International Journal of Radiation... Mar 2017Radiation therapy (RT) is a crucial component of cancer care, used in the treatment of over 50% of cancer patients. Patients undergoing image guided RT or brachytherapy... (Review)
Review
Radiation therapy (RT) is a crucial component of cancer care, used in the treatment of over 50% of cancer patients. Patients undergoing image guided RT or brachytherapy routinely have inert RT biomaterials implanted into their tumors. The single function of these RT biomaterials is to ensure geometric accuracy during treatment. Recent studies have proposed that the inert biomaterials could be upgraded to "smart" RT biomaterials, designed to do more than 1 function. Such smart biomaterials include next-generation fiducial markers, brachytherapy spacers, and balloon applicators, designed to respond to stimuli and perform additional desirable functions like controlled delivery of therapy-enhancing payloads directly into the tumor subvolume while minimizing normal tissue toxicities. More broadly, smart RT biomaterials may include functionalized nanoparticles that can be activated to boost RT efficacy. This work reviews the rationale for smart RT biomaterials, the state of the art in this emerging cross-disciplinary research area, challenges and opportunities for further research and development, and a purview of potential clinical applications. Applications covered include using smart RT biomaterials for boosting cancer therapy with minimal side effects, combining RT with immunotherapy or chemotherapy, reducing treatment time or health care costs, and other incipient applications.
Topics: Artificial Intelligence; Biocompatible Materials; Brachytherapy; Fiducial Markers; Humans; Nanoparticles; Radiotherapy, Image-Guided
PubMed: 28126309
DOI: 10.1016/j.ijrobp.2016.10.034 -
Current Oncology (Toronto, Ont.) May 2023Fiducial markers are utilized for image guided radiotherapy (IGRT) alignment during the delivery of liver stereotactic body radiosurgery (SBRT). There are limited data...
Fiducial markers are utilized for image guided radiotherapy (IGRT) alignment during the delivery of liver stereotactic body radiosurgery (SBRT). There are limited data demonstrating the impact of matching fiducials on the accuracy of liver SBRT. This study quantifies the benefit of fiducial-based alignment and improvements in inter-observer reliability. Nineteen patients with 24 liver lesions were treated with SBRT. Target localization was performed using fiducial markers on cone-beam computed tomography (CBCT). Each CBCT procedure was retrospectively realigned to match both the liver edge and fiducial markers. The shifts were recorded by seven independent observers. Inter-observer variability was analyzed by calculating the mean error and uncertainty for the set-up. The mean absolute Cartesian error observed from fiducial and liver edge-based alignment was 1.5 mm and 5.3 mm, respectively. The mean uncertainty from fiducial and liver edge-based alignment was 1.8 mm and 4.5 mm, respectively. An error of 5 mm or greater was observed 50% of the time when aligning to the liver surface versus 5% of the time when aligning to fiducial markers. Aligning to the liver edge significantly increased the error, resulting in increased shifts when compared to alignment to fiducials. Tumors of 3 cm or farther from the liver dome had higher mean errors when aligned without fiducials (4.8 cm vs. 4.4 cm, = 0.003). Our data support the use of fiducial markers for safer and more accurate liver SBRT.
Topics: Humans; Fiducial Markers; Radiosurgery; Retrospective Studies; Reproducibility of Results; Liver
PubMed: 37232840
DOI: 10.3390/curroncol30050382 -
Radiation Oncology (London, England) Dec 2019In this study, a novel pelvic phantom was developed and used to assess the visibility and presence of artefacts from different types of commercial fiducial markers (FMs)...
BACKGROUND
In this study, a novel pelvic phantom was developed and used to assess the visibility and presence of artefacts from different types of commercial fiducial markers (FMs) on multi-modality imaging relevant to prostate cancer.
METHODS AND MATERIALS
The phantom was designed with 3D printed hollow cubes in the centre. These cubes were filled with gel to mimic the prostate gland and two parallel PVC rods were used to mimic bones in the pelvic region. Each cube was filled with gelatine and three unique FMs were positioned with a clinically-relevant spatial distribution. The FMs investigated were; Gold Marker (GM) CIVCO, GM RiverPoint, GM Gold Anchor (GA) line and ball shape, and polymer marker (PM) from CIVCO. The phantom was scanned using several imaging modalities typically used to image prostate cancer patients; MRI, CT, CBCT, planar kV-pair, ExacTrac, 6MV, 2.5MV and integrated EPID imaging. The visibility of the markers and any observed artefacts in the phantom were compared to in-vivo scans of prostate cancer patients with FMs.
RESULTS
All GMs were visible in volumetric scans, however, they also had the most visible artefacts on CT and CBCT scans, with the magnitude of artefacts increasing with FM size. PM FMs had the least visible artefacts in volumetric scans but they were not visible on portal images and had poor visibility on lateral kV images. The smallest diameter GMs (GA) were the most difficult GMs to identify on lateral kV images.
CONCLUSION
The choice between different FMs is also dependent on the adopted IGRT strategy. PM was found to be superior to investigated gold markers in the most commonly used modalities in the management of prostate cancer; CT, CBCT and MRI imaging.
Topics: Artifacts; Cone-Beam Computed Tomography; Fiducial Markers; Gold; Humans; Image Processing, Computer-Assisted; Magnetic Resonance Imaging; Male; Multimodal Imaging; Organs at Risk; Phantoms, Imaging; Prostatic Neoplasms; Radiotherapy Dosage; Radiotherapy Planning, Computer-Assisted; Radiotherapy, Image-Guided; Radiotherapy, Intensity-Modulated; Signal-To-Noise Ratio
PubMed: 31878967
DOI: 10.1186/s13014-019-1447-1 -
European Respiratory Review : An... Mar 2022Stereotactic body radiation therapy is an alternative to surgery for early-stage, inoperable peripheral non-small cell lung cancer. As opposed to linear accelerator... (Review)
Review
Stereotactic body radiation therapy is an alternative to surgery for early-stage, inoperable peripheral non-small cell lung cancer. As opposed to linear accelerator (linac)-based ( gating) and free-breathing techniques, CyberKnife® with Synchrony® technology allows accurate radiation delivery by means of a real-time respiratory motion tracking system using, in most cases, metal fiducial markers (FMs) placed in the vicinity of the target. The aims of this review are as follows. First, to describe the safety and efficacy of the transthoracic, endovascular and endobronchial FM insertion techniques for peripheral pulmonary lesions (PPLs). Second, to analyse performance in terms of the migration and tracking rates of different FM types. Recent developments in FM tracking for central lesions will also be reviewed. In conclusion, for PPLs, the endobronchial approach provides a low rate of pneumothorax, offers the possibility of concurrent diagnostic sampling for both the PPL and the lymph nodes, and, finally, reduces the intervention time compared to other techniques. In this context, coil-tailed and coil-spring FMs have shown the lowest migration rate with a consequently high tracking rate.
Topics: Carcinoma, Non-Small-Cell Lung; Fiducial Markers; Humans; Lung; Lung Neoplasms; Radiosurgery
PubMed: 35022258
DOI: 10.1183/16000617.0149-2021 -
Otology & Neurotology : Official... Jan 2011Using image-guided surgical techniques, we propose that an industrial robot can be programmed to safely, effectively, and efficiently perform a mastoidectomy.
HYPOTHESIS
Using image-guided surgical techniques, we propose that an industrial robot can be programmed to safely, effectively, and efficiently perform a mastoidectomy.
BACKGROUND
Whereas robotics is a mature field in many surgical applications, robots have yet to be clinically used in otologic surgery despite significant advantages including reliability and precision.
METHODS
We designed a robotic system that incorporates custom software with an industrial robot to manipulate a surgical drill through a complex milling profile. The software controls the movements of the robot based on real-time feedback from a commercially available optical tracking system. The desired path of the drill to remove the desired volume of mastoid bone was planned using computed tomographic scans of cadaveric specimens and then implemented using the robotic system. Bone-implanted fiducial markers were used to provide accurate registration between computed tomographic and physical space.
RESULTS
A mastoid cavity was milled on 3 cadaveric specimens with a 5-mm fluted ball bit. Postmilling computed tomographic scans showed that, for the 3 specimens, 97.70%, 99.99%, and 96.05% of the target region was ablated without violation of any critical feature.
CONCLUSION
To the best of our knowledge, this is the first time that a robot has been used to perform a mastoidectomy. Although significant hurdles remain to translate this technology to clinical use, we have shown that it is feasible. The prospect of reducing surgical time and enhancing patient safety by replacing human hand-eye coordination with machine precision motivates future work toward translating this technique to clinical use.
Topics: Fiducial Markers; Humans; Mastoid; Otologic Surgical Procedures; Robotics
PubMed: 21042227
DOI: 10.1097/MAO.0b013e3181fcee9e -
PloS One 2018Cardiac toxicity after definitive chemoradiotherapy for esophageal cancer is a critical issue. To reduce irradiation doses to organs at risk, individual internal margins...
INTRODUCTION
Cardiac toxicity after definitive chemoradiotherapy for esophageal cancer is a critical issue. To reduce irradiation doses to organs at risk, individual internal margins need to be identified and minimized. The purpose of this study was to quantify esophageal motion using fiducial makers based on four-dimensional computed tomography, and to evaluate the inter-CBCT session marker displacement using breath-hold.
MATERIALS AND METHODS
Sixteen patients with early stage esophageal cancer, who received endoscopy-guided metallic marker placement for treatment planning, were included; there were 35 markers in total, with 9, 15, and 11 markers in the upper thoracic, middle thoracic, and lower thoracic/esophagogastric junction regions, respectively. We defined fiducial marker motion as motion of the centroidal point of the markers. Respiratory esophageal motion during free-breathing was defined as the amplitude of individual marker motion between the consecutive breathing and end-expiration phases, derived from four-dimensional computed tomography. The inter-CBCT session marker displacement using breath-hold was defined as the amplitudes of marker motion between the first and each cone beam computed tomography image. Marker motion was analyzed in the three regions (upper thoracic, middle thoracic, and lower thoracic/esophagogastric junction) and in three orthogonal directions (right-left; anterior-posterior; and superior-inferior).
RESULTS
Respiratory esophageal motion during free-breathing resulted in median absolute maximum amplitudes (interquartile range), in right-left, anterior-posterior, and superior-inferior directions, of 1.7 (1.4) mm, 2.0 (1.5) mm, and 3.6 (4.1) mm, respectively, in the upper thoracic region, 0.8 (1.1) mm, 1.4 (1.2) mm, and 4.8 (3.6) mm, respectively, in the middle thoracic region, and 1.8 (0.8) mm, 1.9 (2.0) mm, and 8.0 (4.5) mm, respectively, in the lower thoracic/esophagogastric region. The inter-CBCT session marker displacement using breath-hold resulted in median absolute maximum amplitudes (interquartile range), in right-left, anterior-posterior, and superior-inferior directions, of 1.3 (1.0) mm, 1.1 (0.7) mm, and 3.3 (1.8) mm, respectively, in the upper thoracic region, 0.7 (0.7) mm, 1.1 (0.4) mm, and 3.4 (1.4) mm, respectively, in the middle thoracic region, and 2.0 (0.8) mm, 2.6 (2.2) mm, and 3.5 (1.8) mm, respectively, in the lower thoracic/esophagogastric region.
CONCLUSIONS
During free-breathing, esophageal motion in the superior-inferior direction in all sites was large, compared to the other directions, and amplitudes showed substantial inter-individual variability. The breath-hold technique is feasible for minimizing esophageal displacement during radiotherapy in patients with esophageal cancer.
Topics: Aged; Aged, 80 and over; Cone-Beam Computed Tomography; Esophageal Neoplasms; Esophagogastric Junction; Esophagus; Female; Fiducial Markers; Four-Dimensional Computed Tomography; Humans; Male; Middle Aged; Neoplasm Staging; Radiotherapy Planning, Computer-Assisted; Respiration
PubMed: 29889910
DOI: 10.1371/journal.pone.0198844 -
International Journal of Hyperthermia :... 2021Fiducial markers improve accuracy in external beam radiation therapy (EBRT) for treatment of prostate cancer (PCa). However, many patients recur after EBRT necessitating...
OBJECTIVES
Fiducial markers improve accuracy in external beam radiation therapy (EBRT) for treatment of prostate cancer (PCa). However, many patients recur after EBRT necessitating additional treatment, such as MR-guided transurethral ultrasound ablation (TULSA). Residual markers may compromise TULSA through ultrasound field distortions and generation of local susceptibility artifacts. The objective was to investigate how markers affect the ablation outcome during clinical TULSA treatments.
SUBJECTS AND METHODS
A retrospective analysis was performed on nine patients with radiorecurrent PCa and residual markers who received TULSA. The MR susceptibility artifact was quantified as a function of marker type, size and orientation, in particular for thermometry. The spatial distribution of markers inside the prostate was recorded, and the resulting impact on the thermal dose was measured. The thermal dose measurements were directly compared to the residual enhancing prostatic tissue observed on the immediate and control post-TULSA contrast enhanced (CE) image.
RESULTS
Successful thermal dose accumulation to the target boundary occurred for 14/20 (70%) of markers, confirmed with CE imaging. Gold markers situated simultaneously close to the urethra (≤12 mm) and far from the target boundary (≥13 mm) reduced the ultrasound depth of heating. Nitinol markers produced large, hypointense artifacts that disrupted thermometry and compromised treatment. Artifacts from gold markers were less pronounced, but when located near the target boundary, also affected treatment.
CONCLUSION
Marker composition, orientation and location inside the prostate can all potentially impact treatment outcome. Proper patient selection through detailed MRI screening is critical to ensure successful radiorecurrent PCa treatment outcomes with TULSA.
Topics: Fiducial Markers; High-Intensity Focused Ultrasound Ablation; Humans; Magnetic Resonance Imaging; Male; Neoplasm Recurrence, Local; Prostatic Neoplasms; Retrospective Studies
PubMed: 34927517
DOI: 10.1080/02656736.2021.2008519 -
The British Journal of Radiology Nov 2016We reviewed the literature on the use of margins in radiotherapy of patients with prostate cancer, focusing on different options for image guidance (IG) and technical... (Review)
Review
We reviewed the literature on the use of margins in radiotherapy of patients with prostate cancer, focusing on different options for image guidance (IG) and technical issues. The search in PubMed database was limited to include studies that involved external beam radiotherapy of the intact prostate. Post-prostatectomy studies, brachytherapy and particle therapy were excluded. Each article was characterized according to the IG strategy used: positioning on external marks using room lasers, bone anatomy and soft tissue match, usage of fiducial markers, electromagnetic tracking and adapted delivery. A lack of uniformity in margin selection among institutions was evident from the review. In general, introduction of pre- and in-treatment IG was associated with smaller planning target volume (PTV) margins, but there was a lack of definitive experimental/clinical studies providing robust information on selection of exact PTV values. In addition, there is a lack of comparative research regarding the cost-benefit ratio of the different strategies: insertion of fiducial markers or electromagnetic transponders facilitates prostate gland localization but at a price of invasive procedure; frequent pre-treatment imaging increases patient in-room time, dose and labour; online plan adaptation should improve radiation delivery accuracy but requires fast and precise computation. Finally, optimal protocols for quality assurance procedures need to be established.
Topics: Fiducial Markers; Humans; Male; Prostatic Neoplasms; Radiotherapy Dosage; Radiotherapy Planning, Computer-Assisted; Radiotherapy, Image-Guided
PubMed: 27377353
DOI: 10.1259/bjr.20160312 -
Physica Medica : PM : An International... Aug 2020This work investigates the time and frequency to observe fiducial markers in MLC-modulated fields during intensity-modulated radiotherapy (IMRT) and volumetric-modulated...
OBJECTIVE
This work investigates the time and frequency to observe fiducial markers in MLC-modulated fields during intensity-modulated radiotherapy (IMRT) and volumetric-modulated arc therapy (VMAT) beam delivery for real-time prostate localization.
METHODS
Thirty seven prostate patients treated with IMRT or VMAT were included in this retrospective study. DRR images were generated for all MLC segments/control points using the TPS. The MLC leaf pattern of each control point was overlaid on the DRR, and the number of fiducials within the MLC opening was analyzed. EPID images of fiducials in a pelvic phantom were obtained to demonstrate the fiducial visibility during modulated beam delivery.
RESULTS
Gold fiducials were visible on EPID images. The probability of seeing a number of fiducials within the MLC opening was analyzed. At least one fiducial was visible during 42 ± 2% and 52 ± 2% beam-on time for IMRT of the prostate with and without lymph nodes, and during 81 ± 4% and 80 ± 5% beam-on time for VMAT of the prostate with and without lymph nodes, respectively. The mean time interval to observe at least one fiducial was 8.4 ± 0.7 and 5.9 ± 0.5 s for IMRT of the prostate with and without the lymph nodes, respectively, and 1.6 ± 0.1 s for VMAT prostate patients. The estimated potential dosimetric uncertainty was 7% and 2% for IMRT and VMAT, respectively.
CONCLUSIONS
Our results demonstrated that the time and frequency to observe fiducial markers in MLC-modulated fields during IMRT/VMAT beam delivery were adequate for real-time prostate localization. The beam's eye view fiducial positions could be used for intrafractional target monitoring and motion correction in prostate radiotherapy.
Topics: Fiducial Markers; Humans; Male; Prostatic Neoplasms; Radiotherapy Dosage; Radiotherapy Planning, Computer-Assisted; Radiotherapy, Intensity-Modulated; Retrospective Studies
PubMed: 32679409
DOI: 10.1016/j.ejmp.2020.06.026 -
The British Journal of Radiology Dec 2016Image-guided radiotherapy (IGRT) is an essential tool in the accurate delivery of modern radiotherapy techniques. Prostate radiotherapy positioned using skin marks or... (Review)
Review
Image-guided radiotherapy (IGRT) is an essential tool in the accurate delivery of modern radiotherapy techniques. Prostate radiotherapy positioned using skin marks or bony anatomy may be adequate for delivering a relatively homogeneous whole-pelvic radiotherapy dose, but these surrogates are not reliable when using reduced margins, dose escalation or hypofractionated stereotactic radiotherapy. Fiducial markers (FMs) for prostate IGRT have been in use since the 1990s. They require surgical implantation and provide a surrogate for the position of the prostate gland. A variety of FMs are available and they can be used in a number of ways. This review aimed to establish the evidence for using prostate FMs in terms of feasibility, implantation procedures, types of FMs used, FM migration, imaging modalities used and the clinical impact of FMs. A search strategy was defined and a literature search was carried out in Medline. Inclusion and exclusion criteria were applied, which resulted in 50 articles being included in this review. The evidence demonstrates that FMs provide a more accurate surrogate for the position of the prostate than either external skin marks or bony anatomy. A combination of FM alignment and soft-tissue analysis is currently the most effective and widely available approach to ensuring accuracy in prostate IGRT. FM implantation is safe and well tolerated. FM migration is possible but minimal. Standardization of all techniques and procedures in relation to the use of prostate FMs is required. Finally, a clinical trial investigating a non-surgical alternative to prostate FMs is introduced.
Topics: Fiducial Markers; Humans; Male; Prostate; Prostatic Neoplasms; Radiotherapy, Image-Guided; Radiotherapy, Intensity-Modulated
PubMed: 27585736
DOI: 10.1259/bjr.20160296