The Cochrane Database of Systematic... Nov 2015The diagnosis of glaucoma is traditionally based on the finding of optic nerve head (ONH) damage assessed subjectively by ophthalmoscopy or photography or by... (Meta-Analysis)
Meta-Analysis Review
BACKGROUND
The diagnosis of glaucoma is traditionally based on the finding of optic nerve head (ONH) damage assessed subjectively by ophthalmoscopy or photography or by corresponding damage to the visual field assessed by automated perimetry, or both. Diagnostic assessments are usually required when ophthalmologists or primary eye care professionals find elevated intraocular pressure (IOP) or a suspect appearance of the ONH. Imaging tests such as confocal scanning laser ophthalmoscopy (HRT), optical coherence tomography (OCT) and scanning laser polarimetry (SLP, as used by the GDx instrument), provide an objective measure of the structural changes of retinal nerve fibre layer (RNFL) thickness and ONH parameters occurring in glaucoma.
OBJECTIVES
To determine the diagnostic accuracy of HRT, OCT and GDx for diagnosing manifest glaucoma by detecting ONH and RNFL damage.
SEARCH METHODS
We searched several databases for this review. The most recent searches were on 19 February 2015.
SELECTION CRITERIA
We included prospective and retrospective cohort studies and case-control studies that evaluated the accuracy of OCT, HRT or the GDx for diagnosing glaucoma. We excluded population-based screening studies, since we planned to consider studies on self-referred people or participants in whom a risk factor for glaucoma had already been identified in primary care, such as elevated IOP or a family history of glaucoma. We only considered recent commercial versions of the tests: spectral domain OCT, HRT III and GDx VCC or ECC.
DATA COLLECTION AND ANALYSIS
We adopted standard Cochrane methods. We fitted a hierarchical summary ROC (HSROC) model using the METADAS macro in SAS software. After studies were selected, we decided to use 2 x 2 data at 0.95 specificity or closer in meta-analyses, since this was the most commonly-reported level.
MAIN RESULTS
We included 106 studies in this review, which analysed 16,260 eyes (8353 cases, 7907 controls) in total. Forty studies (5574 participants) assessed GDx, 18 studies (3550 participants) HRT, and 63 (9390 participants) OCT, with 12 of these studies comparing two or three tests. Regarding study quality, a case-control design in 103 studies raised concerns as it can overestimate accuracy and reduce the applicability of the results to daily practice. Twenty-four studies were sponsored by the manufacturer, and in 15 the potential conflict of interest was unclear.Comparisons made within each test were more reliable than those between tests, as they were mostly based on direct comparisons within each study.The Nerve Fibre Indicator yielded the highest accuracy (estimate, 95% confidence interval (CI)) among GDx parameters (sensitivity: 0.67, 0.55 to 0.77; specificity: 0.94, 0.92 to 0.95). For HRT measures, the Vertical Cup/Disc (C/D) ratio (sensitivity: 0.72, 0.60 to 0.68; specificity: 0.94, 0.92 to 0.95) was no different from other parameters. With OCT, the accuracy of average RNFL retinal thickness was similar to the inferior sector (0.72, 0.65 to 0.77; specificity: 0.93, 0.92 to 0.95) and, in different studies, to the vertical C/D ratio.Comparing the parameters with the highest diagnostic odds ratio (DOR) for each device in a single HSROC model, the performance of GDx, HRT and OCT was remarkably similar. At a sensitivity of 0.70 and a high specificity close to 0.95 as in most of these studies, in 1000 people referred by primary eye care, of whom 200 have manifest glaucoma, such as in those who have already undergone some functional or anatomic testing by optometrists, the best measures of GDx, HRT and OCT would miss about 60 cases out of the 200 patients with glaucoma, and would incorrectly refer 50 out of 800 patients without glaucoma. If prevalence were 5%, e.g. such as in people referred only because of family history of glaucoma, the corresponding figures would be 15 patients missed out of 50 with manifest glaucoma, avoiding referral of about 890 out of 950 non-glaucomatous people.Heterogeneity investigations found that sensitivity estimate was higher for studies with more severe glaucoma, expressed as worse average mean deviation (MD): 0.79 (0.74 to 0.83) for MD < -6 db versus 0.64 (0.60 to 0.69) for MD ≥ -6 db, at a similar summary specificity (0.93, 95% CI 0.92 to 0.94 and, respectively, 0.94; 95% CI 0.93 to 0.95; P < 0.0001 for the difference in relative DOR).
AUTHORS' CONCLUSIONS
The accuracy of imaging tests for detecting manifest glaucoma was variable across studies, but overall similar for different devices. Accuracy may have been overestimated due to the case-control design, which is a serious limitation of the current evidence base.We recommend that further diagnostic accuracy studies are carried out on patients selected consecutively at a defined step of the clinical pathway, providing a description of risk factors leading to referral and bearing in mind the consequences of false positives and false negatives in the setting in which the diagnostic question is made. Future research should report accuracy for each threshold of these continuous measures, or publish raw data.
Topics: Diagnostic Errors; Glaucoma; Humans; Nerve Fibers; Odds Ratio; Ophthalmoscopy; Optic Disk; Prospective Studies; Retrospective Studies; Scanning Laser Polarimetry; Sensitivity and Specificity; Tomography, Optical Coherence; Visual Field Tests
PubMed: 26618332
DOI: 10.1002/14651858.CD008803.pub2 -
Health Technology Assessment... Oct 2007To assess whether open angle glaucoma (OAG) screening meets the UK National Screening Committee criteria, to compare screening strategies with case finding, to estimate... (Review)
Review
OBJECTIVES
To assess whether open angle glaucoma (OAG) screening meets the UK National Screening Committee criteria, to compare screening strategies with case finding, to estimate test parameters, to model estimates of cost and cost-effectiveness, and to identify areas for future research.
DATA SOURCES
Major electronic databases were searched up to December 2005.
REVIEW METHODS
Screening strategies were developed by wide consultation. Markov submodels were developed to represent screening strategies. Parameter estimates were determined by systematic reviews of epidemiology, economic evaluations of screening, and effectiveness (test accuracy, screening and treatment). Tailored highly sensitive electronic searches were undertaken.
RESULTS
Most potential screening tests reviewed had an estimated specificity of 85% or higher. No test was clearly most accurate, with only a few, heterogeneous studies for each test. No randomised controlled trials (RCTs) of screening were identified. Based on two treatment RCTs, early treatment reduces the risk of progression. Extrapolating from this, and assuming accelerated progression with advancing disease severity, without treatment the mean time to blindness in at least one eye was approximately 23 years, compared to 35 years with treatment. Prevalence would have to be about 3-4% in 40 year olds with a screening interval of 10 years to approach cost-effectiveness. It is predicted that screening might be cost-effective in a 50-year-old cohort at a prevalence of 4% with a 10-year screening interval. General population screening at any age, thus, appears not to be cost-effective. Selective screening of groups with higher prevalence (family history, black ethnicity) might be worthwhile, although this would only cover 6% of the population. Extension to include other at-risk cohorts (e.g. myopia and diabetes) would include 37% of the general population, but the prevalence is then too low for screening to be considered cost-effective. Screening using a test with initial automated classification followed by assessment by a specialised optometrist, for test positives, was more cost-effective than initial specialised optometric assessment. The cost-effectiveness of the screening programme was highly sensitive to the perspective on costs (NHS or societal). In the base-case model, the NHS costs of visual impairment were estimated as 669 pounds. If annual societal costs were 8800 pounds, then screening might be considered cost-effective for a 40-year-old cohort with 1% OAG prevalence assuming a willingness to pay of 30,000 pounds per quality-adjusted life-year. Of lesser importance were changes to estimates of attendance for sight tests, incidence of OAG, rate of progression and utility values for each stage of OAG severity. Cost-effectiveness was not particularly sensitive to the accuracy of screening tests within the ranges observed. However, a highly specific test is required to reduce large numbers of false-positive referrals. The findings that population screening is unlikely to be cost-effective are based on an economic model whose parameter estimates have considerable uncertainty. In particular, if rate of progression and/or costs of visual impairment are higher than estimated then screening could be cost-effective.
CONCLUSIONS
While population screening is not cost-effective, the targeted screening of high-risk groups may be. Procedures for identifying those at risk, for quality assuring the programme, as well as adequate service provision for those screened positive would all be needed. Glaucoma detection can be improved by increasing attendance for eye examination, and improving the performance of current testing by either refining practice or adding in a technology-based first assessment, the latter being the more cost-effective option. This has implications for any future organisational changes in community eye-care services. Further research should aim to develop and provide quality data to populate the economic model, by conducting a feasibility study of interventions to improve detection, by obtaining further data on costs of blindness, risk of progression and health outcomes, and by conducting an RCT of interventions to improve the uptake of glaucoma testing.
Topics: Age Factors; Cost-Benefit Analysis; Disease Progression; Glaucoma, Open-Angle; Humans; Sensitivity and Specificity; Technology Assessment, Biomedical; Time Factors; Treatment Outcome; Vision Screening
PubMed: 17927922
DOI: 10.3310/hta11410