JMIR Biomedical Engineering Mar 2024Pulse oximeters work within the red-infrared wavelengths. Therefore, these oximeters produce erratic results in dark-skinned subjects and in subjects with cold...
BACKGROUND
Pulse oximeters work within the red-infrared wavelengths. Therefore, these oximeters produce erratic results in dark-skinned subjects and in subjects with cold extremities. Pulse oximetry is routinely performed in patients with fever; however, an elevation in body temperature decreases the affinity of hemoglobin for oxygen, causing a drop in oxygen saturation or oxyhemoglobin concentrations.
OBJECTIVE
We aimed to determine whether our new investigational device, the Shani device or SH1 (US Patent 11191460), detects a drop in oxygen saturation or a decrease in oxyhemoglobin concentrations.
METHODS
An observational study (phase 1) was performed in two separate groups to validate measurements of hemoglobin and oxygen concentrations, including 39 participants recruited among current university students and staff aged 20-40 years. All volunteers completed baseline readings using the SH1 device and the commercially available Food and Drug Administration-approved pulse oximeter Masimo. SH1 uses two light-emitting diodes in which the emitted wavelengths match with absorption peaks of oxyhemoglobin (hemoglobin combined with oxygen) and deoxyhemoglobin (hemoglobin without oxygen or reduced hemoglobin). Total hemoglobin was calculated as the sum of oxyhemoglobin and deoxyhemoglobin. Subsequently, 16 subjects completed the "heat jacket study" and the others completed the "blood donation study." Masimo was consistently used on the finger for comparison. The melanin level was accounted for using the von Luschan skin color scale (VLS) and a specifically designed algorithm. We here focus on the results of the heat jacket study, in which the subject wore a double-layered heated jacket and pair of trousers including a network of polythene tubules along with an inlet and outlet. Warm water was circulated to increase the body temperature by 0.5-0.8 °C above the baseline body temperature. We expected a slight drop in oxyhemoglobin concentrations in the heating phase at the tissue level.
RESULTS
The mean age of the participants was 24.1 (SD 0.8) years. The skin tone varied from 12 to 36 on the VLS, representing a uniform distribution with one-third of the participants having fair skin, brown skin, and dark skin, respectively. Using a specific algorithm and software, the reflection ratio for oxyhemoglobin was displayed on the screen of the device along with direct hemoglobin values. The SH1 device picked up more minor changes in oxyhemoglobin levels after a change in body temperature compared to the pulse oximeter, with a maximum drop in oxyhemoglobin concentration detected of 6.5% and 2.54%, respectively.
CONCLUSIONS
Our new investigational device SH1 measures oxygen saturation at the tissue level by reflectance spectroscopy using green wavelengths. This device fared well regardless of skin color. This device can thus eliminate racial disparity in these key biomarker assessments. Moreover, since the light is shone on the wrist, SH1 can be readily miniaturized into a wearable device.
PubMed: 38875701
DOI: 10.2196/46974 -
Archives de Pediatrie : Organe Officiel... Jun 2024Periodic breathing (PB)-related intermittent hypoxia can have long-lasting deleterious consequences in preterm infants. Olfactory stimulation using vanilla odor is...
BACKGROUND
Periodic breathing (PB)-related intermittent hypoxia can have long-lasting deleterious consequences in preterm infants. Olfactory stimulation using vanilla odor is beneficial for apnea of prematurity in the first postnatal days/weeks. We aimed to determine for the first time whether vanilla odor can also decrease PB-related intermittent hypoxia.
METHOD
This pilot study was a balanced crossover clinical trial including 27 premature infants born between 30 and 33 weeks of gestation. We performed 12-h recordings on two nights separated by a 24-h period. All infants were randomly exposed to vanilla odor on the first or second study night. The primary outcome was the desaturation index, defined as the number per hour of pulse oximetry (SpO) values <90 % for at least 5 s, together with a drop of ≥5 % from the preceding value. Univariate mixed linear models were used for the statistical analysis.
RESULTS
Overall, exposure to vanilla odor did not significantly decrease the desaturation index (52 ± 22 events/h [mean ± SD] on the intervention night vs. 57 ± 26, p = 0.2); furthermore, it did not significantly alter any secondary outcome. In a preliminary post hoc subgroup analysis, however, the effect of vanilla odor was statistically significant in infants with a desaturation index of ≥70/h (from 86 ± 12 to 65 ± 23, p = 0.04).
CONCLUSION
In this pilot study, vanilla odor overall did not decrease PB-related intermittent hypoxia in infants born at 30-33 weeks of gestation, which is when they are close to term. Preliminary results suggesting a beneficial effect in infants with the highest desaturation index, however, justify further studies in the presence of PB-related intermittent hypoxia as well as in infants born more prematurely.
PubMed: 38871544
DOI: 10.1016/j.arcped.2024.03.002 -
Journal of Clinical Anesthesia Jun 2024Elderly patients undergoing pathophysiological changes necessitate clinical tools for cerebral monitoring. This prospective randomized controlled study aimed to explore...
STUDY OBJECTIVE
Elderly patients undergoing pathophysiological changes necessitate clinical tools for cerebral monitoring. This prospective randomized controlled study aimed to explore how cerebral monitoring using ΔoHbi, ΔHHbi, and ΔcHbi manifests in elderly patients under either propofol or sevoflurane anesthesia.
DESIGN
Single-center, prospective, randomization.
SETTING
A single tertiary hospital (Eunpyeong St. Mary's Hospital, College of Medicine, The Catholic University of Korea).
PATIENTS
Enrolled 100 patients scheduled for urologic surgery under general anesthesia. Inclusion criteria were (a) age 70-80 years, (b) American Society of Anesthesiologists (ASA) physical status I-II.
INTERVENTION
Patients were double-blind randomized to receive propofol-based or sevoflurane anesthesia. Cerebral oximetry-related parameters were measured at 5, 10, 15, 20, and 30 min in a setting devoid of surgery-related factors.
MEASUREMENTS
The primary outcome focused on the ΔoHbi pattern in the left and right sides within the propofol and sevoflurane groups.
MAIN RESULTS
We analyzed 100 patients, 50 patients in each group. In the propofol group, the left ΔoHbi decreased from 1.4 (3.7) at 5 min to -0.1 (1.8) at 30 min (P < 0.0001), and the right ΔoHbi decreased from 2.9 (4.2) at 5 min to -0.06 (2.3) at 30 min (P < 0.0001). In the sevoflurane group, the left ΔoHbi decreased from 1.1 (3.4) at 5 min to -1.4 (4.4) at 30 min (P < 0.0001), and the right ΔoHbi decreased from 2.0 (3.2) at 5 min to -1.2 (3.9) at 30 min (P < 0.0001). There were no significant differences between the two groups. ΔHHbi did not exhibit significant changes after an initial decrease at 5 min and showed no significant differences between the two groups.
CONCLUSIONS
In cerebral oximetry, ΔoHbi and ΔHHbi could emerge as a valuable approach for discerning changes in the underlying baseline status of the brain in elderly patients during anesthesia.
PubMed: 38870700
DOI: 10.1016/j.jclinane.2024.111519 -
Neurology Jul 2024Generalized convulsive seizures (GCSs) are the main risk factor of sudden unexpected death in epilepsy (SUDEP), which is likely due to peri-ictal cardiorespiratory... (Observational Study)
Observational Study
BACKGROUND AND OBJECTIVES
Generalized convulsive seizures (GCSs) are the main risk factor of sudden unexpected death in epilepsy (SUDEP), which is likely due to peri-ictal cardiorespiratory dysfunction. The incidence of GCS-induced cardiac arrhythmias, their relationship to seizure severity markers, and their role in SUDEP physiopathology are unknown. The aim of this study was to analyze the incidence of seizure-induced cardiac arrhythmias, their association with electroclinical features and seizure severity biomarkers, as well as their specific occurrences in SUDEP cases.
METHODS
This is an observational, prospective, multicenter study of patients with epilepsy aged 18 years and older with recorded GCS during inpatient video-EEG monitoring for epilepsy evaluation. Exclusion criteria were status epilepticus and an obscured video recording. We analyzed semiologic and cardiorespiratory features through video-EEG (VEEG), electrocardiogram, thoracoabdominal bands, and pulse oximetry. We investigated the presence of bradycardia, asystole, supraventricular tachyarrhythmias (SVTs), premature atrial beats, premature ventricular beats, nonsustained ventricular tachycardia (NSVT), atrial fibrillation (Afib), ventricular fibrillation (VF), atrioventricular block (AVB), exaggerated sinus arrhythmia (ESA), and exaggerated sinus arrhythmia with bradycardia (ESAWB). A board-certified cardiac electrophysiologist diagnosed and classified the arrhythmia types. Bradycardia, asystole, SVT, NSVT, Afib, VF, AVB, and ESAWB were classified as arrhythmias of interest because these were of SUDEP pathophysiology value. The main outcome was the occurrence of seizure-induced arrhythmias of interest during inpatient VEEG monitoring. Moreover, yearly follow-up was conducted to identify SUDEP cases. Binary logistic generalized estimating equations were used to determine clinical-demographic and peri-ictal variables that were predictive of the presence of seizure-induced arrhythmias of interest. The -score test for 2 population proportions was used to test whether the proportion of seizures and patients with postconvulsive ESAWB or bradycardia differed between SUDEP cases and survivors.
RESULTS
This study includes data from 249 patients (mean age 37.2 ± 23.5 years, 55% female) who had 455 seizures. The most common arrhythmia was ESA, with an incidence of 137 of 382 seizures (35.9%) (106/224 patients [47.3%]). There were 50 of 352 seizure-induced arrhythmias of interest (14.2%) in 41 of 204 patients (20.1%). ESAWB was the commonest in 22 of 394 seizures (5.6%) (18/225 patients [8%]), followed by SVT in 18 of 397 seizures (4.5%) (17/228 patients [7.5%]). During follow-up (48.36 ± 31.34 months), 8 SUDEPs occurred. Seizure-induced bradycardia (3.8% vs 12.5%, = -16.66, < 0.01) and ESAWB (6.6% vs 25%; = -3.03, < 0.01) were over-represented in patients who later died of SUDEP. There was no association between arrhythmias of interest and seizure severity biomarkers ( > 0.05).
DISCUSSION
Markers of seizure severity are not related to seizure-induced arrhythmias of interest, suggesting that other factors such as occult cardiac abnormalities may be relevant for their occurrence. Seizure-induced ESAWB and bradycardia were more frequent in SUDEP cases, although this observation was based on a very limited number of SUDEP patients. Further case-control studies are needed to evaluate the yield of arrhythmias of interest along with respiratory changes as potential SUDEP biomarkers.
Topics: Humans; Female; Male; Adult; Arrhythmias, Cardiac; Incidence; Middle Aged; Prospective Studies; Electroencephalography; Sudden Unexpected Death in Epilepsy; Seizures; Epilepsy, Generalized; Aged; Young Adult; Electrocardiography; Adolescent
PubMed: 38870452
DOI: 10.1212/WNL.0000000000209501 -
Tuberkuloz Ve Toraks Jun 2024Partial carbondioxide pressure of the arterial blood (PaCO2) is used to evaluate alveolar ventilation. Transcutaneous carbon dioxide pressure (TcCO2) monitoring has been...
INTRODUCTION
Partial carbondioxide pressure of the arterial blood (PaCO2) is used to evaluate alveolar ventilation. Transcutaneous carbon dioxide pressure (TcCO2) monitoring has been developed as a non-invasive (NIV) alternative to arterial blood gas analysis (ABG). Studies have shown that decreased tissue perfusion leads to increased carbondioxide (CO2). The use of transcutaneous capnometry may be unreliable in patients with perfusion abnormalities. In this study, we aimed to evaluate the relation between TcCO2-PaCO2 and lactate level which is recognized as a marker of hypoperfusion.
MATERIALS AND METHODS
In this prospective cohort study in critical care patients with hypercapnic respiratory failure (PaCO2 ≥45 mmHg) who received NIV between April 2019 and January 2020 in the intensive care unit were enrolled in the study. Patients' simultaneously measured TcCO2 and PaCO2 values of hypercapnic patients were recorded. Each paired measurement was categorized into two groups; normal lactate (<2 mmol/L) and increased lactate (≥2 mmol/L).
RESULT
A total of 116 paired TcCO2 and PaCO2 measurements of 29 patients were recorded. Bland-Altman analysis showed the mean bias between the TcCO2 and PaCO2 and 95% limits of agreement (LOA) in all measurements (1.75 mmHg 95% LOA -3.67 to 7.17); in the normal lactate group (0.66 mmHg 95% LOA -1.71 to 3.03); and in the increased lactate group (5.17 mmHg 95% LOA -1.63 to 11.97). The analysis showed a correlation between lactate level and the difference between TcCO2 and PaCO2 (r= 0.79, p< 0.001) and a negative correlation between mean blood pressure and the difference between TcCO2 and PaCO2 (r= -0.54, p= 0.001). Multiple regression analysis results showed that lactate level was independently associated with increased differences between TcCO2 and PaCO2 (Beta= 0.875, p< 0.001).
CONCLUSIONS
TcCO2 monitoring may not be reliable in patients with increased lactate levels. TcCO2 levels should be checked by ABG analysis in these patients.
Topics: Humans; Carbon Dioxide; Prospective Studies; Male; Female; Blood Gas Monitoring, Transcutaneous; Lactic Acid; Middle Aged; Aged; Blood Gas Analysis; Hypercapnia; Respiratory Insufficiency; Noninvasive Ventilation; Critical Care
PubMed: 38869204
DOI: 10.5578/tt.202402920 -
Magnetic Resonance in Medical Sciences... Jun 2024The metabolic rate of oxygen (MRO) is fundamental to tissue metabolism. Determination of MRO demands knowledge of the arterio-venous difference in hemoglobin-bound...
The metabolic rate of oxygen (MRO) is fundamental to tissue metabolism. Determination of MRO demands knowledge of the arterio-venous difference in hemoglobin-bound oxygen concentration, typically expressed as oxygen extraction fraction (OEF), and blood flow rate (BFR). MRI is uniquely suited for measurement of both these quantities, yielding MRO in absolute physiologic units of µmol O min/100 g tissue. Two approaches are discussed, both relying on hemoglobin magnetism. Emphasis will be on cerebral oxygen metabolism expressed in terms of the cerebral MRO (CMRO), but translation of the relevant technologies to other organs, including kidney and placenta will be touched upon as well. The first class of methods exploits the blood's bulk magnetic susceptibility, which can be derived from field maps. The second is based on measurement of blood water T, which is modulated by diffusion and exchange in the local-induced fields within and surrounding erythrocytes. Some whole-organ methods achieve temporal resolution adequate to permit time-series studies of brain energetics, for instance, during sleep in the scanner with concurrent electroencephalogram (EEG) sleep stage monitoring. Conversely, trading temporal for spatial resolution has led to techniques for spatially resolved approaches based on quantitative blood oxygen level dependent (BOLD) or calibrated BOLD models, allowing regional assessment of vascular-metabolic parameters, both also exploiting deoxyhemoglobin paramagnetism like their whole-organ counterparts.
PubMed: 38866481
DOI: 10.2463/mrms.rev.2024-0028 -
European Journal of Pediatrics Jun 2024Cerebral monitoring during immediate fetal-to-neonatal transition is of increasing interest. The cerebral fractional tissue oxygen extraction (cFTOE) is a useful... (Review)
Review
UNLABELLED
Cerebral monitoring during immediate fetal-to-neonatal transition is of increasing interest. The cerebral fractional tissue oxygen extraction (cFTOE) is a useful parameter to gain insight in the balance between tissue oxygen delivery and consumption during this complex process. The aim of this study was to review the literature on cFTOE during the first 15 min immediately after birth. A systematic qualitative literature research was last performed on 23 November 2023 of PubMed and EMBASE with the following search terms: neonate, infant, newborn, transition, after birth, delivery room, NIRS, near-infrared spectroscopy, spectroscopy, cFTOE, cerebral fractional tissue oxygenation extraction, cerebral oxygenation, and fractional oxygen extraction. Additional published reports were identified through a manual search of references in retrieved articles and in review articles. The methodological quality of the included studies was assessed by predefined quality criteria. Only human studies with data of cFTOE in the first 15 min after birth were included. Accordingly, exclusion criteria were defined as no measurement of cFTOE or no measurement within the first 15 min after birth. Across all studies, a total of 3566 infants (2423 term, 1143 preterm infants) were analysed. Twenty-five studies were identified describing cFTOE within the first 15 min after birth. Four studies established reference ranges for cFTOE and another four studies focused on the effect of pre-/perinatal circumstances on cFTOE in the first 15 min after birth. Six studies investigated the course of cFTOE after transition in infants without complications. Eleven studies analysed different potentially influencing parameters on cFTOE during transition.
CONCLUSION
This systematic review provides a comprehensive insight on cFTOE during uncomplicated transition as well as the influence of perinatal circumstances, respiratory, haemodynamic, neurological, and laboratory parameters in preterm and term infants.
WHAT IS KNOWN
• The NIRS-measured cerebral fractional tissue oxygen extraction (cFTOE) is a useful parameter to estimate the balance between oxygen delivery and consumption. • During normal transition, the cFTOE decreases in the first minutes after birth and then remains at a stable plateau.
WHAT IS NEW
• The cFTOE is a promising parameter that gives additional information on cerebral oxygenation and perfusion in preterm and term infants. • Several hemodynamic, metabolic, respiratory, and perinatal factors are identified, influencing the oxygen extraction of the newborn's brain after birth.
PubMed: 38861023
DOI: 10.1007/s00431-024-05631-2 -
Indian Pediatrics Jun 2024Pre discharge pulse oximetry screening (POS) is recommended to pick up critical congenital heart diseases in apparently well neonates. However, it is possible that cases...
Pre discharge pulse oximetry screening (POS) is recommended to pick up critical congenital heart diseases in apparently well neonates. However, it is possible that cases may be missed during the early POS in the presence of delayed closure of the ductus arteriosus. Repeat POS in the second week of life was found to be helpful and feasible for early detection of pathological states causing hypoxemia in seemingly well neonates. Studies with larger sample size are recommended to establish the role of an additional POS in the second week for enhanced CCHD detection.
PubMed: 38859649
DOI: No ID Found -
Zhonghua Jie He He Hu Xi Za Zhi =... Jun 2024To evaluate the application value of portable pulse oximeter in adult obstructive sleep apnea (OSA). This study prospectively enrolled adult patients who underwent...
To evaluate the application value of portable pulse oximeter in adult obstructive sleep apnea (OSA). This study prospectively enrolled adult patients who underwent polysomnography (PSG) due to snoring at the Respiratory and Sleep Medicine Department of Peking University People's Hospital from July 2022 to July 2023. During PSG monitoring, CS-WOxi was continuously used to monitor blood oxygen levels. The consistency between 3% oxygen desaturation index (ODI) measured by portable pulse oximeter and ODI of polysomnography was evaluated using difference test, Pearson's correlation coefficient, and Bland-altman method. Receiver operating characteristic curve was used to determine the optimal threshold for diagnosing OSA. A total of 184 subjects were included, including 121 males (65.8%) and 63 females (34.2%). The mean age was 46.0 (34.3, 59.0) years, body mass index was 26.0 (23.3, 29.6) kg/m², and the apnea-hypopnea index was 18.2 (5.8, 40.8) events/h. There was a significant difference between CS-ODI and PSG-ODI [17.1(6.2, 42.7) 14.0(2.9, 32.6), <0.001], and the Pearson correlation coefficient was 0.93 (<0.001). There was a good correlation between CS-ODI and PSG-AHI (=0.92, <0.001). Bland-Altman consistency test showed that the average difference between the two was 0.7 events/h, and the 95% consistency limit was (-17.9, 19.3 events/h). When the CS-ODI≥5 events/h was used to identify OSA, the sensitivity was 94.4%, the specificity was 80.0%, and the accuracy was 91.3%. When PSG-AHI≥5 events/h was used as the diagnostic criteria, the area under the receiver operating characteristic curve was 0.933. Portable pulse oximeter can monitor pulse oxygen saturation accurately and has good sensitivity and specificity for OSA high-risk patients, and is a reliable tool for OSA screening.
Topics: Humans; Sleep Apnea, Obstructive; Oximetry; Female; Male; Middle Aged; Adult; Polysomnography; Prospective Studies; ROC Curve; Sensitivity and Specificity; Body Mass Index; Oxygen
PubMed: 38858202
DOI: 10.3760/cma.j.cn112147-20231113-00306 -
Lung Jun 2024Skin pigmentation influences peripheral oxygen saturation (SpO) compared to arterial saturation of oxygen (SaO). Occult hypoxemia (SaO ≤ 88% with SpO ≥ 92%)...
Characterizing the Racial Discrepancy in Hypoxemia Detection in Venovenous Extracorporeal Membrane Oxygenation: An Extracorporeal Life Support Organization Registry Analysis.
PURPOSE
Skin pigmentation influences peripheral oxygen saturation (SpO) compared to arterial saturation of oxygen (SaO). Occult hypoxemia (SaO ≤ 88% with SpO ≥ 92%) is associated with increased in-hospital mortality in venovenous-extracorporeal membrane oxygenation (VV-ECMO) patients. We hypothesized VV-ECMO cannulation, in addition to race/ethnicity, accentuates the SpO-SaO discrepancy due to significant hemolysis.
METHODS
Adults (≥ 18 years) supported with VV-ECMO with concurrently measured SpO and SaO measurements from over 500 centers in the Extracorporeal Life Support Organization Registry (1/2018-5/2023) were included. Multivariable logistic regressions were performed to examine whether race/ethnicity was associated with occult hypoxemia in pre-ECMO and on-ECMO SpO-SaO calculations.
RESULTS
Of 13,171 VV-ECMO patients, there were 7772 (59%) White, 2114 (16%) Hispanic, 1777 (14%) Black, and 1508 (11%) Asian patients. The frequency of on-ECMO occult hypoxemia was 2.0% (N = 233). Occult hypoxemia was more common in Black and Hispanic patients versus White patients (3.1% versus 1.7%, P < 0.001 and 2.5% versus 1.7%, P = 0.025, respectively). In multivariable logistic regression, Black patients were at higher risk of pre-ECMO occult hypoxemia versus White patients (adjusted odds ratio [aOR] = 1.55, 95% confidence interval [CI] = 1.18-2.02, P = 0.001). For on-ECMO occult hypoxemia, Black patients (aOR = 1.79, 95% CI = 1.16-2.75, P = 0.008) and Hispanic patients (aOR = 1.71, 95% CI = 1.15-2.55, P = 0.008) had higher risk versus White patients. Higher pump flow rates (aOR = 1.29, 95% CI = 1.08-1.55, P = 0.005) and on-ECMO 24-h lactate (aOR = 1.06, 95% CI = 1.03-1.10, P < 0.001) significantly increased the risk of on-ECMO occult hypoxemia.
CONCLUSION
SaO should be carefully monitored if using SpO during ECMO support for Black and Hispanic patients especially for those with high pump flow and lactate values at risk for occult hypoxemia.
PubMed: 38856932
DOI: 10.1007/s00408-024-00711-4