), and intraocular pressure (IOP) control after phacoemulsification in eyes with primary angle closure disease (PACD) with different angle closure mechanisms.
METHODS
PACD eyes were classified into 3 groups according to angle closure mechanisms using preoperative SS AS-OCT images; pupillary block (PB), plateau iris configuration (PIC), exaggerated lens vault (ELV). This retrospective, clinical cohort study included eighty-five eyes of 85 PACD patients: 34 with PB, 23 with PIC, and 28 with ELV. ACA parameters were measured preoperatively and 1 month postoperatively using SS AS-OCT. IOP measurements were performed preoperatively and during six months postoperatively. Postoperative IOP reduction and fluctuation were calculated, and their correlations with SS AS-OCT parameters were analyzed.
RESULTS
PIC group showed the lowest postoperative IOP reduction compared to the other groups (P=0.023). Preoperative ACA measurements were significantly associated with postoperative IOP reduction in ELV and PB groups, while postoperative measurements were in PIC group. Preoperative and postoperative change of iridotrabecular contact (ITC) index and area were correlated with postoperative IOP reduction in PB and ELV groups but not in PIC group. Postoperative ITC index (P=0.031) and area (P=0.003) showed significant correlations with postoperative IOP fluctuation only in PIC group.
CONCLUSIONS
SS AS-OCT parameters including ITC index and area showed different associations with postoperative IOP control, which should be considered in determination of lens extraction and treatment of PACD eyes.
PubMed: 38934849
DOI: 10.1097/IJG.0000000000002454 -
Diagnostics (Basel, Switzerland) Jun 2024Phacoemulsification is the standard of care in cataract surgery in the developed world, with patients having high expectations regarding visual results. Postoperative...
Phacoemulsification is the standard of care in cataract surgery in the developed world, with patients having high expectations regarding visual results. Postoperative dissatisfaction due to negative dysphotopsia (ND) ranges from rare to very frequent; its etiology is unclear, and it affects postoperative satisfaction. Since one of the most frequently used strategies to avoid ND is related to intraocular lens (IOL) haptic orientation, we conducted a prospective interventional study that enrolled 197 patients who underwent standard phacoemulsification. All patients had a one-piece hydrophobic acrylic IOL implanted; in one group, the haptics were placed in any meridional axis except inferotemporal (IT) meridians, and in the other group, the IOL was implanted with the haptics in an IT position. Our results showed no statistically significant differences between groups when analyzing the correlation between the position of IOL haptics and the presence of ND in week one and month one. Also, pupillary diameter showed no statistically significant differences between patients with or without ND. Despite some studies claiming that haptic orientation prevents ND, we found that haptic orientation does not correlate with ND incidence and that ND decreases from day 1 to month 1. Our results support previous findings on the decrease in ND over time and that haptic orientation should not be considered an intraoperative strategy to avoid this unwanted phenomenon.
PubMed: 38928703
DOI: 10.3390/diagnostics14121288 -
International Ophthalmology Jun 2024To examine the rate of ciliary body detachment in patients with choroidal detachment following glaucoma surgery and its effect on the clinical course, management, and... (Observational Study)
Observational Study
PURPOSE
To examine the rate of ciliary body detachment in patients with choroidal detachment following glaucoma surgery and its effect on the clinical course, management, and prognosis.
METHODS
A prospective observational case-series study. Patients with choroidal detachment following glaucoma surgery in 2018-2019 were included. All underwent complete ophthalmological examination and ultrasound biomicroscopy for evaluation of the presence and extent of ciliary body detachment. Follow-up examinations including ultrasound biomicroscopy scans were performed at 1 week, 1 month, 3 months, and 6 months.
RESULTS
Eight patients (8 eyes) were enrolled, 4 male and 4 female, of mean age 72 years (range 60-83). Five patients underwent trabeculectomy with mitomycin C (0.02%), which was combined with phacoemulsification cataract extraction in one; two underwent Ahmed glaucoma valve implantations, and one underwent ab-interno Xen45 gel stent implantation with mitomycin C (0.02%). The mean intraocular pressure was 26.0 ± 7.65 mmHg preoperatively, dropping to 6.9 ± 2.64 mmHg on first postoperative day one. Mean time from surgery to diagnosis of choroidal detachment was 11.6 ± 5.73 days. Ciliary body detachment was identified by ultrasound biomicroscopy in all patients, ranging between one and four quadrants. All patients were treated with topical steroids and cycloplegics; three (37.5%) received oral steroids. No surgical intervention for the choroidal or ciliary body detachments was indicated.
CONCLUSIONS
In this real-world prospective study, concurrent ciliary body detachment was identified in all patients who presented with choroidal detachment following glaucoma surgery. This observation may deepen our understanding of the mechanism underlying the hypotony that is often seen after glaucoma surgery.
Topics: Humans; Male; Female; Ciliary Body; Aged; Prospective Studies; Middle Aged; Aged, 80 and over; Intraocular Pressure; Choroidal Effusions; Glaucoma; Postoperative Complications; Microscopy, Acoustic; Follow-Up Studies; Trabeculectomy; Glaucoma Drainage Implants; Visual Acuity; Uveal Diseases; Tomography, Optical Coherence
PubMed: 38922523
DOI: 10.1007/s10792-024-03219-1 -
Geriatrics (Basel, Switzerland) Jun 2024The aim of this study was to evaluate the influence of ultrasounds used in phacoemulsification during cataract surgery on the corneal structure and morphology in...
The aim of this study was to evaluate the influence of ultrasounds used in phacoemulsification during cataract surgery on the corneal structure and morphology in patients over 65 years. We compared the outcomes of phacoemulsification techniques in terms of corneal cell morphology in 77 patients over 65 years old and 43 patients under 65 years old. Corneal cell density, central corneal thickness and hexagonality were measured preoperatively and post-surgery (at 1 and 4 weeks) by specular microscopy. The effect of gender, axial length and anterior chamber depth on the parameters of corneal endothelium were evaluated. In both groups, a progressive decrease in endothelial cells was observed, starting from the first week post-surgery until the fourth postoperative week. The central corneal thickness increased in both groups with maximum values at the first week postoperatively, while their initial values were restored in the fourth week post-surgery, with no statistical difference between groups. Statistically significant differences were noticed in terms of cell hexagonality in the group over 65, showing smaller hexagonality at all preoperative and postoperative time points compared to group under 65. Our result highlights the importance of routine specular microscopy performed before surgery, regardless the age of the patients, with caution and careful attention to the phaco power intensity, ultrasound energy consumption and intraoperative manipulation of instruments, as well as proper use of viscoelastic substances to reduce corneal endothelium damage, especially in elderly patients.
PubMed: 38920433
DOI: 10.3390/geriatrics9030077 -
BMC Ophthalmology Jun 2024To evaluate the optical performance and safety of a new multifocal lens with a novel optical design featuring two additional foci (or intensifiers) in patients with... (Observational Study)
Observational Study
BACKGROUND
To evaluate the optical performance and safety of a new multifocal lens with a novel optical design featuring two additional foci (or intensifiers) in patients with cataract and presbyopia.
METHODS
In this single-center, non-randomized prospective observational study, 31 patients underwent implantation of the new multifocal IOL between March 2020 and November 2021 at a tertiary clinical center in Buenos Aires and Ramos Mejia, Argentina. Postoperative examinations with emphasis on uncorrected and corrected visual acuity at distance and near and at two different intermediate distances (80 cm and 60 cm) were performed during the 3 postoperative months.
RESULTS
Of the 31 patients who underwent implantation of the new IOL, 30 underwent bilateral surgery (61 eyes in total). At 3 months, all 61 eyes had an uncorrected distance visual acuity (UCDVA) of at least 0.15 logMAR; 57 eyes (93%) had an uncorrected distance visual acuity (UCDVA) of 0.1 logMAR and 27 eyes (44%) had an UCDVA of 0.0 logMAR. At 80 cm, 60 eyes (98%) had an uncorrected intermediate visual acuity (UCIVA) of at least 0.1 log MAR and 48 eyes (79%) had an UCIVA of 0.0 logMAR.
CONCLUSION
The new multifocal IOL with a novel optical concept (5 foci) showed a wide range of visual acuity especially at intermediate and near distances in patients undergoing cataract surgery. Uncorrected visual acuity was excellent at all tested distances, monocularly and binocularly, spectacle independence and patient satisfaction were high.
Topics: Humans; Visual Acuity; Prospective Studies; Female; Male; Aged; Multifocal Intraocular Lenses; Middle Aged; Prosthesis Design; Presbyopia; Refraction, Ocular; Lens Implantation, Intraocular; Pseudophakia; Phacoemulsification; Cataract; Lenses, Intraocular; Aged, 80 and over; Follow-Up Studies
PubMed: 38918765
DOI: 10.1186/s12886-024-03521-7 -
Journal of Cataract and Refractive... Jun 2024To evaluate microbiological cultures of cataract surgical devices and products that were reused for multiple cases.
PURPOSE
To evaluate microbiological cultures of cataract surgical devices and products that were reused for multiple cases.
SETTING
Aravind Eye Hospital, Pondicherry, Tamil Nadu, India.
DESIGN
Prospective cohort study.
METHODS
Samples from multiple surgical instruments and products that were reused for consecutive cataract surgeries underwent bacterial and fungal cultures and were monitored alongside positive controls for 7 days. This included instruments that were processed using immediate use steam sterilization (IUSS) between cases (e.g., surgical cannulas, syringes, phacoemulsification and coaxial/bimanual irrigation and aspiration (IA) tips, phacoemulsification and IA sleeves) (Group 1), instruments that were used without sterilization between cases (e.g., phacoemulsification tubing/handpieces, coaxial IA handpieces) (Group 2) and the residual (unused) fluid from balanced salt solution bags after being used for multiple patients (Group 3).
RESULTS
3,333 discrete samples were collected from all 3 product groups that were reused across multiple patients. In all collected samples, no bacterial or fungal growth was observed. Of the 3,241 cataract surgeries that utilized reused and IUSS-sterilized instruments alongside instrument sets cultured on the same day and balanced salt solution bags shared across multiple patients, no eyes developed endophthalmitis over a 6-week follow-up period.
CONCLUSION
Bacterial or fungal growth was not found in extensive microbiological cultures of IUSS-sterilized ophthalmic surgical instruments, and cataract surgical products that were reused in multiple patients. This microbiological data complements clinical endophthalmitis data from 2 million consecutive cases at the Aravind Eye Hospital, suggesting that their instrument and surgical supply processing practices may allow for safe and sustainable ophthalmic care.
PubMed: 38915155
DOI: 10.1097/j.jcrs.0000000000001509 -
International Journal of Retina and... Jun 2024The EVA Nexus system offers several technical improvements over its predecessor. The newly designed Aveta cannula system for vitrectomy surgery avoids the need for...
BACKGROUND
The EVA Nexus system offers several technical improvements over its predecessor. The newly designed Aveta cannula system for vitrectomy surgery avoids the need for removal of the valve from the infusion cannula. The chamfered leading edge of the cannula also reduces the insertion force needed. The new EquiPhaco needles in combination with SmartIOP provide excellent anterior chamber stability during phaco-emulsification surgery, enabling to work at lower infusion pressures, and the multiburst phaco mode allows easier removal of hard cataracts. The system offers a secondary active infusion line for independent control of pressure to the anterior and posterior chambers, monitoring of flow rate/reflux and warning of infusion bottle emptying. This study evaluated whether these technical improvements result in improved surgical safety.
METHODS
In total, 250 eyes that underwent vitrectomy (53%) or phaco-vitrectomy (47%) using the EVA Nexus system were prospectively included. The occurrence of intraoperative adverse events was compared to that of historically operated eyes using the EVA system.
RESULTS
The average age of the patients was 63 years. A total of 33% of the patients were operated on for retinal detachment, 17% for macular pucker, 11% for treating floaters, 9% for removing silicone oil, 8% for macular hole repair and 22% for other diseases. In 75% of surgeries, 23 G instruments were used, and 27 G instruments were used in 25% of cases. Device issues that occurred included priming cycle issues (n = 4), eye pressure stability problems (n = 6) and vitrectome performance issues (n = 1), all of which in the first 100 patients who were included and were fixed with software updates. The frequency of surgical complications in the anterior segment was lower than that in the historically recorded surgical reports. Intraoperative events in the posterior segment included hemorrhage from retinal vessels, choroidal hematoma, iatrogenic retinal damage/tear, and subchoroidal infusion. Again, these events occurred rarely and less frequently than in the historical surgical reports.
CONCLUSIONS
The EVA Nexus provides a surgical platform that reduces the incidence of intraoperative adverse events and iatrogenic complications in both anterior and posterior segment surgery. This could increase surgical safety during cataract and vitrectomy surgery. TRIAL REGISTRATION NUMBER CLINICALTRIALS.GOV: : NCT05229094 Data 22/5/2021.
PubMed: 38915097
DOI: 10.1186/s40942-024-00563-3 -
International Ophthalmology Jun 2024Evaluation of anterior segment parameters using the Scheimpflug corneal topography 1 year after surgery in patients who underwent sutureless scleral fixation...
PURPOSE
Evaluation of anterior segment parameters using the Scheimpflug corneal topography 1 year after surgery in patients who underwent sutureless scleral fixation intraocular lens (SFIOL) implantation using the modified Yamane technique and retropupillary iris-claw intraocular lens (RPIOL) implantation.
METHODS
A total of 60 eyes from 57 patients who underwent sutureless SFIOL implantation and 57 eyes from 52 patients who underwent RPIOL implantation were included. Anterior chamber depth (ACD), anterior chamber angle (ACA), anterior chamber volume (ACV), anterior-posterior corneal astigmatism, and keratometric values were assessed using the Scheimpflug corneal topography (Pentacam HR, Germany).
RESULTS
There was no statistically significant difference in postoperative UCVA and BCVA between the sutureless SFIOL and the RPIOL group (p = 0.236, p = 0.293, respectively). While there was no statistically significant difference in postoperative IOP between the two groups (p = 0.223), a statistically significant decrease in IOP was observed in both groups (p < 0.001). While there was no statistical difference between the sutureless SFIOL group and the RPIOL group in terms of spherical value (p = 0.441) and spherical equivalence (p = 0.237), there was a statistically significant difference in cylindrical value (p < 0.001). While there was a statistical difference in anterior astigmatism (p < 0.001), there was no statistical difference in posterior astigmatism (p = 0.405). There was no statistical difference in terms of ACV, ACD, and ACA between the sutureless SFIOL and the RPIOL group (p = 0.812, p = 0.770, p = 0.401, respectively).
CONCLUSION
In this study, although there was a statistical difference in cylindrical value and anterior corneal astigmatism between the sutureless SFIOL and RPIOL groups, vision was not affected. According to this study, sutureless SFIOL and RPIOL are two successful methods in terms of visual acuity, anterior segment, and keratometry outcomes in aphakic patients after phacoemulsification.
Topics: Humans; Male; Female; Visual Acuity; Sclera; Lens Implantation, Intraocular; Middle Aged; Corneal Topography; Aged; Iris; Sutureless Surgical Procedures; Lenses, Intraocular; Retrospective Studies; Anterior Eye Segment; Adult; Treatment Outcome; Prosthesis Design; Aphakia, Postcataract; Follow-Up Studies; Aphakia
PubMed: 38914871
DOI: 10.1007/s10792-024-03187-6 -
Clinical Case Reports Jul 2024If an intraocular lens (IOL) is incorrectly labeled, problems can arise for surgeons and patients. Precise biometry is important to accurately determine the IOL power...
KEY CLINICAL MESSAGE
If an intraocular lens (IOL) is incorrectly labeled, problems can arise for surgeons and patients. Precise biometry is important to accurately determine the IOL power and prevent the implantation of the wrong IOL. Labeling and verifying IOLs with accuracy is crucial to ensuring the best possible results of cataract surgery.
ABSTRACT
Mislabeling of IOLs can cause unpredictable problems for surgeons. However, we can prevent incorrect IOLs by using precise biometry to determine accurate IOL power and safely implant the correct IOL. A 50-year-old female with no medical or ocular history came to our clinic complaining of decreased vision in both eyes that had been ongoing for several months. After being diagnosed with cataracts, primary angle closure suspect, and high hyperopia, the patient underwent phacoemulsification surgery. A posterior chamber IOL was implanted, and visco-goniosynechialysis was performed. During follow-up appointments, it was discovered that the patient had an uncorrected visual acuity of 20/50 in her right eye, which was corrected to 20/20 with a + 7.00 D lens. Upon further evaluation, it was determined that the source of the error was due to a manufacturing mislabeling of the IOL power. The patient then underwent successful IOL exchange surgery, and her best-corrected visual acuity became 20/20 with no significant refraction. This case highlights an uncommon source of refractive surprise after phacoemulsification surgery, successfully managed with IOL exchange surgery using the same IOL power from a different brand.
PubMed: 38911921
DOI: 10.1002/ccr3.9104 -
International Ophthalmology Jun 2024To explore ocular characteristics of patients with cataracts after renal transplantation and analyze the results of phacoemulsification combined with intraocular lens...
PURPOSE
To explore ocular characteristics of patients with cataracts after renal transplantation and analyze the results of phacoemulsification combined with intraocular lens (IOL) implantation.
METHODS
Patients with cataracts after renal transplantation and control patients who underwent phacoemulsification combined with IOL implantation were enrolled. All patients underwent phacoemulsification combined with IOL implantation. Visual acuity, intraocular pressure, type of lens opacity, corneal endothelial cell density, and ocular biological parameters were evaluated before surgery. Visual prognosis, dry eye, and postoperative complications were monitored for 6 months after phacoemulsification.
RESULTS
We analyzed 25 eyes of 16 patients after renal transplantation and 30 eyes of 21 control patients. The most common type of cataract of renal transplantation group was posterior subcapsular, while the most common type of cataract of control group was cortical. Significant differences in corneal astigmatism, white-to-white ratio, and keratometry values were observed between the groups. The postoperative visual acuity of both groups significantly improved following surgery. Postoperative complications, such as the degree of anterior and posterior capsule opacification and the incidence of a requirement of neodymium-doped yttrium aluminum garnet laser capsulotomy, were significantly lower in the renal transplantation group. Moreover, secondary glaucoma occurred in two eyes in the renal transplantation group.
CONCLUSION
This study showed that cataracts after renal transplantation were mostly posterior subcapsular. Postoperative visual acuity recovered well in most patients, with reduced incidence of postoperative complications. This study suggested that phacoemulsification combined with IOL implantation was safe and effective, providing a reference for multi-focal IOL implantation in kidney transplant recipients.
Topics: Humans; Phacoemulsification; Male; Female; Cataract; Visual Acuity; Kidney Transplantation; Middle Aged; Postoperative Complications; Adult; Intraocular Pressure; Lens Implantation, Intraocular; Retrospective Studies; Treatment Outcome; Follow-Up Studies; Aged
PubMed: 38909160
DOI: 10.1007/s10792-024-03161-2