Endoscopy International Open Jun 2022Duodenoscopes that are contaminated due to inadequate reprocessing are well-documented. However, studies have demonstrated poor reprocessing of other kinds of... (Review)
Review
Duodenoscopes that are contaminated due to inadequate reprocessing are well-documented. However, studies have demonstrated poor reprocessing of other kinds of endoscopes as well, including echoendoscopes, gastroscopes, and colonoscopes. We estimated the contamination rate beyond the elevator of gastrointestinal endoscopes based on available data. We searched PubMed and Embase from January 1, 2010 to October 10, 2020, for studies investigating contamination rates of reprocessed gastrointestinal endoscopes. A random-effects model was used to calculate the contamination rate of patient-ready gastrointestinal endoscopes. Subgroup analyses were conducted to investigate differences among endoscope types, countries, and colony-forming unit (CFU) thresholds. Twenty studies fulfilled the inclusion criteria, including 1,059 positive cultures from 7,903 samples. The total contamination rate was 19.98 % ± 0.024 (95 % confidence interval [Cl]: 15.29 %-24.68 %; I = 98.6 %). The contamination rates of colonoscope and gastroscope channels were 31.95 % ± 0.084 and 28.22 % ± 0.076, respectively. Duodenoscope channels showed a contamination rate of 14.41 % ± 0.029. The contamination rates among studies conducted in North America and Europe were 6.01 % ± 0.011 and 18.16% ± 0.053 %, respectively. The contamination rate among studies using a CFU threshold > 20 showed contamination of 30.36 % ± 0.094, whereas studies using a CFU threshold < 20 showed a contamination rate of 11 % ± 0.026. On average, 19.98 % of reprocessed gastrointestinal endoscopes may be contaminated when used in patients and varies between different geographies. These findings highlight that the elevator mechanism is not the only obstacle when reprocessing reusable endoscopes; therefore, guidelines should recommend more surveillance of the endoscope channels as well.
PubMed: 35692921
DOI: 10.1055/a-1795-8883 -
Gastroenterology & Hepatology Mar 2021
PubMed: 34035774
DOI: No ID Found -
Journal of Clinical Microbiology Feb 2016Recent outbreaks of carbapenem-resistant Enterobacteriaceae infections associated with duodenoscopes used for endoscopic retrograde cholangiopancreatography have...
Recent outbreaks of carbapenem-resistant Enterobacteriaceae infections associated with duodenoscopes used for endoscopic retrograde cholangiopancreatography have highlighted the challenge of cleaning and high-level disinfection of these instruments. The Food and Drug Administration has suggested that duodenoscope surveillance by microbiological culturing, along with strict adherence to reprocessing protocols, may help reduce the risk of duodenoscope-associated infection transmission. We developed and validated an effective, user-friendly duodenoscope sampling and culture protocol and compared its performance to the interim Centers for Disease Control and Prevention-recommended guidelines. Our protocol resulted in a 65% recovery rate for Gram-negative organisms, demonstrating a 2-fold increased recovery rate compared to the CDC method. The implementation of this protocol may increase the feasibility of duodenoscope surveillance for microbiology laboratories and endoscopy departments.
Topics: Bacteriological Techniques; Cholangiopancreatography, Endoscopic Retrograde; Cross Infection; Duodenoscopes; Humans
PubMed: 26582839
DOI: 10.1128/JCM.02754-15 -
Antimicrobial Agents and Chemotherapy Aug 2018With the dissemination of carbapenemase-producing (CPE) strains worldwide, carbapenem-hydrolyzing enzymes are increasingly reported among isolates of , the first...
With the dissemination of carbapenemase-producing (CPE) strains worldwide, carbapenem-hydrolyzing enzymes are increasingly reported among isolates of , the first hospital and community-acquired opportunistic pathogen. Here, we have performed an epidemiological survey of carbapenemase-producing (CP-) isolates received at the French National Reference Centre (F-NRC) in 2012 and 2013. Antimicrobial susceptibilities for last-resort antibiotics and antimicrobial compounds commonly used to treat urinary tract infections were determined by broth microdilution. Clonal relationship was assessed using repetitive sequence-based PCR (rep-PCR) and multilocus sequence typing (MLST). From this collection of 140 carbapenemase-producing isolates, 74% produced an OXA-48-like carbapenemase and 21% produced an NDM carbapenemase. A link with a foreign country was suspected for 37% of infected/colonized patients. Most of the isolates were from screening (56%) and from urine samples (26%). Colistin, fosfomycin, and nitrofurantoin possessed the most consistent activity, with 100%, 95%, and 96% isolates susceptible, respectively. A wide diversity of carbapenemase-producing isolates has been found (50 different sequence types [STs]). The most prevalent clones were (i) sequence type 38 (ST38) producing OXA-48 ( = 21), a clone linked to Turkey and North African countries, (ii) ST-90 producing OXA-204 ( = 9), which was responsible for an outbreak related to a contaminated duodenoscope, and (iii) ST-410 producing OXA-181 ( = 5), which was recovered from patients of different geographical origins. These specific clones might be considered high-risk clones for the dissemination of carbapenemases in The wide diversity of STs, combined with the increasing number of CP- isolates received by the F-NRC, suggests a likely dissemination of CP- isolates in the community.
Topics: Africa, Northern; Anti-Bacterial Agents; Bacterial Proteins; Carbapenems; Colistin; Epidemiological Monitoring; Escherichia coli; Escherichia coli Infections; Fosfomycin; France; Gene Expression; Genetic Variation; Humans; Incidence; Multilocus Sequence Typing; Nitrofurantoin; Phylogeny; Turkey; Urinary Tract Infections; beta-Lactamases
PubMed: 29866863
DOI: 10.1128/AAC.00266-18 -
Infection Control and Hospital... Sep 2023To evaluate variables that affect risk of contamination for endoscopic retrograde cholangiopancreatography and endoscopic ultrasound endoscopes. (Observational Study)
Observational Study
OBJECTIVE
To evaluate variables that affect risk of contamination for endoscopic retrograde cholangiopancreatography and endoscopic ultrasound endoscopes.
DESIGN
Observational, quality improvement study.
SETTING
University medical center with a gastrointestinal endoscopy service performing ∼1,000 endoscopic retrograde cholangiopancreatography and ∼1,000 endoscopic ultrasound endoscope procedures annually.
METHODS
Duodenoscope and linear echoendoscope sampling (from the elevator mechanism and instrument channel) was performed from June 2020 through September 2021. Operational changes during this period included standard reprocessing with high-level disinfection with ethylene oxide gas sterilization (HLD-ETO) was switched to double high-level disinfection (dHLD) (June 16, 2020-July 15, 2020), and duodenoscopes changed to disposable tip model (March 2021). The frequency of contamination for the co-primary outcomes were characterized by calculated risk ratios.
RESULTS
The overall pathogenic contamination rate was 4.72% (6 of 127). Compared to duodenoscopes, linear echoendoscopes had a contamination risk ratio of 3.64 (95% confidence interval [CI], 0.69-19.1). Reprocessing using HLD-ETO was associated with a contamination risk ratio of 0.29 (95% CI, 0.06-1.54). Linear echoendoscopes undergoing dHLD had the highest risk of contamination (2 of 18, 11.1%), and duodenoscopes undergoing HLD-ETO and the lowest risk of contamination (0 of 53, 0%). Duodenoscopes with a disposable tip had a 0% contamination rate (0 of 27).
CONCLUSIONS
We did not detect a significant reduction in endoscope contamination using HLD-ETO versus dHLD reprocessing. Linear echoendoscopes have a risk of contamination similar to that of duodenoscopes. Disposable tips may reduce the risk of duodenoscope contamination.
Topics: Humans; Cholangiopancreatography, Endoscopic Retrograde; Disinfection; Duodenoscopes; Endosonography; Equipment Contamination
PubMed: 36645014
DOI: 10.1017/ice.2022.319 -
The Journal of Hospital Infection Apr 2021At present, the most frequent method for processing flexible gastrointestinal (GI) endoscopes is cleaning followed by high-level disinfection as terminal sterilization...
BACKGROUND
At present, the most frequent method for processing flexible gastrointestinal (GI) endoscopes is cleaning followed by high-level disinfection as terminal sterilization is often not practicable. Post-processing monitoring studies consistently show high levels of positive cultures remaining on endoscopes, which can lead to patient infection and even fatality. The processing deficiency is attributed to the complex design of endoscopes, incomplete cleaning, formation of biofilms and lack of margin of safety with high-level disinfection.
OBJECTIVE
To demonstrate that flexible GI endoscopes can be practicably terminally sterilized.
METHODS
An endoscope sterilization cycle was developed in a vaporized hydrogen peroxide sterilization system. The cycle was used to study the sterilization of flexible GI endoscopes which included colonoscopes and duodenoscope and material compatibility for both original flexible GI endoscopes and those experimentally modified endoscopes using compatible materials.
RESULTS
Testing demonstrated that the vaporized hydrogen peroxide can sterilize flexible GI endoscopes (colonoscopes, duodenoscope) with a sterility assurance level of 10. Additionally, no recoverable survivors were detected when devices were artificially soiled with hard water and serum. Material compatibility test results demonstrated that replacing molybdenum disulphide lubricant with a graphite-based inert lubricant can make them compatible with vaporized hydrogen peroxide sterilizers.
CONCLUSION
Flexible GI endoscopes can be practicably terminally sterilized using vaporized hydrogen peroxide sterilization technologies if their materials are revised to become compatible.
Topics: Disinfection; Duodenoscopes; Endoscopes, Gastrointestinal; Equipment Contamination; Hydrogen Peroxide; Sterilization
PubMed: 33549767
DOI: 10.1016/j.jhin.2021.01.021 -
Journal of Clinical Microbiology Oct 2015Inadequate flexible endoscope reprocessing has been associated with infection outbreaks, most recently caused by carbapenem-resistant Enterobacteriaceae. Lapses in... (Review)
Review
Inadequate flexible endoscope reprocessing has been associated with infection outbreaks, most recently caused by carbapenem-resistant Enterobacteriaceae. Lapses in essential device reprocessing steps such as cleaning, disinfection/sterilization, and storage have been reported, but some outbreaks have occurred despite claimed adherence to established guidelines. Recommended changes in these guidelines include the use of sterilization instead of high-level disinfection or the use of routine microbial culturing to monitor efficacy of reprocessing. This review describes the current standards for endoscope reprocessing, associated outbreaks, and the complexities associated with both microbiological culture and sterilization approaches to mitigating the risk of infection associated with endoscopy.
Topics: Cross Infection; Decontamination; Disinfection; Endoscopes; Humans; Practice Guidelines as Topic
PubMed: 26202125
DOI: 10.1128/JCM.01394-15 -
Acta Gastro-enterologica Belgica 2021Data on procedural outcome and quality of endoscopic retrograde cholangiopancreatography (ERCP) in Belgian practice are scarce. The aim of this study is to assess...
BACKGROUND AND STUDY AIMS
Data on procedural outcome and quality of endoscopic retrograde cholangiopancreatography (ERCP) in Belgian practice are scarce. The aim of this study is to assess current status of ERCP-performance in Belgium.
METHODS
National multi-institutional survey (online questionnaire) among members of the Belgian Society of Gastrointestinal Endoscopy (BSGIE), conducted in the period June-August 2018. The RIZIV/INAMI provided real-life data on the total number of ERCPs performed in Belgium.
RESULTS
Forty-five responders completed the survey (for 43 centers performing ERCP), providing information for 8368 ERCPs performed in 45% (43/95) of institutions performing ERCP in Belgium. Fifty-eight percent of centers performed > 100 ERCPs/year and 7% of centers (n=3) performed < 50 ERCPs/year. According to the RIZIV/INAMI data, low case-volume centers are underrepresented in this survey. The most common ERCPindication was stone extraction (52%). 74% of endoscopists had more than 10 years of experience in performing ERCP. The majority of centers had their own written protocol (84%) for microbiological duodenoscope surveillance. Monitoring of cannulation rate and post-ERCP pancreatitis (PEP) was only performed in a minority of centers (30%). The majority of centers (76%) provided verbal informed consent relating to the ERCP-procedure ; a minority also requested a written informed consent (23%). 65% of centers systematically use NSAIDs for PEP prophylaxis.
CONCLUSION
This is the first survey of ERCP performance in Belgium. There were wide variations in practice. Adherence to key performance measures and measurement and evaluation of ERCP performance in daily practice at center and endoscopist level are not uniformly widespread.
Topics: Belgium; Cholangiopancreatography, Endoscopic Retrograde; Endoscopy, Gastrointestinal; Humans; Pancreatitis; Surveys and Questionnaires
PubMed: 33639696
DOI: 10.51821/84.1.128 -
Clinical and Translational... Aug 2020
Review
Topics: Cholangiopancreatography, Endoscopic Retrograde; Disinfection; Duodenoscopes; Equipment Design; Focus Groups; Humans; Infection Control; Qualitative Research; Surgical Wound Infection
PubMed: 32955193
DOI: 10.14309/ctg.0000000000000214 -
Endoscopy Feb 2022The risk of exogenous infections from endoscopic procedures is often cited as almost negligible (1 infection in 1.8 million procedures); however, this risk is based on...
BACKGROUND
The risk of exogenous infections from endoscopic procedures is often cited as almost negligible (1 infection in 1.8 million procedures); however, this risk is based on older literature and does not seem to match the number of infectious outbreaks due to contaminated duodenoscopes reported after endoscopic retrograde cholangiopancreatography (ERCP). Using Dutch data, we aimed to estimate the minimum risk of duodenoscope-associated infection (DAI) and colonization (DAC) in patients undergoing ERCP.
METHODS
A systematic literature search identified all DAI outbreaks in the Netherlands reported between 2008 and 2019. Included cases were confirmed by molecular matching of patient and duodenoscope cultures. Risk ratios were calculated based on the total number of ERCPs performed during the study period.
RESULTS
Three outbreaks were reported and published between 2008 and 2018, including 21 confirmed DAI cases and 52 confirmed DAC cases. The estimated number of ERCPs performed during the same period was 181 209-227 006. The calculated minimum estimated DAI risk was approximately 0.01 % and the minimum estimated DAC risk was 0.023 %-0.029 %.
CONCLUSIONS
The estimated risk of DAI in Dutch ERCP practice was at least 180 times higher than previously published risk estimates. The actual risk is likely to be (much) higher due to underreporting of infections caused by multidrug-resistant organisms and sensitive bacteria. Greater awareness by healthcare personnel involved in endoscopy and endoscope cleaning is required, as well as innovative technical solutions to contain and ultimately eliminate DAIs.
Topics: Cholangiopancreatography, Endoscopic Retrograde; Disease Outbreaks; Duodenoscopes; Equipment Contamination; Humans; Netherlands
PubMed: 33915575
DOI: 10.1055/a-1467-6294