Journal of Cataract and Refractive... May 2024To systematically compare the effectiveness of conventional phacoemulsification surgery (CPS) and low-energy femtosecond laser-assisted cataract surgery (FLACS) in...
TOPIC
To systematically compare the effectiveness of conventional phacoemulsification surgery (CPS) and low-energy femtosecond laser-assisted cataract surgery (FLACS) in patients with cataract.
CLINICAL RELEVANCE
Cataract surgery is a common procedure, and comparing different techniques such as CPS and low-energy FLACS is crucial for optimizing patient outcomes.
METHODS
The PubMed, Web of Science, MEDLINE, EMBASE, and Cochrane library databases were searched for clinical trials. Outcomes of procedure time, effective phacoemulsification time, balanced salt solution usage, cumulative dissipated energy, mean change of best-corrected visual acuity, endothelial cells reduction, central corneal thickness, and aqueous cytokine level were evaluated. The effect measures were weighted mean differences with 95% confidence interval. The protocol was registered at the Prospective Register for Systematic Reviews (PROSPERO, registration number CRD42023420173).
RESULTS
11 studies were included in the present meta-analysis, of which 1680 eyes were analyzed (637 eyes in the low-energy FLACS group, and 1043 eyes in the CPS group). Low-energy FLACS demonstrated significantly fewer reductions in endothelial cell count at six months (p<0.001) compared to CPS. It also exhibited a shorter effective phacoemulsification time (p<0.001) and less balanced salt solution usage (P<0.001). However, there were no differences in cumulative dissipated energy, corrected distance visual acuity, central corneal thickness changes, or aqueous cytokine levels between the two groups.
CONCLUSION
Both low-energy FLACS and CPS are effective in treating cataracts, but low-energy FLACS may offer advantages such as reduced phacoemulsification time and less endothelial cell loss.
PubMed: 38861345
DOI: 10.1097/j.jcrs.0000000000001501 -
International Ophthalmology Jun 2024To explore the application effect of cervical pillow in phacoemulsification surgery for age-related cataract patients. (Randomized Controlled Trial)
Randomized Controlled Trial
PURPOSE
To explore the application effect of cervical pillow in phacoemulsification surgery for age-related cataract patients.
METHODS
104 cases of age-related cataract patients admitted to our hospital in June 2023 were enrolled and divided into the control group (traditional supine position) and the experimental group (the cervical pillow supine position) by the digital parity method (52 cases per group). The two groups were evaluated for the discomfort score, the satisfaction of patients and doctors, the head displacement rate, the number of displacement, the operation time and the time of body position during the operation and after the operation.
RESULTS
There was no significant difference in the gender (P = 0.84), age (P = 0.86), course of disease (P = 0.82) and the time spent on position placement (P = 0.15) of the two groups. The patient in the experimental group had lower discomfort score (P = 0.0001), higher patients satisfaction (P = 0.0001) and higher doctors satisfaction (P = 0.0001) than patients in the control group. There was no significant difference between the experimental group and the control group in the proportion of intraoperative (P = 0.36) and postoperative pain (P = 0.65). Besides, the number of head transfers (P = 0.001), number of head shifts (P = 0.0001), the surgical time (P = 0.0001) and laparoscopic time (P = 0.0001) in the experimental group were significantly lower than those in the control group.
CONCLUSION
The additional cervical pillow for age-related cataract patients in the traditional supine position during the operation will not increase the preparation time before the operation, but will help improve the patient satisfaction, improve the comfort and maintain a good position of the operative eye field, bringing obvious comfort and smooth operation for the surgeon in the operation, thus reducing the risk of the operation, shortening the operation time.
Topics: Humans; Female; Male; Phacoemulsification; Prospective Studies; Aged; Middle Aged; Patient Satisfaction; Cataract; Visual Acuity; Operative Time
PubMed: 38861013
DOI: 10.1007/s10792-024-03090-0 -
European Journal of Ophthalmology Jun 2024To compare the efficacy and safety of iStent versus Endocyclophotocoagulation (ECP) as an adjunct to cataract surgery by Phacoemulsification for treating glaucoma...
PURPOSE
To compare the efficacy and safety of iStent versus Endocyclophotocoagulation (ECP) as an adjunct to cataract surgery by Phacoemulsification for treating glaucoma patients in a tertiary eye center.
METHODS
Retrospective study of 67 eyes of 61 patients with glaucoma and cataract who underwent either phaco-ECP or phaco-iStent. Primary efficacy endpoint is the Intraocular pressure (IOP) reduction, while reduction of glaucoma medications is the secondary outcome. In addition to IOP and number of glaucoma medications; visual acuity, degree of disc cupping, safety profiles were all assessed at different intervals up to 12 months.
RESULTS
A total of 40 eyes underwent phaco-ECP, and 27 eyes underwent phaco-iStent. Both groups were associated with a significant reduction in the number of glaucoma medications; however phaco-iStent group achieved slightly lower IOP levels than the phaco-ECP group. Furthermore, iStent inject had better control of IOP at the last follow-up compared to first-generation stents. Moreover; 2 or more stents significantly reduced IOP than single stent ( = 0.009 vs. = 0.618, respectively). Phaco-iStent achieved a better reduction in the number of glaucoma medications for primary open-angle glaucoma ( = 0.007) compared to pseudoexfoliation glaucoma patients ( = 0.084). Complications were seen in 12 eyes (18%), of which five eyes in phaco-ECP (7.4%) and 7 eyes in phaco-iStent (10.4%), majority were mild and treated conservatively.
CONCLUSIONS
Both groups had equal efficacy in reducing the IOP. However, phaco-iStent seems superior in reducing the number of glaucoma medications after 1 year of follow-up compared to phaco-ECP, particularly when 2 or more stents are used. Both groups showed an overall good safety profile.
PubMed: 38860310
DOI: 10.1177/11206721241261418 -
Clinical Ophthalmology (Auckland, N.Z.) 2024To determine if the changes in stereoacuity and aniseikonia, following bilateral implantation of presbyopia correcting intraocular lenses could be predicted from...
Stereoacuity and Aniseikonia: Evaluation Before and After Bilateral Implantation of Three Types of Presbyopia-Correcting Intraocular Lenses in Uncomplicated Phacoemulsification with Due Consideration of Interocular Differences in Higher Order Aberrations, Axial Lengths, Refractive Errors, and...
PURPOSE
To determine if the changes in stereoacuity and aniseikonia, following bilateral implantation of presbyopia correcting intraocular lenses could be predicted from preoperative measurements of higher order aberrations (HOAs), axial lengths (AL), refractive errors (RE) and corrected visual acuities (CVAs).
PATIENTS AND METHODS
Stereoacuity (Randot tests, @6m & 40cm, in steps of 20 arcsecs") vertical and horizontal aniseikonia (Awaya test @6m, in steps of 1%) with best correction and HOAs (Shack-Hartmann aberrometer) were measured before, 3 and 6 months after uncomplicated bilateral phacoemulsification. Twenty patients (I) underwent a mix-and-match procedure (Tecnis MF, ZKB00 in one eye and ZLB00 in the other), 17 (II) were implanted with a trifocal (AT LISA 839 triMP) and 18 (III) with a one-piece diffractive (Synergy OU) intraocular lens. The resultant aniseikonia (A) of vertical and horizontal pairs of aniseikonia measurements was calculated using the Pythagorean theorem. Twenty untreated age/gender matched cases were recruited as controls (IV).
RESULTS
The key results (p < 0.001) were a) stereoacuity at distance (SAD) and near (SAN) improved, A reduced in groups I, II & III remaining unchanged in group IV; b) some significant intergroup differences in SAD, SAN & A were detected at postop; c) at 6 months postop, changes (Δ=pre- minus postoperative value) correlated with preoperative values (x). Linear regression revealed, I ΔSAD=0.66x-57.47 [0.832, ±66.4], ΔSAN=0.96x-34.59 [0.821, ±16.9], ΔA=0.93A-2.12 [0.795, ±1.4] II ΔSAD=0.79x-62.91 [0.916, ±38.1], ΔSAN=0.96x-31.49 [0.892, ±8.0], ΔA=0.91A-0.91 [0.839, ±1.3] III ΔSAD=0.67x-35.50 [0.991, ±23.7], ΔSAN=0.88x-38.51[0.988, ±10.6], ΔA=0.86A-0.96 [0.900, ±1.3]. Figures in parentheses are the corresponding and ±limits of agreement between actual and estimated values. Definitive overarching associations connecting interocular differences in HOAs, AL, RE, and CVAs with SAD, SAN and A were not found.
CONCLUSION
Changes in stereoacuity and aniseikonia can be predicted using preoperative values. ΔSAN can be predicted within ±1, and ΔA within ±2, scale divisions. In group III ΔSAD can be predicted within ±1, and in group I ±3, scale divisions.
PubMed: 38855012
DOI: 10.2147/OPTH.S459684 -
Clinical & Experimental Ophthalmology Jun 2024This systematic review and meta-analysis quantitatively examines the efficacy of angle-based minimally invasive glaucoma surgery (MIGS) in normal tension glaucoma (NTG).
BACKGROUND
This systematic review and meta-analysis quantitatively examines the efficacy of angle-based minimally invasive glaucoma surgery (MIGS) in normal tension glaucoma (NTG).
METHODS
A literature search was performed on Medline, Embase, PubMed, CINAHL and Cochrane Library from inception until 20 December 2022. Pilot, cohort, observational studies and randomised controlled trials including at least 5 subjects undergoing angle-based MIGS (trabecular-bypass devices, excisional trabeculotomy, goniotomy and ab-interno canaloplasty) for NTG, with or without cataract surgery, were included. Meta-analysis of continuous outcome using the meta routine in R version 2022.12.0+353 was performed to determine mean intraocular pressure (IOP) and anti-glaucoma medication (AGM) reduction post-operatively.
RESULTS
Of the 846 studies initially identified, 15 studies with a pooled total of 367 eyes which underwent combined phacoemulsification and angle-based MIGS were included for final meta-analysis. Outcomes of the iStent were reported in 5 studies, iStent inject in 7 studies, Hydrus Microstent in 1 study, Kahook Dual Blade in 3 studies, and Trabectome in 2 studies. There was significant reduction in both IOP and AGM post-operatively at 6 months (2.44 mmHg, 95%CI: 1.83-3.06; 1.21 AGM, 95%CI: 0.99-1.44), 12 months (2.28 mmHg, 95%CI: 1.71-2.84; 1.18 AGM, 95%CI: 0.90-1.47), 24 months (2.10 mmHg, 95%CI: 1.51-2.68; 1.26 AGM, 95%CI: 0.85-1.68) and 36 months (2.43 mmHg, 95%CI: 1.71-3.15, 0.87 AGM, 95%CI: 0.21-1.53) (all p < 0.05). Subgroup analysis on combined phacoemulsification-iStent inject surgery demonstrated a reduction in both IOP (2.31 mmHg, 95%CI: 1.07-3.56, p < 0.001) and AGM (1.07 AGM, 95%CI: 0.86-1.29, p < 0.001) at 12 months post-operatively.
CONCLUSIONS
Angle-based MIGS combined with phacoemulsification effectively reduces IOP and AGM in NTG eyes for up to 36 months after surgery.
PubMed: 38853535
DOI: 10.1111/ceo.14408 -
American Journal of Ophthalmology Jun 2024To compare intraindividual monocular visual performance of a monofocal EDOF and a trifocal EDOF IOL following bilateral cataract surgery.
PURPOSE
To compare intraindividual monocular visual performance of a monofocal EDOF and a trifocal EDOF IOL following bilateral cataract surgery.
DESIGN
Single-center, prospective, fellow-eye comparison clinical trial.
METHODS
All patients received uneventful bilateral cataract surgery with implantation of the monofocal EDOF Isopure IOL in the dominant eye and the trifocal EDOF FineVision Triumf IOL in the non-dominant eye. Intraindividual monocular comparison included visual acuity at various distances, defocus curves, internal total higher-order aberration (HOA), spherical aberration (SA) Z(4.0), IOL decentration and tilt. Additionally, subjective outcomes were evaluated using Halo and Glare simulation and the VF-7 questionnaire.
RESULTS
In total, 25 patients (50 eyes) were included. Intraindividual monocular BCDVA, DCIVA, and BCNVA were comparable (p> .05). However, monocular contrast acuity (p= .019), DCNVA (p< .001) and defocus curves at defocus levels of 0.0D (p= .005) and between -1.5 and -4.0D (p< .001) differed significantly. At 5mm, internal HOA and SA Z(4.0) were significantly different (p< .001) and comparable at 3mm pupil diameter, as were IOL tilt and decentration (p> .05).
CONCLUSION
In this combined fellow-eye comparison, near vision was significantly better with the trifocal EDOF IOL. The monofocal EDOF IOL demonstrated similar distance and intermediate visual performance as the trifocal EDOF IOL. Overall, low values of photic phenomena and visual impairment was observed.
PubMed: 38851443
DOI: 10.1016/j.ajo.2024.05.029 -
Graefe's Archive For Clinical and... Jun 2024To explore the molecular mechanism underlying the protective effect of hypothermic perfusion on the corneal endothelium during phacoemulsification.
AIM
To explore the molecular mechanism underlying the protective effect of hypothermic perfusion on the corneal endothelium during phacoemulsification.
METHODS
Phacoemulsification was performed on New Zealand white rabbits. Perfusate at different temperatures was used during the operation, and the aqueous humor was collected for proteomic sequencing after the operation. Corneal endothelial cell injury was simulated by a corneal endothelial cell oxygen-glucose deprivation/reoxygenation (OGD/R) model in vitro. Flow cytometry and evaluation of fluorescent LC3B puncta were used to detect apoptosis and autophagy, and western blotting was used to detect protein expression.
RESULTS
A total of 381 differentially expressed proteins were identified between the two groups. In vitro, 4 ℃ hypothermia significantly reduced apoptosis and promoted autophagy. Apoptosis increased after autophagy was inhibited by 3-Methyladenine (3-MA). Furthermore, adiponectin (ADIPOQ) knockdown inhibited phospho-AMPK and blocked the protective effect of hypothermia on corneal endothelial cells.
CONCLUSIONS
We investigated the differential expression of proteins between the hypothermia group and normothermia group by proteomics. Moreover, hypothermia-induced ADIPOQ can reduce apoptosis by promoting AMPK-mediated autophagy.
PubMed: 38850333
DOI: 10.1007/s00417-024-06542-6 -
Journal of Refractive Surgery... May 2024To assess the visual and refractive outcomes of patients implanted with a toric extended depth-of-focus (EDOF) intraocular lens (IOL) following cataract surgery.
PURPOSE
To assess the visual and refractive outcomes of patients implanted with a toric extended depth-of-focus (EDOF) intraocular lens (IOL) following cataract surgery.
METHODS
A total of 44 eyes implanted with the EDOF LuxSmart toric IOL were evaluated 4 to 6 months postoperatively. The main outcomes measurements evaluated were refractive error, rotational stability, distance, intermediate, and near visual acuities, defocus curve, photopic and mesopic contrast sensitivity, wavefront aberrations, and modulation transfer function, and the Catquest-9SF-questionnaire.
RESULTS
The mean postoperative spherical equivalent and cylinder were -0.02 ± 0.26 and -0.17 ± 0.29 diopters (D), respectively. A total of 90.45% and 100% of the eyes had a postoperative spherical equivalent within ±0.50 and ±1.00 D, respectively (this being 93.18% and 100% for the refractive cylinder). The mean rotational stability was 0.61 ± 1.61 degrees. The mean binocular corrected distance visual acuity (CDVA), corrected distance intermediate visual acuity (CDIVA), and corrected distance near visual acuity (CDNVA) were -0.02 ± 0.06, 0.07 ± 0.08, and 0.26 ± 0.09 logMAR, respectively. The CDVA was 20/25 or better in 95.45% of patients, CDIVA was 20/25 or better in 72.73%, and CDNVA was 20/40 or better in 72.73%. The defocus curve showed good visual acuity at distance and intermediate vergences. The contrast sensitivity and optical quality outcomes were good with mean higher order, spherical, and coma aberration values of 0.161 ± 0.155, -0.019 ± 0.048, and 0.080 ± 0.065 µm, respectively. A total of 90.9% of patients were either fairly satisfied or very satisfied with their vision after the surgery, and 77.7% of patients reported no difficulties when reading text in newspapers.
CONCLUSIONS
Implantation of the toric pure refractive EDOF technology IOL provides good refractive, optical, and visual quality at different distances, with high levels of patient satisfaction being reported. .
Topics: Humans; Visual Acuity; Lenses, Intraocular; Female; Male; Middle Aged; Aged; Contrast Sensitivity; Phacoemulsification; Lens Implantation, Intraocular; Pseudophakia; Refraction, Ocular; Corneal Wavefront Aberration; Depth Perception; Vision, Binocular; Prosthesis Design; Surveys and Questionnaires; Prospective Studies; Patient Satisfaction; Aged, 80 and over; Treatment Outcome
PubMed: 38848057
DOI: 10.3928/1081597X-20240501-02 -
Journal of Refractive Surgery... May 2024To assess the predictive accuracy of new-generation online intraocular lens (IOL) power formulas in eyes with previous myopic laser refractive surgery (LRS) and to...
PURPOSE
To assess the predictive accuracy of new-generation online intraocular lens (IOL) power formulas in eyes with previous myopic laser refractive surgery (LRS) and to evaluate the influence of corneal asphericity on the predictive accuracy.
METHODS
The authors retrospectively evaluated 52 patients (78 eyes) with a history of laser in situ keratomileusis (LASIK) or photorefractive keratectomy (PRK) who subsequently underwent cataract surgery. Refractive prediction errors were calculated for 12 no-history new online formulas: 8 formulas with post-LRS versions (Barrett True-K, EVO 2.0, Hoffer QST, and Pearl DGS) using keratometry and posterior/total keratometry measured by IOLMaster 700 and 4 formulas without post-LRS versions (Cooke K6 and Kane) using keratometry and total keratometry. The refractive prediction error, mean absolute error (MAE), and percentages of eyes with prediction errors of ±0.25, ±0.50, ±0.75, ±1.00, and ±1.50 diopters (D) were compared.
RESULTS
The MAEs of the 12 formulas were significantly different (F = 83.66, < .001). The MAEs ranged from 0.62 to 0.94 D and from 1.07 to 1.84 D in the formulas with and without post-LRS versions, respectively. The EVO formula produced the lowest MAE (0.60) and MedAE (0.47), followed by the Barrett True-K (0.69 and 0.50, respectively). Each percentage of eyes with refractive prediction error was also significantly different among the 12 formulas ( < .001).
CONCLUSIONS
The EVO and Barrett True-K formulas demonstrate comparable performance to the other existing formulas in eyes with a history of myopic LASIK/PRK. Surgeons should use these formulas with post-LRS versions and input keratometric values whenever possible. .
Topics: Humans; Retrospective Studies; Myopia; Lenses, Intraocular; Female; Male; Refraction, Ocular; Middle Aged; Photorefractive Keratectomy; Keratomileusis, Laser In Situ; Lens Implantation, Intraocular; Adult; Optics and Photonics; Visual Acuity; Lasers, Excimer; Cornea; Reproducibility of Results; Biometry; Phacoemulsification; Aged
PubMed: 38848053
DOI: 10.3928/1081597X-20240422-01 -
Journal of Refractive Surgery... May 2024To assess clinical outcomes after cataract surgery with bilateral implantation of a new bi-aspheric diffractive intraocular lens (IOL).
PURPOSE
To assess clinical outcomes after cataract surgery with bilateral implantation of a new bi-aspheric diffractive intraocular lens (IOL).
METHODS
Thirty patients underwent bilateral implantation of the Asqelio Trifocal Toric IOL (AST Products, Inc) and were evaluated 3 months postoperatively. Main outcomes included refractive error, photopic monocular and binocular uncorrected and corrected distance (UDVA, CDVA), intermediate (UIVA, CDIVA) at 60 cm, and near (UNVA, CDNVA) at 40 cm visual acuities. Mesopic monocular and binocular CDNVA were also measured. Defocus curves, binocular contrast sensitivity under photopic and mesopic conditions with and without glare and rotational stability were determined. Patients completed Catquest-9SF and visual symptoms questionnaires.
RESULTS
Average values of binocular photopic CDVA, CDIVA, and CDNVA, and mesopic CDNVA were -0.04 ± 0.06, 0.02 ± 0.08, 0.02 ± 0.07, and 0.22 ± 0.11 logMAR, respectively. All patients achieved cumulative CDVA ⩾ 20/25, and CDIVA and CDNVA of 20/32 or better. Binocular depth of focus was approximately 3.25 diopters (D). Mean postoperative spherical equivalent was -0.08 ± 0.26 D, with 95% of eyes within ±0.50 D. Mean postoperative refractive cylinder was -0.22 ± 0.27 D, with 91.67% of eyes within 0.50 D or less, respectively. IOL rotation averaged 0.25 ± 0.65 degrees, all eyes having rotation of less than 5 degrees. Contrast sensitivity was within or above normal levels under photopic and mesopic conditions, with or without glare, except for 12 cpd under mesopic conditions with glare. Questionnaire responses indicated 96.67% of patients were satisfied or very satisfied with postoperative vision, and 80.00% to 96.67% reported no difficulty in different daily activities.
CONCLUSIONS
The Asqelio Trifocal Toric IOL demonstrated favorable outcomes, providing excellent visual performance at all distances, precise refractive results, and remarkable rotational stability. Patients reported high satisfaction levels and minimal difficulty in daily activities. .
Topics: Humans; Visual Acuity; Vision, Binocular; Lens Implantation, Intraocular; Male; Female; Aged; Refraction, Ocular; Phacoemulsification; Pseudophakia; Middle Aged; Contrast Sensitivity; Multifocal Intraocular Lenses; Surveys and Questionnaires; Prosthesis Design; Patient Satisfaction; Prospective Studies; Aged, 80 and over; Lenses, Intraocular; Treatment Outcome
PubMed: 38848051
DOI: 10.3928/1081597X-20240506-01