The Spine Journal : Official Journal of... Jun 2024Thread shape is regarded as an important factor influencing the fixation strength and osseointegration of bone screws. However, commercial pedicle screws with a V-shaped...
BACKGROUND CONTEXT
Thread shape is regarded as an important factor influencing the fixation strength and osseointegration of bone screws. However, commercial pedicle screws with a V-shaped thread are prone to generating stress concentration at the bone-screw interface, thereby increasing the risk of screw loosening. Thus, modification of the pedicle-screw thread is imperative.
PURPOSE
This study aimed to investigate the fixation stability of pedicle screws with the new undercut thread design in comparison to pedicle screws with a V-shaped thread.
STUDY DESIGN
In vitro cadaveric biomechanical test and finite element analysis (FEA).
METHODS
Pedicle screws with the undercut thread (characterized by a flat crest feature and a tip-facing undercut feature) were custom-manufactured, whereas those with the V-shaped thread were procured from a commercial supplier. Fixation stability was assessed by the cyclic nonpullout compressive biomechanical testing on cadaveric female osteoporotic vertebrae. The vertical displacement and rotation angle of the 2 types of pedicle screws were calculated every 100 cycles to evaluate their resistance to migration and rotation. FEA was conducted to investigate the stress distribution and bone damage at the bone-screw interface for both types of pedicle screws.
RESULTS
Biomechanical testing revealed that the pedicle screws with the undercut thread exhibited significantly lower vertical displacement and rotation angles than the pedicle screws with the V-shape thread (P < 0.05). FEA results demonstrated a more uniform stress distribution in the bone surrounding the thread in the undercut design than in the V-shape design. Additionally, bone damage resulting from the pedicle screw was lower in the undercut design than in the V-shape design.
CONCLUSIONS
Pedicle screws with an undercut thread are less prone to migration and rotation and thus more stable in the bone than those with a V-shape thread.
CLINICAL SIGNIFICANCE
The undercut thread design may reduce the incidence of pedicle-screw loosening.
PubMed: 38843954
DOI: 10.1016/j.spinee.2024.05.006 -
Frontiers in Bioengineering and... 2024Various implant surface treatment methods have been developed to achieve good osseointegration in implant treatment. However, some cases remain impossible to treat with...
Various implant surface treatment methods have been developed to achieve good osseointegration in implant treatment. However, some cases remain impossible to treat with implants because osseointegration is not obtained after implantation, and the implants fail. Thus, this study focused on phosphorylated pullulan because of its adhesiveness to titanium (Ti) and bone, high biocompatibility, and early replacement with bone. In this study, the response of bone-related cells to phosphorylated pullulan was evaluated to develop a new surface treatment method. Saos-2 (human osteosarcoma-derived osteoblast-like cells), MC3T3-E1 (mouse calvaria-derived osteoblast-like cells), and RAW264.7 (mouse macrophage-like cells) were used. In evaluating cellular responses, phosphorylated pullulan was added to the culture medium, and cell proliferation and calcification induction tests were performed. The proliferation and calcification of cells on the surface of Ti disks coated with phosphorylated pullulan were also evaluated. In addition, bone morphogenetic protein-2 (BMP-2), an osteogenic factor, was used to evaluate the role of phosphorylated pullulan as a drug carrier in inducing calcification on Ti disks. Phosphorylated pullulan tended to promote the proliferation of osteoblast-like cells and the formation of calcification on Ti disks coated with phosphorylated pullulan. Ti disks coated with phosphorylated pullulan loaded with BMP-2 enhanced calcification. Phosphorylated pullulan inhibited osteoclast-like cell formation. These results are due to the properties of phosphorylated pullulan, such as adhesiveness to titanium and drug-loading function. Therefore, phosphorylated pullulan effectively promotes bone regeneration when coated on titanium implants and is useful for developing a new surface treatment method.
PubMed: 38840668
DOI: 10.3389/fbioe.2024.1378039 -
BMC Oral Health Jun 2024Peri-implantitis (PI) is a frequent inflammatory disorder characterised by progressive loss of the supporting bone. Not all patients with recognised risk factors develop...
BACKGROUND
Peri-implantitis (PI) is a frequent inflammatory disorder characterised by progressive loss of the supporting bone. Not all patients with recognised risk factors develop PI. The aim of this study is to evaluate the presence of single nucleotide polymorphisms (SNP) of inflammatory and bone metabolism related proteins in a population treated with dental implants from the Basque Country (Spain).
METHODS
We included 80 patients with diagnosis of PI and 81 patients without PI, 91 women and 70 men, with a mean age of 60.90 years. SNPs of BMP-4, BRINP3, CD14, FGF-3, FGF-10, GBP-1, IL-1α, IL-1β, IL-10, LTF, OPG and RANKL proteins were selected. We performed a univariate and bivariate analysis using IBM SPSS® v.28 statistical software.
RESULTS
Presence of SNPs GBP1 rs7911 (p = 0.041) and BRINP3 rs1935881 (p = 0.012) was significantly more common in patients with PI. Patients with PI who smoked (> 10 cig/day) showed a higher presence of OPG rs2073617 SNP (p = 0.034). Also, BMP-4 rs17563 (p = 0.018) and FGF-3 rs1893047 (p = 0.014) SNPs were more frequent in patients with PI and Type II diabetes mellitus.
CONCLUSIONS
Our findings suggest that PI could be favoured by an alteration in the osseointegration of dental implants, based on an abnormal immunological response to peri-implant infection in patients from the Basque Country (Spain).
Topics: Humans; Male; Female; Case-Control Studies; Middle Aged; Spain; Polymorphism, Single Nucleotide; Peri-Implantitis; Dental Implants; Osteoprotegerin; Aged; Bone Morphogenetic Protein 4; GTP-Binding Proteins; RANK Ligand; Interleukin-1alpha; Phosphoric Diester Hydrolases; Pyrophosphatases
PubMed: 38840172
DOI: 10.1186/s12903-024-04319-1 -
International Journal of Implant... Jun 2024This study evaluated the implant stability, volumetric changes, and patient-reported outcome measures (PROMs) of hydroxyapatite (HA) nano-coated sandblasted/acid-etched... (Randomized Controlled Trial)
Randomized Controlled Trial Comparative Study
A randomized controlled trial of immediate implant placement comparing hydroxyapatite nano-coated and uncoated sandblasted/acid-etched implants using a digital surgical guide.
PURPOSE
This study evaluated the implant stability, volumetric changes, and patient-reported outcome measures (PROMs) of hydroxyapatite (HA) nano-coated sandblasted/acid-etched (SLA) implants compared to uncoated SLA implants.
METHODS
Forty patients were recruited and randomly allocated to HA nano-coated SLA group (test, n = 20) and uncoated SLA group (control, n = 20) using single-blinded/block randomization. Implants were immediately placed in maxillary posterior region using a digital surgical guide. Insertion torque and implant stability quotient (ISQ) were measured at implant surgery and 1, 2, 3, and 4 months postoperatively. Intraoral scans, PROMs and soft tissue inflammation data were collected, and multivariable linear regression analysis of ISQ was performed.
RESULTS
In total, 48 implants (test; n = 24, control; n = 24) in 37 patients (test; n = 19, control; n = 18) were analyzed. Despite no significant between-group difference at surgery, the test group showed higher ISQ values than the control group at 2 (76.53 ± 4.17 vs. 71.32 ± 4.79, p < 0.01), 3 (77.45 ± 4.41 vs. 73.85 ± 4.69, p < 0.05), and 4 months (79.08 ± 2.96 vs. 73.43 ± 3.52, p < 0.0001) postoperatively. There were no significant differences in linear and volumetric changes, PROMs, and soft tissue inflammation analysis between two groups. The ISQ at implant surgery was influenced by age and diabetes mellitus (DM) at the implant level and DM and predicted total bone-to-implant contact area at the patient level.
CONCLUSION
HA nano-coated SLA implants promoted favorable immediate implants stability during early osseointegration phase compared to uncoated SLA implants, but displayed similar dimensional changes, PROMs, and soft tissue inflammation outcomes.
TRIAL REGISTRATION
Clinical Research Information Service (CRIS), KCT0006364. Registered 21 July 2021, https://cris.nih.go.kr/cris/search/detailSearch.do?seq=24221&search_page=L .
Topics: Humans; Durapatite; Male; Female; Middle Aged; Single-Blind Method; Dental Implants; Immediate Dental Implant Loading; Adult; Coated Materials, Biocompatible; Acid Etching, Dental; Aged; Patient Reported Outcome Measures; Osseointegration; Surface Properties
PubMed: 38839621
DOI: 10.1186/s40729-024-00549-8 -
International Journal of Nanomedicine 2024Atomic layer deposition (ALD) is a method that can deposit zirconia uniformly on an atomic basis. The effect of deposited zirconia on titanium implants using ALD was...
PURPOSE
Atomic layer deposition (ALD) is a method that can deposit zirconia uniformly on an atomic basis. The effect of deposited zirconia on titanium implants using ALD was evaluated in vivo.
METHODS
Machined titanium implants (MTIs) were used as the Control. MTIs treated by sandblasting with large grit and acid etching (SA) and MTIs deposited with zirconia using ALD are referred to as Groups S and Z, respectively. Twelve implants were prepared for each group. Six rabbits were used as experimental animals. To evaluate the osteogenesis and osteocyte aspects around the implants, radiological and histological analyses were performed. The bone-to-implant contact (BIC) ratio was measured and statistically analyzed to evaluate the osseointegration capabilities.
RESULTS
In the micro-CT analysis, more radiopaque bone tissues were observed around the implants in Groups S and Z. Histological observation found that Groups S and Z had more and denser mature bone tissues around the implants in the cortical bone area. Many new and mature bone tissues were also observed in the medullary cavity area. For the BIC ratio, Groups S and Z were significantly higher than the Control in the cortical bone area ( < 0.017), but there was no significant difference between Groups S and Z.
CONCLUSION
MTIs deposited with zirconia using ALD (Group Z) radiologically and histologically showed more mature bone formation and activated osteocytes compared with MTIs (Control). Group Z also had a significantly higher BIC ratio than the Control. Within the limitations of this study, depositing zirconia on the surface of MTIs using ALD can improve osseointegration in vivo.
Topics: Animals; Zirconium; Rabbits; Titanium; Osseointegration; Surface Properties; X-Ray Microtomography; Coated Materials, Biocompatible; Bone-Implant Interface; Osteogenesis; Dental Implants; Prostheses and Implants
PubMed: 38832337
DOI: 10.2147/IJN.S460392 -
The Japanese Dental Science Review Dec 2024Peri-implant diseases, characterized by inflammatory conditions affecting peri-implant tissues, encompass peri-implant mucositis and peri-implantitis. Peri-implant... (Review)
Review
Peri-implant diseases, characterized by inflammatory conditions affecting peri-implant tissues, encompass peri-implant mucositis and peri-implantitis. Peri-implant mucositis is an inflammatory lesion limited to the mucosa around an implant, while peri-implantitis extends from the mucosa to the supporting bone, causing a loss of osseointegration. For non-surgical treatments, we tested the null hypothesis that the presence or absence of air-polishing made no difference. The study focused on randomized controlled trials (RCTs) comparing air-polishing with mechanical or ultrasonic debridement, evaluating outcomes such as bleeding on probing (BOP), probing depth (PD), plaque index/plaque score (PI/PS), clinical attachment level (CAL), bone loss, and mucosal recession (MR). Two independent reviewers conducted data extraction and quality assessments, considering short-term (<6 months) and long-term (≥6 months) follow-up periods. After screening, ten articles were included in the meta-analysis. In nonsurgical peri-implant disease management, air-polishing moderately mitigated short-term PI/PS for peri-implant mucositis and showed a similar improvement in long-term BOP and bone loss for peri-implantitis compared to the control group. The Egger test found no evidence of publication bias except for the long-term PI/PS of peri-implant mucositis. Leave-one-out analysis confirmed the stability of the results. The findings highlight the need for future research with longer-term follow-up and high-quality, multi-center, large-sample RCTs.
PubMed: 38828461
DOI: 10.1016/j.jdsr.2024.05.003 -
Biomaterials Research 2024With the high incidence rate, distinctive implant characteristic and unique infection pattern, peri-implantitis (PI) requires a specially designed implant animal model...
With the high incidence rate, distinctive implant characteristic and unique infection pattern, peri-implantitis (PI) requires a specially designed implant animal model for the researches on the pathogenesis and treatments. Previous small-animal PI models exhibit variability in implant site selection, design, and surgical procedures resulting in unnecessary tissue damage and less effectivity. Herein, a quantitative-analysis-based standardized rat model for transmucosal PI-related research was proposed. After dissecting the anatomic structures of the rat maxilla, we determined that placing the implant anterior to the molars in the rat maxilla streamlined the experimental period and enhanced animal welfare. We standardized the model by controlling the rat strain, gender, and size. The customized implant and a series of matched surgical instruments were appropriately designed. A clear, step-by-step surgical process was established. These designs ensured the success rate, stability, and replicability of the model. Each validation method confirmed the successful construction of the model. This study proposed a quantitative-analysis-based standardized transmucosal PI rat model with improved animal welfare and reliable procedures. This model could provide efficient in vivo insights to study the pathogenesis and treatments of PI and preliminary screening data for further large-animal and clinical trials.
PubMed: 38828365
DOI: 10.34133/bmr.0021 -
Cureus Apr 2024Since ancient times, several sorts of nanoparticles have been employed in the quickly expanding field of nanotechnology. These features include size, shape, and chemical... (Review)
Review
Since ancient times, several sorts of nanoparticles have been employed in the quickly expanding field of nanotechnology. These features include size, shape, and chemical as well as physical properties. Because of their small size and huge surface area, carbon-based nanoparticles, including fullerenes, carbon nanotubes, graphene, graphene oxide, and carbon-based quantum dots, have attracted a lot of attention in a variety of sectors, including biomedical applications. Lipid bilayers form the spherical vesicles known as liposomes. Magnetic resonance imaging (MRI) contrast agents are iron oxide nanoparticles. These materials are perfect for drug and delivery of genes, bioimaging, and bone repair because of their remarkable mechanical, electrical, visual, and chemical properties. However, concerns about potential asbestos-related diseases have arisen due to their length-to-width aspect ratio. Ceramic nanoparticles, on the other hand, are a common material in daily life and play a crucial role in bone repair, multiscale hybridisation, and aerospace structures. These nanoparticles can enhance osseointegration and bone development by mimicking the nanocomposition and nanoscale characteristics of bone tissue and enhance osteoconductive and osteoinductive capacities. Ceramic nanoparticles, however, have the potential to generate oxidative stress, which can result in irritation of the reticuloendothelial system, cytotoxicity to the heart, liver, and lungs, as well as toxicity to the cells that are attached. Additionally, oxidative stress, cell damage, and genotoxicity might result from the generation of free radicals by ceramic nanoparticles. Metal nanoparticles exhibit linear optical properties similar to molecular systems but arise from a different physical process. Semiconductor nanocrystals (NCs) are made from various compounds, such as silicon and germanium. Polyandry nanoparticles are particles approximately 10 and 10000 nanometers (nm) in size that can contain active substances. They have applications in vaccine delivery, gene therapy, and polymer nanoparticles (nanomedicine) for therapeutic applications.
PubMed: 38813303
DOI: 10.7759/cureus.59234 -
Plastic and Reconstructive Surgery.... May 2024Upper extremity limb loss profoundly impacts a patient's quality of life and well-being and carries a significant societal cost. Although osseointegration allows the...
BACKGROUND
Upper extremity limb loss profoundly impacts a patient's quality of life and well-being and carries a significant societal cost. Although osseointegration allows the attachment of the prosthesis directly to the bone, it is a relatively recent development as an alternative to conventional socket prostheses. The objective of this review was to identify reports on osseointegrated prosthetic embodiment for transhumeral amputations and assess the implant systems used, postoperative outcomes, and complications.
METHODS
A systematic review following PRISMA and AMSTAR guidelines assessed functional outcomes, implant longevity and retention, activities of daily living, and complications associated with osseointegrated prostheses in transhumeral amputees.
RESULTS
The literature search yielded 794 articles, with eight of these articles (retrospective analyses and case series) meeting the inclusion criteria. Myoelectric systems equipped with Osseointegrated Prostheses for the Rehabilitation of Amputees implants have been commonly used as transhumeral osseointegration systems. The transhumeral osseointegrated prostheses offered considerable improvements in functional outcomes, with participants demonstrating enhanced range of motion and improved performance of activities compared with traditional socket-based prostheses. One study demonstrated the advantage of an osseointegrated implant as a bidirectional gateway for signal transmission, enabling intuitive control of a bionic hand.
CONCLUSIONS
Osseointegrated prostheses hold the potential to significantly improve the quality of life for individuals with transhumeral amputations. Continued research and clinical expansion are expected to lead to the realization of enhanced efficacy and safety in this technique, accompanied by cost reductions over time as a result of improved efficiencies and advancements in device design.
PubMed: 38808147
DOI: 10.1097/GOX.0000000000005850 -
International Journal of Nanomedicine 2024There is an ongoing need for improved healing response and expedited osseointegration on the Ti implants in acetabular fracture sites. To achieve adequate bonding and...
INTRODUCTION
There is an ongoing need for improved healing response and expedited osseointegration on the Ti implants in acetabular fracture sites. To achieve adequate bonding and mechanical stability between the implant surface and the acetabular fracture, a new coating technology must be developed to promote bone integration and prevent bacterial growth.
METHODS
A cylindrical Ti substrate mounted on a rotating specimen holder was used to implant Ca, P, and Sr ions at energies of 100 KeV, 75 KeV and 180 KeV, respectively, using a low-energy accelerator to synthesize strontium-substituted hydroxyapatite at varying conditions. Ag ions of energy 100 KeV were subsequently implanted on the as-formed surface at the near-surface region to provide anti-bacterial properties to the as-formed specimen.
RESULTS
The properties of the as-formed ion-implanted specimen were compared with the SrHA-Ag synthesized specimens by cathodic deposition and low-temperature high-speed collision technique. The adhesion strength of the ion-implanted specimen was 43 ± 2.3 MPa, which is well above the ASTM standard for Ca-P coating on Ti. Live/dead cell analysis showed higher osteoblast activity on the ion-implanted specimen than the other two. Ag in the SrHA implanted Ti by ion implantation process showed superior antibacterial activity.
DISCUSSION
In the ion implantation technique, nano-topography patterned surfaces are not concealed after implantation, and their efficacy in interacting with the osteoblasts is retained. Although all three studies examined the antibacterial effects of Ag ions and the ability to promote bone tissue formation by MC3T3-E1 cells on SrHA-Ag/Ti surfaces, ion implantation techniques demonstrated superior ability. The synthesized specimen can be used as an effective implant in acetabular fracture sites based on their mechanical and biological properties.
Topics: Titanium; Silver; Strontium; Anti-Bacterial Agents; Acetabulum; Animals; Coated Materials, Biocompatible; Osseointegration; Mice; Surface Properties; Fractures, Bone; Durapatite; Osteoblasts; Hydroxyapatites; Prostheses and Implants; Ions; Humans; Cell Line
PubMed: 38803996
DOI: 10.2147/IJN.S464905