Clinical Microbiology and Infection :... Oct 2018Several outbreaks of severe infections due to contamination of gastrointestinal (GI) endoscopes, mainly duodenoscopes, have been described. The rate of microbial...
OBJECTIVES
Several outbreaks of severe infections due to contamination of gastrointestinal (GI) endoscopes, mainly duodenoscopes, have been described. The rate of microbial endoscope contamination varies dramatically in literature. The aim of this multicentre prospective study was to evaluate the hygiene quality of endoscopes and automated endoscope reprocessors (AERs) in Tyrol/Austria.
METHODS
In 2015 and 2016, a total of 463 GI endoscopes and 105 AERs from 29 endoscopy centres were analysed by a routine (R) and a combined routine and advanced (CRA) sampling procedure and investigated for microbial contamination by culture-based and molecular-based analyses.
RESULTS
The contamination rate of GI endoscopes was 1.3%-4.6% according to the national guideline, suggesting that 1.3-4.6 patients out of 100 could have had contacts with hygiene-relevant microorganisms through an endoscopic intervention. Comparison of R and CRA sampling showed 1.8% of R versus 4.6% of CRA failing the acceptance criteria in phase I and 1.3% of R versus 3.0% of CRA samples failing in phase II. The most commonly identified indicator organism was Pseudomonas spp., mainly Pseudomonas oleovorans. None of the tested viruses were detected in 40 samples. While AERs in phase I failed (n = 9, 17.6%) mainly due to technical faults, phase II revealed lapses (n = 6, 11.5%) only on account of microbial contamination of the last rinsing water, mainly with Pseudomonas spp.
CONCLUSIONS
In the present study the contamination rate of endoscopes was low compared with results from other European countries, possibly due to the high quality of endoscope reprocessing, drying and storage.
Topics: Austria; Cross Infection; Decontamination; Endoscopes, Gastrointestinal; Equipment Contamination; Europe; Humans; Prospective Studies; Pseudomonas
PubMed: 29408276
DOI: 10.1016/j.cmi.2018.01.017 -
BMC Gastroenterology Apr 2023Endoscopic retrograde cholangiography (ERC) possesses a translocation risk of microbes to the biliary system. We studied bile contamination during ERC and its impact on...
BACKGROUND
Endoscopic retrograde cholangiography (ERC) possesses a translocation risk of microbes to the biliary system. We studied bile contamination during ERC and its impact on patients' outcome in a real-life-situation.
METHODS
Ninety-nine ERCs were analyzed and microbial samples were taken from the throat before and from bile during ERC and from irrigation fluid of the duodenoscope before and after ERC.
RESULTS
91.2% of cholangitis patients had detectable microbes in the bile (sensitivity 91%), but the same was true for 86.2% in the non-cholangitis group. Bacteroides fragilis (p=0.015) was significantly associated with cholangitis. In 41.7% of ERCs with contaminated endoscopes these microbes were found in the bile after the procedure. Analysis of duodenoscopes' irrigation liquid after ERC matched the microbial bile analysis of these patients in 78.8%. Identical microbial species were in throat and in bile samples of the same ERC in 33% of all cases and in 45% in the non-cholangitis group. Transmission of microbes to the biliary tract did not result in more frequent cholangitis, longer hospital stays, or worse outcome.
CONCLUSIONS
During ERC bile samples are regularly contaminated with microbes of the oral cavity but it did not affect clinical outcome.
Topics: Humans; Cholangiopancreatography, Endoscopic Retrograde; Biliary Tract; Cholangiography; Biliary Tract Surgical Procedures; Cholangitis; Microbiota
PubMed: 37013522
DOI: 10.1186/s12876-023-02721-7 -
Digestion 2018Endoscopic resection (ER) is becoming the first choice of treatment for treating superficial nonampullary duodenal epithelial tumors (SNADETs), but ER procedures for... (Review)
Review
BACKGROUND
Endoscopic resection (ER) is becoming the first choice of treatment for treating superficial nonampullary duodenal epithelial tumors (SNADETs), but ER procedures for SNADETs remain challenging because of the difficulty experienced in maneuvering the endoscope toward the thin duodenal wall, which results in a high rate of adverse events. Although several ER methods were used to overcome these technical difficulties and complications, ER methods for SNADETs are not standardized. A new technique, underwater endoscopic mucosal resection (UEMR), was developed recently in a western country, and its usefulness was reported. Beginning in 2014, we were the first to use UEMR for SNADETs in Japan. Thus, in our experience, we would propose an indication of the various ER methods for SNADETs according to the lesion size.
SUMMARY
Endoscopic mucosal resection (EMR) and UEMR were effective and safe for small lesions (≤20 mm), but for large lesions (>20 mm), piecemeal removal of lesion by EMR and UEMR had high incidence of recurrence and adverse events. Especially, piecemeal EMR could cause delayed perforation. Cold snare polypectomy was useful for small lesions (≤10 mm), but further study of its recurrence is warranted. Endoscopic submucosal dissection (ESD) achieved a high complete resection rate regardless of the lesion size, but its rate of adverse events, including morbid complications, was high. Thus, after ESD for large lesions, secure prevention method for adverse events, such as closure of the wound by laparoscopic-endoscopic cooperative surgery, should be required. Key Messages: ER methods for treating SNADETs were proposed based on the lesion size. For large lesions, prophylactic methods for adverse events should be implemented.
Topics: Duodenal Neoplasms; Duodenoscopes; Duodenoscopy; Duodenum; Endoscopic Mucosal Resection; Humans; Incidence; Intestinal Mucosa; Japan; Laparoscopy; Neoplasm Recurrence, Local; Neoplasms, Glandular and Epithelial; Postoperative Complications; Treatment Outcome
PubMed: 29393159
DOI: 10.1159/000484112 -
Journal of Personalized Medicine May 2021Endoscopic retrograde cholangiopancreatography (ERCP) in patients who have undergone a Billroth II gastrectomy is a major challenge. This study aimed to evaluate the...
Endoscopic retrograde cholangiopancreatography (ERCP) in patients who have undergone a Billroth II gastrectomy is a major challenge. This study aimed to evaluate the outcomes of the road-map technique for duodenal intubation using a side-viewing duodenoscope for ERCP in Billroth II gastrectomy patients with naïve papilla, and to analyze the formation and release patterns of common bowel loops that occur when the duodenoscope navigates the afferent limb. The duodenoscopy approach success rate was 85.8% (97/113). In successful duodenoscopy approach patients, there were five bowel looping patterns that occurred when the preceding catheter-connected duodenoscope was advanced into the duodenum: (1) reverse ɣ-loop (29.9%), (2) fixed reverse ɣ-loop (5.2%), (3) simple U-loop (22.7%), (4) N-loop (28.9%), and (5) reverse alpha loop (13.4%). The duodenoscopy cannulation and duodenoscopy therapeutic success rates were 81.4% (92/113) and 80.5% (91/113), respectively, while the overall cannulation and therapeutic success rates were 92.0% (104/113) and 87.6% (99/113), respectively. Bowel perforation occurred in three patients (2.7%). The road-map technique may benefit duodenoscope-based ERCP in Billroth II gastrectomy patients by minimizing the tangential axis alignment between the duodenoscopic tip and driving of the afferent limb, and by predicting and counteracting bowel loops that occur when the duodenoscope navigates the afferent limb.
PubMed: 34066235
DOI: 10.3390/jpm11050404 -
Gastrointestinal Endoscopy Clinics of... Oct 2020Flexible endoscopes require cleaning, high-level disinfection, and sterilization between each patient use to reduce risk of transmitting pathogens. Public health... (Review)
Review
Flexible endoscopes require cleaning, high-level disinfection, and sterilization between each patient use to reduce risk of transmitting pathogens. Public health investigations have identified concerns, including endoscope damage, mishandling, and reprocessing deficiencies, placing patients at risk for transmission of bacterial, viral, and other pathogens. Findings from outbreak investigations and other studies have led to innovations in endoscope design, use, and reprocessing, yet infection risks related to contaminated or damaged endoscopes remain. Strict adherence to infection control guidelines and manufacturer instructions for use, utilization of supplemental guidance, and training and oversight of reprocessing personnel, reduce risk of pathogen transmission by flexible endoscopes.
Topics: Centers for Disease Control and Prevention, U.S.; Cross Infection; Disease Outbreaks; Endoscopes, Gastrointestinal; Guideline Adherence; Guidelines as Topic; Humans; Infection Control; United States
PubMed: 32891228
DOI: 10.1016/j.giec.2020.06.009 -
Digestive Diseases and Sciences Mar 2022The challenging disinfection process for the elevator mechanism on duodenoscopes and linear echoendoscopes has been identified as a source of clinically significant... (Comparative Study)
Comparative Study
BACKGROUND
The challenging disinfection process for the elevator mechanism on duodenoscopes and linear echoendoscopes has been identified as a source of clinically significant bacterial transmission. Despite increased awareness, there continues to be a lack of definitive guidelines for bacterial culturing protocols for elevator-containing endoscopes.
AIMS
To compare two different prospective bacterial surveillance protocols for duodenoscopes and linear echoendoscopes with regard to accuracy, efficiency, and cost.
METHODS
Consecutive duodenoscopes and linear echoendoscopes used at a single tertiary care center were reprocessed following hospital and manufacturer guidelines, dried using an automatic endoscope-drying machine, and hung overnight in an upright position. Following reprocessing, culture samples were sequentially obtained from each endoscope using two methods, first, the brush protocol followed immediately by the swab protocol.
RESULTS
A total of 532 primary cultures were collected from 17 duodenoscopes and eight linear echoendoscopes. Of these, 266 cultures gathered using the brush protocol were negative, while 266 cultures gathered using the swab protocol resulted in three positive cultures (1.1%). Positive cultures showed Enterobacter cloacae and Klebsiella pneumoniae from one duodenoscope and two linear echoendoscopes. Yearly, the brush protocol amounts to approximately 520 nursing hours, and the swab protocol takes an estimated 42 nursing hours. Annually, the swab protocol could save over $26,500 and 478 nursing hours.
CONCLUSIONS
The proposed swab protocol was superior to the brush protocol when evaluating the presence of residual bacteria on elevator-containing endoscopes following reprocessing and saves cost and nursing hours.
Topics: Bacteria; Disinfection; Duodenoscopes; Endoscopes; Enterobacter cloacae; Equipment Contamination; Humans; Klebsiella pneumoniae; Prospective Studies
PubMed: 33712967
DOI: 10.1007/s10620-021-06930-6 -
The Journal of Hospital Infection Feb 2023Microbiological cultures are the gold standard in the monitoring of duodenoscope reprocessing. However, many different sampling and culturing techniques are used, making...
AIM
Microbiological cultures are the gold standard in the monitoring of duodenoscope reprocessing. However, many different sampling and culturing techniques are used, making it difficult to compare results. The latest Centers for Disease Control and Prevention protocol advises the use of a neutralizer to deactivate any remaining disinfectants in the samples. This study compared culturing results of duodenoscope samples collected with and without addition of a neutralizer.
METHODS
Six duodenoscopes were soiled with gut bacteria in a non-clinical experimental setting and reprocessed afterwards. Samples of the tip and working channel were collected immediately after decontamination or after drying. Dey-Engley (DE) broth was added as a neutralizer to the samples of four duodenoscopes; samples for the other two duodenoscopes were collected without the addition of DE broth.
RESULTS
Post-decontamination cultures were significantly more likely to be positive for growth of the applied micro-organisms in the group of samples with DE broth (88.1% vs 20.2%; P<0.0001). Post-drying samples were significantly more likely to be positive in the group of samples without DE broth (75.7% vs 33.4%; P<0.001).
CONCLUSION
The addition of DE broth to samples collected from wet duodenoscopes increases the yield of those cultures. Remaining disinfectants in wet duodenoscopes can lead to false-negative results. This can be overcome by adding a neutralizer, such as DE broth, to the samples. The higher yield after drying in the group without neutralizer could be due to biofilm formation in these two duodenoscopes, but this was not investigated. Standardization of the sampling method can help to compare both clinical and study results regarding duodenoscope contamination.
Topics: Humans; Duodenoscopes; Equipment Contamination; Disinfectants; Bacteria; Disinfection
PubMed: 36414167
DOI: 10.1016/j.jhin.2022.11.008 -
Current Gastroenterology Reports Jul 2021Hepatobiliary complications are common in Roux-en-Y gastric bypass (RYGB) patients. Despite development of multiple surgical and endoscopic access techniques over the... (Review)
Review
PURPOSE OF REVIEW
Hepatobiliary complications are common in Roux-en-Y gastric bypass (RYGB) patients. Despite development of multiple surgical and endoscopic access techniques over the years, ERCP using standard duodenoscope remains challenging in these patients due to the altered anatomy.
RECENT FINDINGS
Limited success with enteroscope-assisted and laparoscope-assisted ERCP led to the evolution of the novel EUS-directed transgastric ERCP (EDGE) procedure, with variations of this technique termed as Gastric Access Temporary for Endoscopy (GATE), EUS-guided TransGastric ERCP (EUS-TG-ERCP), EUS-guided GastroGastrostomy-assisted ERCP (EUS-GG-ERCP), and EUS-directed transgastric intervention (EDGI). EDGE has high technical (100%) and clinical success rates (60-100%), lower adverse event rate (1.5-7.6%), and up to 20% access stent migration rate; without any significant weight changes. EDGE has significantly shorter procedure time (73vs184min), post-procedural hospital stays (0.8vs2.65 days) and is more cost effective compared to other modalities. EDGE technique addresses the challenges of RYGB anatomy as a minimally invasive, clinically successful, fully endoscopic, and cost-effective option. We present a literature review of the EDGE technique from its inception to current, in addition to reviewing other access techniques, their advantages, disadvantages and outcomes.
Topics: Cholangiopancreatography, Endoscopic Retrograde; Endosonography; Gastric Bypass; Humans; Stomach
PubMed: 34212281
DOI: 10.1007/s11894-021-00808-3 -
The Journal of Hospital Infection Jan 2023Microbiological surveillance of endoscopes is a safety measure for verifying the quality of reprocessing procedures and identifying contaminated devices, but...
INTRODUCTION
Microbiological surveillance of endoscopes is a safety measure for verifying the quality of reprocessing procedures and identifying contaminated devices, but duodenoscope-related outbreaks are still reported.
AIM
To assess the effectiveness of duodenoscope reprocessing procedures in Italy.
METHODS
Between December 2019 and April 2020, data obtained from microbiological surveillance post-reprocessing in 15 Italian endoscopy units were collected. Sampling was carried out after reprocessing or during storage in a cabinet. In keeping with international guidelines and the Italian position paper, the micro-organisms were classified as high-concern organisms (HCOs) and low-concern organisms (LCOs).
FINDINGS
In total, 144 samples were collected from 51 duodenoscopes. Of these, 36.81% were contaminated: 22.92% were contaminated with HCOs and 13.89% were contaminated with LCOs [2.08% with an LCO load of 11-100 colony-forming units (CFU)/device and 0.69% with an LCO load of >100 CFU/device]. The contamination rate was 27.5% in samples collected after reprocessing, 40% in samples collected during storage in a cabinet that was compliant with EN 16442:2015 (C-I), and 100% in samples collected during storage in a cabinet that was not compliant with EN 16442:2015 (NC-I). The respective HCO rates were 15.00%, 27.27% and 66.67%. Correlation between LCO contamination and storage time was demonstrated (Spearman's rho=0.3701; P=0.0026). The Olympus duodenoscope TJFQ180V demonstrated the lowest rate of contamination (29.82%), although the contamination rate was 100% for duodenoscopes stored in an NC-I cabinet.
CONCLUSION
Microbiological surveillance, along with strict adherence to reprocessing protocols, may help to detect endoscope contamination at an early stage, and reduce the risk of duodenoscope-associated infections.
Topics: Humans; Equipment Contamination; Disinfection; Endoscopes; Duodenoscopes; Endoscopy, Gastrointestinal
PubMed: 36244520
DOI: 10.1016/j.jhin.2022.09.024 -
Revista Latino-americana de Enfermagem 2022to analyze the cleaning process of gastroscopes, colonoscopes and duodenoscopes in eight in-hospital health services.
OBJECTIVE
to analyze the cleaning process of gastroscopes, colonoscopes and duodenoscopes in eight in-hospital health services.
METHOD
a cross-sectional study conducted with 22 endoscopes (eight gastroscopes, eight colonoscopes and six duodenoscopes), and microbiological analysis of 60 samples of air/water channels (all endoscopes) and elevator (duodenoscopes), in addition to protein testing. Descriptive statistics with calculation of frequencies and central tendency measures was used in data analysis.
RESULTS
the processing of 22 endoscopes was monitored with microbiological analysis for 60 channels. In the pre-cleaning procedure, in 82.3% (14/17) of the devices, gauze was used in cleaning the insertion tube. Incomplete immersion of the endoscope in detergent solution occurred in 72.3% (17/22) of the cases, and in 63.6% (14/22) there was no standardization of filling-in of the channels. Friction of the biopsy channel was not performed in 13.6% (3/22) of the devices. In the microbiological analysis, 25% (7/32) of the samples from the stored endoscopes were positive for microbial growth (from 2x101 to 9.5x104 CFU/mL), while after processing, contamination was 32% (9/28). Protein residues in the elevator channel were detected in 33% of duodenoscopes.
CONCLUSION
the results indicate important gaps in the stages of pre-cleaning and cleaning of endoscopes that, associated with presence of protein residues and growth of microorganisms of epidemiological importance, indicate limitations in safety of the processing procedures, which can compromise the disinfection processes and, consequently, their safe use among patients subjected to such tests.
Topics: Humans; Equipment Contamination; Cross-Sectional Studies; Detergents; Endoscopes; Disinfection; Hospitals
PubMed: 36287399
DOI: 10.1590/1518-8345.5969.3684